光子计数CT:冠状动脉支架的高分辨率成像
本体外研究将18种冠状动脉支架置于模拟心包脂肪的体模中,在相同扫描协议和重建参数下比较光子计数探测器(PCD)与常规能量积分探测器(EID)的CT成像特性。结果显示PCD在支架内可见性、支架内管腔直径和支架内衰减方面均优于EID,且噪声更低。
要点
- PCD在相同扫描协议和重建参数下,支架内管腔显示主观评分优于EID(P < 0.001)。
- PCD的支架内平均直径较EID分别增大28.8%(0度)和8.4%(90度)。
- PCD的平均噪声为5 ± 0.2 HU,显著低于EID的8.3 ± 0.2 HU(P < 0.001)。
核心结论
- 在匹配的CT扫描协议和相同图像重建参数下,PCD对冠状动脉支架的支架内管腔显示优于常规EID阵列。
研究设计与样本
- 体外研究,将18种不同冠状动脉支架扩张于3 mm直径塑料管中,充填稀释至120 kVp下250 HU的对比剂并密封,置于校准至120 kVp下-100 HU的充油定制体模中以模拟心包脂肪;体模以0度和90度两种角度放置,使用研究型双源PCD-CT扫描仪成像,A探测器子系统使用标准64排EID,B探测器子系统使用PCD(等中心处探测器像素尺寸0.250 × 0.250 mm);两种探测器系统在相同管电压(100 kVp)和管电流-时间乘积(100 mA)下采集,均使用B46f卷积核和相同重建参数重建;由两名独立盲法读者评估支架内可见性并测量噪声、支架内管腔直径和支架内衰减。
关键影像表现/征象
- EID和PCD在支架内可见性、噪声、支架内管腔直径和支架内衰减方面的最佳结果均出现在体模沿z轴0度位置,提示面内分辨率高于穿面分辨率。
- PCD获得的冠状动脉支架图像支架内可见性主观评分优于EID(P < 0.001)。
诊断标准与关键数值
- 支架内平均直径:0度位置PCD为0.85 ± 0.24 mm,EID为0.66 ± 0.21 mm,PCD较EID增大28.8%;90度位置PCD为0.83 ± 0.14 mm,EID为0.76 ± 0.13 mm,PCD较EID增大8.4%。
- 平均噪声:PCD为5 ± 0.2 HU,EID为8.3 ± 0.2 HU,PCD显著低于EID(P < 0.001)。
- 支架内衰减增加量:0度位置PCD为Δ 245 ± 163 HU,EID为Δ 156.5 ± 126 HU(P = 0.006);90度位置PCD为Δ 194 ± 141 HU,EID为Δ 126 ± 78 HU(P = 0.001);PCD的支架内衰减增加量显著低于EID。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: The aim of this study was to investigate computed tomography (CT) imaging characteristics of coronary stents using a novel photon-counting detector (PCD) in comparison with a conventional energy-integrating detector (EID).
MATERIALS AND METHODS: In this in vitro study, 18 different coronary stents were expanded in plastic tubes of 3 mm diameter, were filled with contrast agent (diluted to an attenuation of 250 Hounsfield units [HU] at 120 kVp), and were sealed. Stents were placed in an oil-filled custom phantom calibrated to an attenuation of -100 HU at 120 kVp for resembling pericardial fat. The phantom was positioned in the gantry at 2 different angles at 0 degree and 90 degrees relative to the z axis, and was imaged in a research dual-source PCD-CT scanner. Detector subsystem "A" used a standard 64-row EID, while detector subsystem "B" used a PCD, allowing high-resolution scanning (detector pixel-size 0.250 × 0.250 mm in the isocenter). Images were obtained from both detector systems at identical tube voltage (100 kVp) and tube current-time product (100 mA), and were both reconstructed using a typical convolution kernel for stent imaging (B46f) and using the same reconstruction parameters. Two independent, blinded readers evaluated in-stent visibility and measured noise, intraluminal stent diameter, and in-stent attenuation for each detector subsystem. Differences in noise, intraluminal stent diameter, and in-stent attenuation where tested using a paired t test; differences in subjective in-stent visibility were evaluated using a Wilcoxon signed-rank test.
RESULTS: Best results for in-stent visibility, noise, intraluminal stent diameter, and in-stent attenuation in EID and PCD were observed at 0-degree phantom position along the z axis, suggesting higher in-plane compared with through-plane resolution. Subjective in-stent visibility was superior in coronary stent images obtained from PCD compared with EID (P < 0.001). Mean in-stent diameter was 28.8% and 8.4% greater in PCD (0.85 ± 0.24 mm; 0.83 ± 0.14 mm) as compared with EID acquisitions (0.66 ± 0.21 mm; 0.76 ± 0.13 mm) for both 0-degree and 90-degree phantom positions, respectively. Average noise was significantly lower (P < 0.001) for PCD (5 ± 0.2 HU) compared with EID (8.3 ± 0.2 HU). The increase in in-stent attenuation (0 degree: Δ 245 ± 163 HU vs Δ 156.5 ± 126 HU; P = 0.006; 90 degrees: Δ 194 ± 141 HU vs Δ 126 ± 78 HU; P = 0.001) was significantly lower for PCD compared with EID acquisitions.
CONCLUSIONS: At matched CT scan protocol settings and identical image reconstruction parameters, the PCD yields superior in-stent lumen delineation of coronary artery stents as compared with conventional EID arrays.