PubMed 文献 · Invest Radiol · 2018-01-01

前瞻性研究纳入59例患者,用GRASP自由呼吸动态T1WI监测钆塞酸注射后短暂运动,40.0%(23/59)出现短暂运动(平均持续21.5±13.0秒),其中73.9%(17/23)在注射后15秒内发生;呼吸门控重建可显著减少晚期动脉期运动伪影并提高图像质量和病灶检出效能。

要点

  • 钆塞酸注射后约40%患者出现短暂运动,多发生于注射后15秒内,平均持续约21.5秒。
  • 呼吸门控GRASP重建可显著减少晚期动脉期运动伪影,门控13.3秒图像质量最高。
  • 仅门控6秒T1WI较非门控13.3秒T1WI显著提高病灶检出效能(0.69 vs 0.60,P=0.004)。

核心结论

  • 钆塞酸给药后短暂运动发生率40%(23/59),GRASP呼吸门控自由呼吸T1WI能在出现短暂运动的患者中持续提供可接受的动脉期图像。

研究设计与样本

  • 经机构审查委员会批准的前瞻性研究,纳入59例知情同意患者,3 T下行GRASP自由呼吸动态T1WI,钆塞酸标准剂量0.025 mmol/kg、注射速率1 mL/s。

诊断标准与关键数值

  • 短暂运动定义为自门控MR信号的呼吸频率明显改变;短暂运动见于40.0%(23/59)患者,平均持续21.5±13.0秒,其中73.9%(17/23)在给药后15秒内发生;晚期动脉期运动伪影评分门控13.3秒3.64±0.34、门控6秒3.61±0.36,均优于非门控13.3秒3.12±0.51(P<0.0001);总体图像质量门控13.3秒3.76±0.39最高,其次门控6秒3.39±0.55、非门控13.3秒2.57±0.57(P<0.0001);病灶检出效能门控6秒0.69(0.63-0.76)vs 非门控13.3秒0.60(0.56-0.65),P=0.004。

临床应用与阅片要点

  • 作者结论认为,GRASP呼吸门控自由呼吸T1WI能在出现短暂运动的患者中持续提供可接受的动脉期图像。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The aims of this study were to observe the pattern of transient motion after gadoxetic acid administration including incidence, onset, and duration, and to evaluate the clinical feasibility of free-breathing gadoxetic acid-enhanced liver magnetic resonance imaging using golden-angle radial sparse parallel (GRASP) imaging with respiratory gating.

MATERIALS AND METHODS: In this institutional review board-approved prospective study, 59 patients who provided informed consents were analyzed. Free-breathing dynamic T1-weighted images (T1WIs) were obtained using GRASP at 3 T after a standard dose of gadoxetic acid (0.025 mmol/kg) administration at a rate of 1 mL/s, and development of transient motion was monitored, which is defined as a distinctive respiratory frequency alteration of the self-gating MR signals. Early arterial, late arterial, and portal venous phases retrospectively reconstructed with and without respiratory gating and with different temporal resolutions (nongated 13.3-second, gated 13.3-second, gated 6-second T1WI) were evaluated for image quality and motion artifacts. Diagnostic performance in detecting focal liver lesions was compared among the 3 data sets.

RESULTS: Transient motion (mean duration, 21.5 ± 13.0 seconds) was observed in 40.0% (23/59) of patients, 73.9% (17/23) of which developed within 15 seconds after gadoxetic acid administration. On late arterial phase, motion artifacts were significantly reduced on gated 13.3-second and 6-second T1WI (3.64 ± 0.34, 3.61 ± 0.36, respectively), compared with nongated 13.3-second T1WI (3.12 ± 0.51, P < 0.0001). Overall, image quality was the highest on gated 13.3-second T1WI (3.76 ± 0.39) followed by gated 6-second and nongated 13.3-second T1WI (3.39 ± 0.55, 2.57 ± 0.57, P < 0.0001). Only gated 6-second T1WI showed significantly higher detection performance than nongated 13.3-second T1WI (figure of merit, 0.69 [0.63-0.76]) vs 0.60 [0.56-0.65], P = 0.004).

CONCLUSIONS: Transient motion developed in 40% (23/59) of patients shortly after gadoxetic acid administration, and gated free-breathing T1WI using GRASP was able to consistently provide acceptable arterial phase imaging in patients who exhibited transient motion.