数字乳腺X线摄影和乳腺断层摄影中能否减少乳腺压迫
研究分两部分:第一部分在未成像条件下评估四级压迫力对乳腺厚度、感知疼痛和组织覆盖的影响;第二部分在同一乳腺标准与减压迫力下各采集一幅MLO DBT图像,由三名乳腺影像放射科医生评判模糊伪影和组织覆盖。结果显示压迫力平均降低48%(CC位)和47%(MLO位)可显著减轻感知疼痛,压迫后乳腺厚度增加但组织覆盖无变化、运动模糊无增加。作者认为乳腺X线摄影和DBT可能可以使用目前一半的压迫力完成采集。
要点
- 压迫力平均降低约一半(CC位48%、MLO位47%)可显著减轻感知疼痛。
- 减压迫力后压迫乳腺厚度增加,但组织覆盖无变化且运动模糊无增加。
- 仅第一次减力可行,进一步减力导致乳腺制动不足。
核心结论
- 乳腺X线摄影和DBT采集可能可以使用目前所用压迫力的一半完成,感知疼痛显著且大幅降低,乳腺厚度和组织覆盖无临床显著变化。
研究设计与样本
- 两部分研究,招募两组有既往乳腺X线摄影的女性;第一部分受试者摆位CC和MLO体位并施加四级压迫力,不进行成像,评估乳腺厚度、感知疼痛和相对组织覆盖变化。
- 第二部分对每例患者一侧乳腺在标准和减低的压迫力下各采集两幅MLO DBT图像,由三名乳腺影像放射科医生评判模糊伪影和组织覆盖,并记录压迫力、乳腺厚度、相对组织覆盖和感知疼痛。
诊断标准与关键数值
- CC位压迫力平均降低48%、MLO位平均降低47%,感知疼痛水平显著降低。
- 第一部分压迫后乳腺厚度分别增加0.02 cm(CC位)和0.09 cm(MLO位);第二部分MLO位增加0.38 cm。
- 减压迫力后相对组织覆盖无变化,运动模糊无增加。
- 仅第一次减力可行,进一步减力导致乳腺制动不足。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The objective of this study was to investigate the impact of decreasing breast compression during digital mammography and breast tomosynthesis (DBT) on perceived pain and image quality.
MATERIALS AND METHODS: In this two-part study, two groups of women with prior mammograms were recruited. In part 1, subjects were positioned for craniocaudal (CC) and mediolateral oblique (MLO) views, and four levels of compression force were applied to evaluate changes in breast thickness, perceived pain, and relative tissue coverage. No imaging was performed. In part 2, two MLO DBT images of one breast of each patient were acquired at standard and reduced compression. Blurring artifacts and tissue coverage were judged by three breast imaging radiologists, and compression force, breast thickness, relative tissue coverage, and perceived pain were recorded.
RESULTS: Only the first reduction in force was feasible because further reduction resulted in inadequate breast immobilization. Mean force reductions of 48% and 47% for the CC and MLO views, respectively, resulted in a significantly reduced perceived pain level, whereas the thickness of the compressed breast increased by 0.02 cm (CC view) and 0.09 (MLO view, part 1 of the study) and 0.38 cm (MLO view, part 2 of the study), respectively, with no change in tissue coverage or increase in motion blurring.
CONCLUSION: Mammography and DBT acquisitions may be possible using half of the compression force used currently, with a significant and substantial reduction in perceived pain with no clinically significant change in breast thickness and tissue coverage.
引用
- Agasthya GA, D'Orsi E, Kim YJ, et al. Can Breast Compression Be Reduced in Digital Mammography and Breast Tomosynthesis? AJR Am J Roentgenol. 2017. PMID: 28929809. doi:10.2214/AJR.16.17615
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。