神经肿瘤PET数据定量准确性随发射扫描时长的变化:PET/MR与PET/CT对比

Quantification accuracy of neuro-oncology PET data as a function of emission scan duration in PET/MR compared to PET/CT.
PubMed 文献 · Eur J Radiol · 2017-08-26

该研究在体模和脑扫描中比较缩短采集时间对PET/CT与PET/MR定量准确性的影响,发现缩短帧时长会使RCmax和SUVmax显著升高,而RCA50和SUVA50偏差较小、更适合比较不同计数下的PET研究;使用后滤波并将帧时长分别缩短至4分钟(体模)和3分钟(患者)时未见统计学显著差异;作者认为在合适重建设置和定量指标下,帧时长可降至临床常规的30%。

要点

  • 缩短帧时长会显著升高RCmax和SUVmax,解读短采集PET时需注意该偏差。
  • RCA50和SUVA50在不同计数下偏差较小,更适合比较不同采集条件的PET研究。
  • 使用后滤波并将帧时长分别缩短至4分钟(体模)和3分钟(患者)时,定量准确性无统计学显著差异。

核心结论

  • 在混合氨基酸脑成像中,若采用合适重建设置和定量指标,帧时长(或注射活度)可潜在降至临床常规所用标准的30%,而PET图像定量准确性无显著变化;NEMA体模中帧时长低于4分钟时不建议用于获得定量且可重复的测量。

研究设计与样本

  • 研究采集20分钟18F-FDG体模测量和10分钟18F-FET脑扫描,设备为Biograph-True-Point-True-View PET/CT(n=8)和Biograph mMR PET/MR(n=16);列表模式数据反复分割为1至10分钟帧,采用3D-OSEM算法的两种重建设置重建:后滤波(OSEM)和不后滤波但带分辨率恢复(PSF)。

诊断标准与关键数值

  • 评估指标为恢复系数RCmax、RCA50和标准摄取值SUVmax、SUVA50。
  • 缩短帧时长时,体模RCmax和患者SUVmax显著升高。
  • RCA50和SUVA50的偏差分别显著更低,使其更适合比较不同计数下的PET研究。
  • 使用后滤波并将帧时长缩短至4分钟(RCA50,参考20分钟,体模)和3分钟(SUVA50,参考10分钟,患者)时,未观察到统计学显著差异。

临床应用与阅片要点

  • 作者结论认为,在混合氨基酸脑成像中,采用合适重建设置和定量指标时,帧时长(或注射活度)可潜在降至临床常规所用标准的30%,而定量准确性无显著变化;NEMA体模中帧时长低于4分钟时不建议用于获得定量且可重复的测量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To evaluate and compare the effect of reduced acquisition time, as a surrogate of injected activity, on the PET quantification accuracy in PET/CT and PET/MR imaging.

METHODS: Twenty min 18F-FDG phantom measurements and 10min 18F-FET brain scans were acquired in a Biograph-True-Point-True-View PET/CT (n=8) and a Biograph mMR PET/MR (n=16). Listmode data were repeatedly split into frames of 1min to 10min length and reconstructed using two different reconstruction settings of a 3D-OSEM algorithm: with post-filtering ("OSEM"), and without post-filtering but with resolution recovery ("PSF"). Recovery coefficients (RCmax, RCA50) and standard uptake values (SUVmax, SUVA50) were evaluated.

RESULTS: RCmax (phantom) and SUVmax (patients) increased significantly when reducing the frame duration. Significantly lower deviations were observed for RCA50 and SUVA50, respectively, making them more appropriate to compare PET studies at different number of counts. No statistical significant differences were observed when using post-filtering and reducing the frame time to 4min (RCA50, reference 20min, phantom) and to 3min (SUVA50, reference 10min, patients).

CONCLUSIONS: For hybrid aminoacid brain imaging, frame duration (or injected activity) can potentially be reduced to 30% of the standard used in clinical routine without significant changes on the quantification accuracy of the PET images if adequate reconstruction settings and quantitative measures are used. Frame times below 4min in the NEMA phantom are not advisable to obtain quantitative and reproducible measures.