利用虚拟单能图像重建和金属伪影减少算法减少脊柱骨科硬件在能谱探测器CT检查中产生的伪影

Reduction of artifacts caused by orthopedic hardware in the spine in spectral detector CT examinations using virtual monoenergetic image reconstructions and metal-artifact-reduction algorithms.
PubMed 文献 · Skeletal Radiol · 2017-09-21

纳入28例脊柱骨科硬件患者行能谱探测器CT,重建常规迭代、O-MAR及40-200 keV虚拟单能图像。O-MAR和高keV单能图像较常规迭代显著减少低密度伪影并降低噪声,主观上最多24/28例伪影减轻,二者联合似乎有望进一步减少伪影。

要点

  • O-MAR和高keV虚拟单能图像均可减少脊柱骨科硬件伪影
  • 高keV单能图像在椎旁肌肉噪声降低上未达统计学显著
  • O-MAR与高keV单能图像主观上最多24/28例伪影减轻

核心结论

  • 能谱探测器CT的O-MAR和高keV虚拟单能图像重建均带来客观和主观伪影减少,O-MAR与单能图像的组合似乎有望进一步减少伪影。

研究设计与样本

  • 连续纳入28例脊柱骨科硬件行能谱探测器CT检查的患者,重建常规迭代、O-MAR及40-200 keV虚拟单能图像,以常规迭代为对照。

关键影像表现/征象

  • O-MAR和高keV单能图像较常规迭代显示低密度伪影显著减少,表现为衰减值更高:常规迭代-396.5 HU vs O-MAR -115.2 HU、200 keV单能图像-48.1 HU,均p≤0.001。
  • 椎旁肌肉噪声:常规迭代34.7±19.0 HU vs O-MAR 26.4±14.4 HU(p≤0.05),常规迭代 vs 200 keV单能图像27.4±16.1 HU(n.s.)。
  • 椎管噪声:常规迭代103.4±61.3 HU vs O-MAR 72.6±62.6 HU、200 keV单能图像60.9±40.1 HU,均p≤0.001。
  • 两名放射科医师主观评估显示O-MAR和高keV图像在最多24/28例患者中产生伪影减轻。

诊断标准与关键数值

  • 以常规迭代重建为对照,O-MAR和高keV单能图像在低密度伪影(衰减值)和椎管噪声上均p≤0.001。
  • 椎旁肌肉噪声常规迭代 vs O-MAR p≤0.05,常规迭代 vs 200 keV单能图像n.s.。

临床应用与阅片要点

  • 脊柱骨科硬件术后CT阅片时,O-MAR和高keV虚拟单能图像可减少金属伪影、改善椎旁肌肉和椎管显示。
  • 高keV单能图像对椎旁肌肉噪声的降低未达统计学显著,阅片时需结合O-MAR综合评估。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Aim of this study was to assess the artifact reduction in patients with orthopedic hardware in the spine as provided by (1) metal-artifact-reduction algorithms (O-MAR) and (2) virtual monoenergetic images (MonoE) as provided by spectral detector CT (SDCT) compared to conventional iterative reconstruction (CI).

METHODS: In all, 28 consecutive patients with orthopedic hardware in the spine who underwent SDCT-examinations were included. CI, O-MAR and MonoE (40-200 keV) images were reconstructed. Attenuation (HU) and noise (SD) were measured in order to calculate signal-to-noise ratio (SNR) of paravertebral muscle and spinal canal. Subjective image quality was assessed by two radiologists in terms of image quality and extent of artifact reduction.

RESULTS: O-MAR and high-keV MonoE showed significant decrease of hypodense artifacts in terms of higher attenuation as compared to CI (CI vs O-MAR, 200 keV MonoE: -396.5HU vs. -115.2HU, -48.1HU; both p ≤ 0.001). Further, artifacts as depicted by noise were reduced in O-MAR and high-keV MonoE as compared to CI in (1) paravertebral muscle and (2) spinal canal-CI vs. O-MAR/200 keV: (1) 34.7 ± 19.0 HU vs. 26.4 ± 14.4 HU, p ≤ 0.05/27.4 ± 16.1, n.s.; (2) 103.4 ± 61.3 HU vs. 72.6 ± 62.6 HU/60.9 ± 40.1 HU, both p ≤ 0.001. Subjectively both O-MAR and high-keV images yielded an artifact reduction in up to 24/28 patients.

CONCLUSION: Both, O-MAR and high-keV MonoE reconstructions as provided by SDCT lead to objective and subjective artifact reduction, thus the combination of O-MAR and MonoE seems promising for further reduction.

来源:Skeletal Radiol,查看原文。版权归原作者所有。

引用

  1. Große Hokamp N, Neuhaus V, Abdullayev N, et al. Reduction of artifacts caused by orthopedic hardware in the spine in spectral detector CT examinations using virtual monoenergetic image reconstructions and metal-artifact-reduction algorithms. Skeletal Radiol. 2017. PMID: 28932962. doi:10.1007/s00256-017-2776-5