采用磁共振成像测量肝脏质子密度脂肪分数的线性、偏倚与精确性:一项Meta分析
该Meta分析纳入23项研究(1679名受试者)评估线性与偏倚、11项研究(425名受试者)评估精确性,结果显示MR成像-PDFF与MR波谱-PDFF高度线性(R²=0.96),平均Bland-Altman偏倚-0.13%(95%一致性界限:-3.95%,3.40%),重复性和再现性系数分别为2.99%和4.12%,场强、设备制造商和重建方法对再现性影响极小。
要点
- MR成像-PDFF与MR波谱-PDFF高度线性(R²=0.96),斜率0.97。
- 平均偏倚-0.13%(95%一致性界限:-3.95%,3.40%),提示轻度低估。
- 感兴趣区水平重复性系数2.99%、再现性系数4.12%。
核心结论
- MR成像-PDFF在不同场强、设备制造商和重建方法下均具有优异的线性、偏倚和精确性。
研究设计与样本
- Meta分析,按照PRISMA指南进行,系统检索评估MR成像-PDFF线性/偏倚或精确性的研究,使用QUADAS-2工具评价质量,汇总去标识化原始数据集;线性与偏倚分析纳入23项研究(1679名受试者),精确性分析纳入11项研究(425名受试者);参考标准为配准的MR波谱-PDFF。
诊断标准与关键数值
- 线性:MR成像-PDFF与MR波谱-PDFF回归R²=0.96,斜率0.97(P<.001);偏倚:平均Bland-Altman偏倚-0.13%(95%一致性界限:-3.95%,3.40%);精确性:感兴趣区水平重复性系数2.99%、再现性系数4.12%。
局限性
- 本文为2017年在线发表文章的更正版本,更正于2017年10月2日。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Purpose To determine the linearity, bias, and precision of hepatic proton density fat fraction (PDFF) measurements by using magnetic resonance (MR) imaging across different field strengths, imager manufacturers, and reconstruction methods. Materials and Methods This meta-analysis was performed in accordance with Preferred Reporting Items for Systematic Reviews and Meta-Analyses guidelines. A systematic literature search identified studies that evaluated the linearity and/or bias of hepatic PDFF measurements by using MR imaging (hereafter, MR imaging-PDFF) against PDFF measurements by using colocalized MR spectroscopy (hereafter, MR spectroscopy-PDFF) or the precision of MR imaging-PDFF. The quality of each study was evaluated by using the Quality Assessment of Studies of Diagnostic Accuracy 2 tool. De-identified original data sets from the selected studies were pooled. Linearity was evaluated by using linear regression between MR imaging-PDFF and MR spectroscopy-PDFF measurements. Bias, defined as the mean difference between MR imaging-PDFF and MR spectroscopy-PDFF measurements, was evaluated by using Bland-Altman analysis. Precision, defined as the agreement between repeated MR imaging-PDFF measurements, was evaluated by using a linear mixed-effects model, with field strength, imager manufacturer, reconstruction method, and region of interest as random effects. Results Twenty-three studies (1679 participants) were selected for linearity and bias analyses and 11 studies (425 participants) were selected for precision analyses. MR imaging-PDFF was linear with MR spectroscopy-PDFF (R2 = 0.96). Regression slope (0.97; P < .001) and mean Bland-Altman bias (-0.13%; 95% limits of agreement: -3.95%, 3.40%) indicated minimal underestimation by using MR imaging-PDFF. MR imaging-PDFF was precise at the region-of-interest level, with repeatability and reproducibility coefficients of 2.99% and 4.12%, respectively. Field strength, imager manufacturer, and reconstruction method each had minimal effects on reproducibility. Conclusion MR imaging-PDFF has excellent linearity, bias, and precision across different field strengths, imager manufacturers, and reconstruction methods. © RSNA, 2017 Online supplemental material is available for this article. An earlier incorrect version of this article appeared online. This article was corrected on October 2, 2017.
引用
- Yokoo T, Serai SD, Pirasteh A, et al. Linearity, Bias, and Precision of Hepatic Proton Density Fat Fraction Measurements by Using MR Imaging: A Meta-Analysis. Radiology. 2017. PMID: 28892458. doi:10.1148/radiol.2017170550
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。