婴儿胸部无需镇静的影像检查:16厘米宽探测器CT单轴旋转扫描的可行性

Imaging the Infant Chest without Sedation: Feasibility of Using Single Axial Rotation with 16-cm Wide-Detector CT.
PubMed 文献 · Radiology · 2017-09-25

前瞻性纳入32例婴儿(组1)在16厘米宽探测器CT上以单轴旋转模式行无需镇静的非增强胸部CT,并与30例使用64排CT镇静下常规螺旋扫描的历史对照组(组2)比较。两组体重、年龄、平均CT值、图像噪声及主观图像质量评分无统计学显著差异,组1扫描时间缩短83%、准备时间缩短57%、CTDI降低42%、DLP降低52%(均P<.05)。结论认为该方案可在提供相当图像质量的同时降低辐射剂量、准备时间和总扫描时间。

要点

  • 16厘米宽探测器CT单轴旋转可在无需镇静下完成婴儿胸部CT,图像质量与64排CT镇静螺旋扫描相当。
  • 组1扫描时间0.35秒 vs 组2 2.01秒±0.21,缩短83%。
  • 组1 CTDI 2.03 mGy±0.4 vs 组2 3.52 mGy±0.03,降低42%;DLP 27.07 mGy·cm±6.97 vs 55.84 mGy·cm±6.46,降低52%。

核心结论

  • 与镇静下常规64排螺旋CT相比,16厘米宽探测器CT单轴旋转用于婴儿胸部无需镇静成像可降低辐射剂量、准备时间和总扫描时间,同时提供相当的图像质量。

研究设计与样本

  • 前瞻性研究纳入32例婴儿(组1)在16厘米宽探测器CT上行无需镇静单轴旋转非增强胸部CT,历史对照组为30例(组2)使用64排容积CT在镇静下行常规螺旋扫描。

关键影像表现/征象

  • 两组体重、年龄、平均CT衰减值、图像噪声及主观图像质量评分无统计学显著差异。
  • 组1扫描时间0.35秒 vs 组2 2.01秒±0.21,组1缩短83%。
  • 组1准备时间41.25分钟±103.78 vs 组2 96.5分钟±151.77,组1缩短57%。

诊断标准与关键数值

  • 组1采用自动管电流调制,体重5 kg及以下管电压80 kVp,体重大于5 kg管电压100 kVp。
  • 组1 CTDI 2.03 mGy±0.4 vs 组2 3.52 mGy±0.03,组1降低42%(P<.05)。
  • 组1 DLP 27.07 mGy·cm±6.97 vs 组2 55.84 mGy·cm±6.46,组1降低52%(P<.05)。

临床应用与阅片要点

  • 婴儿胸部CT可考虑采用16厘米宽探测器单轴旋转模式,在无需镇静条件下完成扫描。
  • 扫描参数可按体重选择管电压:5 kg及以下用80 kVp,大于5 kg用100 kVp,并采用自动管电流调制。

局限性

  • 对照组为历史对照,非同期随机对照。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Purpose To compare image quality, patient preparation time, and radiation dose using a single axial rotation with 16-cm wide-detector computed tomography (CT) in imaging the infant chest without sedation with those in infants examined by using a 64-row CT and sedation. Materials and Methods Thirty-two infants (group 1) were prospectively enrolled to undergo nonenhanced chest CT without sedation using a single axial rotation on a 16-cm wide-detector CT scanner. Patients were imaged with automatic tube current modulation and tube voltages of 80 kVp for patients weighing 5 kg or less and 100 kVp for patients weighing more than 5 kg. Patient preparation time, CT dose index (CTDI), dose-length product (DLP), and image quality were compared with those in a historical control group consisting of 30 infants (group 2) who underwent conventional helical scanning with sedation performed by using a 64-row volume CT scanner. The Student t test for independent samples was used to assess continuous variables. The Mann-Whitney rank test and the κ test were used to evaluate image quality. Results There was no statistically significant difference in body weight, age, mean CT attenuation value, image noise, and subjective image quality score between the two groups. However, compared with the group scanned by using a 64-row volume CT scanner (group 2), group 1 experienced significantly reduced scan time by 83% (0.35 second vs 2.01 seconds ± 0.21 [standard deviation]), preparation time by 57% (41.25 minutes ± 103.78 vs 96.5 minutes ± 151.77), CTDI by 42% (2.03 mGy ± 0.4 vs 3.52 mGy ± 0.03), and DLP by 52% (27.07 mGy·cm ± 6.97 vs 55.84 mGy·cm ± 6.46) (P < .05 for all). Conclusion Compared with conventional 64-row helical CT with sedation, use of a single axial rotation with 16-cm wide-detector CT in imaging the infant chest without sedation can reduce radiation dose, preparation time, and total scan time, while providing comparable image quality. © RSNA, 2017.