光谱束成形在平扫胸部CT检查中的应用:一项关于剂量降低与图像质量的体模研究

Spectral Beam Shaping in Unenhanced Chest CT Examinations: A Phantom Study on Dose Reduction and Image Quality.
PubMed 文献 · Acad Radiol · 2017-10-18

采用新生儿、5岁儿童和成人三种人体物理体模,在第三代双源CT上以CAREkV半模式和CAREDose4D扫描,评估所有管电压(70-150 kV及Sn150 kV)及Sn100 kV匹配和半剂量条件下的CTDIvol与CNR。结果显示,成人体模在100 kV或Sn150 kV(锡滤过)时可降低辐射剂量,而儿童体模未显示剂量降低。

要点

  • 成人平扫胸部CT采用Sn150 kV可在CNR相当(54.4 vs 56.6)时降低CTDIvol(2.08 mGy vs 2.67 mGy)。
  • 新生儿体模采用Sn150 kV辐射剂量升高(0.75 mGy)且CNR未改善(96.9 vs 101.5)。
  • 儿童体模采用Sn150 kV辐射剂量升高(1.02 mGy)且CNR未改善(92.4 vs 95.8)。

核心结论

  • 成人体模平扫胸部CT在100 kV或Sn150 kV(锡滤过)时可降低辐射剂量,但儿童体模未显示剂量降低。

研究设计与样本

  • 体模研究,使用新生儿、5岁儿童和成人三种人体物理体模,在第三代双源CT上以CAREkV半模式和CAREDose4D(参考kV 120,参考mAs 50)扫描,管电压覆盖70-150 kV及Sn150 kV。

诊断标准与关键数值

  • 新生儿体模最优管电压范围为70-110 kV,CTDIvol为0.36-0.37 mGy。
  • 儿童体模最低CTDIvol出现在100-120 kV,为0.78-0.80 mGy。
  • 成人体模最低CTDIvol出现在140 kV和150 kV,分别为2.68 mGy和2.67 mGy。
  • Sn150 kV扫描成人CTDIvol为2.08 mGy,CNR为54.4,与150 kV的CNR 56.6相当。
  • Sn100-half扫描成人CNR为58.1,与150 kV的56.6相当。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: This study aimed to determine the optimal tube potential for unenhanced chest computed tomographies (CTs) with age-related phantoms.

MATERIALS AND METHODS: Three physical anthropomorphic phantoms (newborn, 5-year-old child, and adult) were scanned on a third-generation dual-source CT using CAREkV in semi-mode and CAREDose4D (ref. KV: 120; ref. mAs 50). Scans were performed with all available tube potentials (70-150 kV and Sn150 kV). The lowest volume computed tomography dose index (CTDIvol) was selected to perform additional Sn100-kV scans with matched and half (Sn100-half) CTDIvol value. Image quality was evaluated on the basis of contrast-to-noise ratio (CNR).

RESULTS: For the newborn phantom, 70-110 kV was selected as the optimal range (0.36-0.37 mGy). Using Sn150 kV led to an increase in radiation dose (0.75 mGy) without improving CNR (96.9 vs 101.5). Sn100-half showed a decrease in CNR (73.1 vs 101.5). The lowest CTDIvol for the child phantom was achieved between 100 and 120 kV (0.78-0.80 mGy). Using Sn150 kV increased radiation dose (1.02 mGy) without improvement of CNR (92.4 vs 95.8). At Sn100-half CNR was decreased (61.4 vs 95.8). For adults, 140 and 150 kV revealed the lowest CTDIvol (2.68 and 2.67 mGy). The Sn150 kV scan delivered comparable CNR (54.4 vs 56.6), but a lower CTDIvol (2.08 mGy). At Sn100-half CNR was comparable to the 150 kV scan (58.1 vs 56.6).

CONCLUSION: Unenhanced chest CT performed at 100 kV or 150 kV with tin filtration enables radiation dose reduction for the adult phantom, but not for the pediatric phantoms.

来源:Acad Radiol,查看原文。版权归原作者所有。