对比增强超声给药装置对微泡增强效果的影响:聚焦儿科应用

Influence of contrast-enhanced ultrasound administration setups on microbubble enhancement: a focus on pediatric applications.
PubMed 文献 · Pediatr Radiol · 2017-09-11

本研究采用体外流动模型和流式细胞术,探讨不同静脉给药装置对两种超声对比剂增强效果和微泡浓度的影响。结果显示,经中性置换接头注射和经三通旋塞垂直注射均导致两种对比剂增强显著下降(P<0.0001)。作者建议避免使用中性置换接头和垂直注射方式,以防止增强效果丢失。

要点

  • 经中性置换接头或经三通旋塞垂直注射,两种超声对比剂的增强效果均显著下降(P<0.0001)。
  • 加用三通旋塞(P>0.24)或使用儿科扩散导管(P>0.28)不影响增强效果。
  • 作者建议避免使用中性置换接头和垂直注射方式,以防增强效果丢失。

核心结论

  • 超声对比剂的增强效果取决于给药途径,但部分效应似乎与所用对比剂种类有关;为避免增强效果丢失,应避免使用中性置换接头和垂直注射。

研究设计与样本

  • 体外实验研究,使用流动模型(ATS Laboratories,Bridgeport,CT)和Acuson S3000超声系统(Siemens Healthineers,Mountain View,CA)配9L4探头,以Cadence脉冲序列模式定量增强效果;用LSRII流式细胞仪(BD Biosciences,San Jose,CA)测定微泡浓度;对比剂为Optison(GE Healthcare,Princeton,NJ)和Lumason(Bracco,Geneva,Switzerland);经1 mL注射器注射后用5 mL生理盐水冲洗,通过22G扩散导管(BD Medical,Franklin Lakes,NJ),比较经三通旋塞顺行注射(加或不加中性置换接头)、经三通旋塞垂直注射(加或不加接头)及不使用三通旋塞等给药方式,另经22G标准血管导管注射。

临床应用与阅片要点

  • 作者建议避免使用中性置换接头和垂直注射方式,以防增强效果丢失;加用三通旋塞或使用儿科扩散导管不影响增强效果,但部分效应似乎与所用超声对比剂种类有关。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: In pediatrics, contrast-enhanced ultrasound offers high-quality imaging with an excellent safety profile.

OBJECTIVE: To investigate the effects of varying intravenous administration setups on in vitro enhancement and concentration of two commercially available ultrasound contrast agents, taking into consideration potential pediatric applications.

MATERIALS AND METHODS: We quantified in vitro enhancement using a flow phantom (ATS Laboratories, Bridgeport, CT) and Acuson S3000 ultrasound system (Siemens Healthineers, Mountain View, CA) with a 9L4 probe in Cadence pulse sequencing mode. We determined microbubble concentration with an LSRII flow cytometer (BD Biosciences, San Jose, CA). We investigated Optison (GE Healthcare, Princeton, NJ) and Lumason (Bracco, Geneva, Switzerland) ultrasound contrast agents. The ultrasound (US) contrast agent was injected via a 1 mL syringe and flushed with 5 mL of saline through a 22-gauge diffusion catheter (BD Medical, Franklin Lakes, NJ) with the following variations: in-line injection through a 3-way stopcock with and without a neutral displacement connector (ICU Medical, San Clemente, CA), perpendicular through a 3-way stopcock with and without a connector, and without a 3-way stopcock. We also conducted injections through a 22-gauge standard angiocatheter.

RESULTS: Injection through the connector and perpendicular injection via the 3-way stopcock resulted in significant decreases in enhancement for both ultrasound contrast agents (P<0.0001). Injection through the connector resulted in significant decrease in concentration for Optison (P<0.05). Neither addition of the 3-way stopcock (P>0.24) nor use of a pediatric diffusion catheter (P>0.28) affected the enhancement.

CONCLUSION: Ultrasound contrast agent enhancement depends on the administration route, although some effects appear to be specific to the ultrasound contrast agent used. To avoid loss of enhancement, neutral displacement connectors and perpendicular injection should be avoided.

来源:Pediatr Radiol,查看原文。版权归原作者所有。