新型超声弹性成像技术测量肝脏剪切波变异性研究

Variability of Liver Shear Wave Measurements Using a New Ultrasound Elastographic Technique.
PubMed 文献 · J Ultrasound Med · 2017-09-29

在55例混合病因慢性肝病患者中评估二维剪切波弹性成像(2D SWE)的测量变异性,发现中位IQR/中位数比值为0.131,5次测量即可近似10次测量的精度;肥胖、腹壁厚度增加、肝包膜下测量及ROI SD/速度比>0.15与变异性增加相关。

要点

  • 2D SWE总体测量变异性低,5次读数可提供与10次采样相当的精度。
  • ROI SD/速度比>0.15提示剪切波轮廓不均,偏差大于≤0.15者(0.42 vs 0.22 m/s)。
  • 肝包膜下1.5 cm内测量、BMI及皮肤至肝包膜距离增加均与变异性升高相关。

核心结论

  • 2D SWE总体测量变异性低,最少5次读数即可提供与现有10次采样方法相当的精度。
  • 肥胖、腹壁厚度增加、肝包膜下测量及ROI SD/速度比>0.15均与测量变异性增加相关。
  • ROI SD/速度比值得进一步评估作为质量评估指标,以便操作者实时客观评估单次2D SWE测量的可靠性。

研究设计与样本

  • 研究纳入55例混合病因慢性肝病患者,使用Aplio 500超声系统(Toshiba Medical Systems Corporation,栃木,日本)进行2D SWE评估。
  • 每例患者由一名对先前所有读数不知情的操作者获取10次测量,结果结合病历中的临床信息分析。

关键影像表现/征象

  • 2D SWE的中位IQR/中位数比值为0.131(四分位距1-3,0.089-0.174)。
  • 5次读数提供的近似值在0.11 m/s以内,即10次测量中位速度4.2%以内。
  • 剪切波轮廓不均(ROI SD/速度比>0.15)的测量偏离集合中位速度大于ROI SD/速度比≤0.15者(0.42 vs 0.22 m/s;P=0.001)。

诊断标准与关键数值

  • ROI SD/速度比>0.15定义为剪切波轮廓不均,其偏离集合中位速度大于≤0.15者(0.42 vs 0.22 m/s;P=0.001)。
  • 5次读数提供的近似值在10次测量中位速度的0.11 m/s或4.2%以内。

临床应用与阅片要点

  • 2D SWE测量时建议获取至少5次读数以达到与10次采样相当的精度。
  • 测量应避开肝包膜下1.5 cm以内区域以减少变异性。
  • ROI SD/速度比>0.15提示单次测量可靠性下降,可作为实时质量评估指标进一步评估。

局限性

  • ROI SD/速度比作为质量评估指标值得进一步评估(原文warrants further evaluation)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: A new 2-dimensional (2D) shear wave elastographic (SWE) device has been developed for the noninvasive assessment of liver fibrosis. Guidelines on measurement acquisition parameters are not yet well established for this technique. Our study aimed to assess 2D SWE measurement variability and to determine the number of measurements required per patient to reliably assess liver stiffness.

METHODS: Two-dimensional SWE was assessed in 55 patients with mixed-etiology chronic liver disease on an Aplio 500 ultrasound system (Toshiba Medical Systems Corporation, Tochigi, Japan). Ten measurements were obtained per patient by an operator blinded to all preceding readings. Results were analyzed with clinical information obtained from medical records.

RESULTS: The median interquartile range/median ratio for 2D SWE was 0.131 (quartiles 1-3, 0.089-0.174). Five readings provided an approximation within 0.11 m/s, or 4.2% of the median velocity of 10 measurements. Factors associated with increased measurement variability included body mass index (ρ = 0.388; P = .01), increased skin-to-liver capsule distance (ρ = 0.426; P = .002), and measurements taken within 1.5 cm of the liver capsule (P < .001). Measurements with heterogeneous shear wave profiles (indicated by a region of interest [ROI] SD/speed ratio > 0.15) showed greater deviation from the set's median velocity than those with an ROI SD/speed ratio of 0.15 or lower (0.42 versus 0.22 m/s; P = .001).

CONCLUSIONS: Two-dimensional SWE showed low overall measurement variability, with a minimum of 5 readings providing equivalent precision to the existing method using 10 samples. Obesity, increasing abdominal wall thickness, subcapsular measurements and an ROI SD/speed ratio of greater than 0.15 were all associated with increased measurement variability. The ROI SD/speed ratio warrants further evaluation as a quality assessment metric, to allow objective operator assessment of individual 2D SWE measurement reliability in real time.

来源:J Ultrasound Med,查看原文。版权归原作者所有。

引用

  1. Nadebaum DP, Nicoll AJ, Sood S, et al. Variability of Liver Shear Wave Measurements Using a New Ultrasound Elastographic Technique. J Ultrasound Med. 2017. PMID: 28960385. doi:10.1002/jum.14375

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。