3T膝关节合成MRI的初步经验:与传统T1加权成像及T2定量图的比较
回顾性纳入19例膝关节正常者,比较合成MRI与多回波自旋回波序列的T2值及T1加权图像对比。软骨和肌肉T2值无显著差异,骨髓T2值MESE序列显著高于合成序列(124.3±3.6 ms vs 73.1±5.3 ms),骨髓对肌肉相对对比传统T1加权更高,骨髓对软骨相对对比相似。合成MRI可同时获得常规图像与定量图,有望缩短检查时间,除骨髓外图像质量与传统MRI相当,作者认为经优化后可能有助于早期骨关节炎评估。
要点
- 软骨和肌肉T2值合成MRI与MESE序列无显著差异,但骨髓T2值MESE显著高于合成序列。
- 骨髓对肌肉相对对比传统T1加权图像显著更高,骨髓对软骨相对对比两序列相似。
- 合成MRI可一次采集同时获得常规图像和定量图,有望缩短总检查时间,但骨髓图像质量需进一步优化。
核心结论
- 合成MRI能同时采集常规图像与定量图,有望缩短整体检查时间;除骨髓外,其图像质量与传统MRI相当,经进一步优化后可利用合成成像的肌骨组织图像质量优势。
研究设计与样本
- 回顾性研究,纳入19例膝关节检查正常者,在3T设备上同时行合成MRI与多回波自旋回波序列进行T2定量,测量软骨、骨髓和肌肉的T2值及T1加权序列上各结构的相对信号强度和相对对比。
关键影像表现/征象
- 软骨和肌肉的平均T2值在多回波自旋回波序列与合成MRI之间无显著差异。
- 骨髓平均T2值多回波自旋回波序列为124.3±3.6 ms,显著高于合成采集的73.1±5.3 ms。
- 两种方法各结构的相对信号强度无显著差异。
- 骨髓对肌肉的相对对比在传统T1加权图像上显著更高,骨髓对软骨的相对对比两序列相似。
诊断标准与关键数值
- 以多回波自旋回波序列T2定量为参照,合成MRI在软骨和肌肉的T2值与之无显著差异,骨髓T2值显著低于参照序列(73.1±5.3 ms vs 124.3±3.6 ms)。
临床应用与阅片要点
- 合成MRI可一次采集同时获得常规图像和定量图,有望缩短总检查时间,阅片时需注意骨髓信号与常规T1加权存在差异。
- 骨髓对肌肉相对对比在传统T1加权图像上显著更高,骨髓对软骨相对对比两序列相似,报告骨髓病变时需结合序列特点。
- 作者认为合成MRI图像对比良好且节省时间,可能有助于评估早期膝关节骨关节炎。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To assess the feasibility and accuracy of synthetic MRI compared to conventional T1 weighted and multi-echo spin-echo (MESE) sequences for obtaining T2 values in the knee joint at 3 Tesla.
METHODS: This retrospective study included 19 patients with normal findings in the knee joint who underwent both synthetic MRI and MESE pulse sequences for T2 quantification. T2 values of the two sequences at the articular cartilage, bone marrow and muscle were measured. Relative signal intensity (SI) of each structure and relative contrast among structures of the knee were measured quantitatively by T1 weighted sequences.
RESULTS: The mean T2 values for cartilage and muscle were not significantly different between MESE pulse sequences and synthetic MRI. For the bone marrow, the mean T2 value obtained by MESE sequences (124.3 ± 3.6 ms) was significantly higher than that obtained by synthetic acquisition (73.1 ± 5.3 ms). There were no significant differences in the relative SI of each structure between the methods. The relative contrast of bone marrow to muscle was significantly higher with conventional T1 weighted images, while that for bone marrow to cartilage was similar for both sequences.
CONCLUSION: Synthetic MRI is able to simultaneously acquire conventional images and quantitative maps, and has the potential to reduce the overall examination time. It provides comparable image quality to conventional MRI for the knee joint, with the exception of the bone marrow. With further optimization, it will be possible to take advantage of the image quality of musculoskeletal tissue with synthetic imaging. Advances in knowledge: Synthetic MRI produces images of good contrast and is also a time-saving technique. Thus, it may be useful for assessing osteoarthritis in the knee joint in the early stages.