MRI对乳腺组织扩张器的影响及安全使用建议

Effect of MRI on breast tissue expanders and recommendations for safe use.
指南 / 共识 · J Plast Reconstr Aesthet Surg · 2017-10-15

该离体研究评估三种品牌乳腺组织扩张器在三种MRI环境下的平移力、扭矩、磁场强度和温度变化,发现3T环境磁场最强且因制造商而异,充盈生理盐水、俯卧位及选用1.5T可减轻风险,作者认为在做好围操作期选择后可为携带扩张器患者考虑MRI。

要点

  • 3T MRI环境下磁场强度、偏转角和扭矩最大,且因制造商而异(Sientra > Mentor > Allergan)。
  • 生理盐水充盈扩张器需240 mL、空气充盈需360 mL才能使扭矩不可测,俯卧位可减轻扭矩效应。
  • 防止生理盐水充盈扩张器移位需120 g重量,空扩张器需870 g,温度无明显变化。

核心结论

  • 通过使用1.5T MRI、选择器械、以生理盐水充盈扩张器及俯卧位,可能降低既往描述的风险;在做好适当围操作期选择后,可为携带乳腺组织扩张器的患者考虑MRI,使获益大于风险。

研究设计与样本

  • 离体实验研究,评估三种品牌乳腺组织扩张器在三种MRI环境下的表现,采用偏转角度法测定平移力,以0-4分量表评估空、生理盐水充盈和空气充盈扩张器的扭矩,用高斯计测量磁场,测定防止空气和生理盐水充盈扩张器移位所需重量,用酒精温度计测量温度随时间变化。

关键影像表现/征象

  • 3T MRI环境下磁场强度、偏转角和扭矩最大,且因制造商而异(Sientra > Mentor > Allergan)。
  • 生理盐水充盈扩张器需240 mL、空气充盈需360 mL体积才能使扭矩不可测,俯卧位可减轻扭矩效应。
  • 防止生理盐水充盈组织扩张器移位需120 g重量,空扩张器需870 g。
  • 温度随时间无明显变化。

临床应用与阅片要点

  • 作者认为,既往描述的风险可通过使用1.5T MRI、选择器械、以生理盐水充盈扩张器及俯卧位来降低。
  • 作者认为,在做好适当围操作期选择后,可为携带乳腺组织扩张器的患者考虑MRI,使获益大于风险。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

INTRODUCTION: Ferromagnetic port-containing breast tissue expanders are currently labeled MRI-unsafe because of the presumption that magnets should not enter the machine. However, designating these devices as MRI-unsafe can lead to unnecessary procedures or suboptimal imaging choices. This study provides an ex vivo analysis of how breast tissue expanders behave when subjected to strong magnetic fields to determine which variables might affect clinical risk.

METHODS: Three different brands of tissue expanders were evaluated in three MRI environments. Translational force was determined using the deflection angle method. Torque on empty, saline-filled, and air-filled expanders was evaluated on a 0-4 scale. Magnetic field was measured using a gaussmeter. The weight required to prevent displacement of the expanders was determined for both air- and saline-filled expanders. Temperature over time was measured using an alcohol thermometer.

RESULTS: Magnetic field strength, deflection angle, and torque were the greatest in 3T MRI environments and varied by device manufacturer (Sientra > Mentor > Allergan). Saline-filled expanders required 240 mL and air-filled required 360 mL volume to make the torque undetectable, and the effect of torque could be mitigated with prone positioning. A weight of 120 g was required to prevent displacement of a saline-filled tissue expander and 870 g for an empty expander. There were no appreciable changes in temperature.

CONCLUSIONS: Previously described risks may be reduced by using a 1.5T MRI, device selection, filling expanders with saline, and prone positioning. MRI can be considered in patients with breast tissue expanders when appropriate peri-procedural choices have been made so that the benefits of undergoing MRI outweigh the risks.

来源:J Plast Reconstr Aesthet Surg,查看原文。版权归原作者所有。

引用

  1. Marano AA, Henderson PW, Prince MR, et al. Effect of MRI on breast tissue expanders and recommendations for safe use. J Plast Reconstr Aesthet Surg. 2017. PMID: 29046262. doi:10.1016/j.bjps.2017.07.012

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。