文献速递胸部技术与安全MR

自由呼吸肺部MRI用于放疗的可行性:基于非笛卡尔k空间采集方案

Feasibility of free breathing Lung MRI for Radiotherapy using non-Cartesian k-space acquisition schemes.
PubMed 文献 · Br J Radiol · 2017-09-13

在5名健康志愿者和10例肺癌患者中评估自由呼吸StarVIBE与SpiralVIBE序列,并与标准屏气VIBE及TWIST对比。屏气配合时VIBE解剖图像优于自由呼吸StarVIBE,但屏气不配合时StarVIBE更优;StarVIBE功能成像优于标准序列,SpiralVIBE超短回波可显示肺实质。

要点

  • 屏气不配合时自由呼吸StarVIBE解剖图像质量优于屏气VIBE。
  • 自由呼吸StarVIBE动态对比增强成像优于标准屏气TWIST序列。
  • SpiralVIBE超短回波序列可显示肺实质结构。

核心结论

  • StarVIBE与SpiralVIBE两种非笛卡尔采集序列可提供自由呼吸肺部成像方案,图像质量足以在放疗期间进行解剖、结构和功能评估。

研究设计与样本

  • 在5名健康志愿者和10例肺癌患者中,于3T宽孔径扫描仪(MAGNETOM Skyra,Siemens Healthcare,Erlangen,Germany)上评估基于T1加权3D梯度回波序列的StarVIBE和SpiralVIBE两种非笛卡尔采集方案。

关键影像表现/征象

  • 屏气配合的VIBE序列解剖图像质量优于自由呼吸StarVIBE。
  • 屏气不配合时StarVIBE解剖图像质量优于屏气VIBE。
  • StarVIBE功能评估优于标准序列,并在存在运动时提供稳健的动态对比增强数据。
  • SpiralVIBE的超短回波使肺实质得以显示。

临床应用与阅片要点

  • 作者结论认为StarVIBE与SpiralVIBE提供的自由呼吸肺部成像方案图像质量足以在放疗期间进行解剖、结构和功能评估。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To test a free-breathing MRI protocol for anatomical and functional assessment during lung cancer radiotherapy by assessing two non-Cartesian acquisition schemes based on T1 weighted 3D gradient recall echo sequence: (i) stack-of stars (StarVIBE) and (ii) spiral (SpiralVIBE) trajectories.

METHODS: MR images on five healthy volunteers were acquired on a wide bore 3T scanner (MAGNETOM Skyra, Siemens Healthcare, Erlangen, Germany). Anatomical image quality was assessed on: (1) free breathing (StarVIBE), (2) the standard clinical sequence (volumetric interpolated breath-hold examination, VIBE) acquired in a 20 second (s) compliant breath-hold and (3) 20 s non-compliant breath-hold. For functional assessment, StarVIBE and the current standard breath-hold time-resolved angiography with stochastic trajectories (TWIST) sequence were run as multiphase acquisitions to replicate dynamic contrast enhancement (DCE) in one healthy volunteer. The potential application of the SpiralVIBE sequence for lung parenchymal imaging was assessed on one healthy volunteer. Ten patients with lung cancer were subsequently imaged with the StarVIBE and SpiralVIBE sequences for anatomical and structural assessment. For functional assessment, free-breathing StarVIBE DCE protocol was compared with breath-hold TWIST sequences on four prior lung cancer patients with similar tumour locations. Image quality was evaluated independently and blinded to sequence information by an experienced thoracic radiologist.

RESULTS: For anatomical assessment, the compliant breath-hold VIBE sequence was better than free-breathing StarVIBE. However, in the presence of a non-compliant breath-hold, StarVIBE was superior. For functional assessment, StarVIBE outperformed the standard sequence and was shown to provide robust DCE data in the presence of motion. The ultrashort echo of the SpiralVIBE sequence enabled visualisation of lung parenchyma.

CONCLUSION: The two non-Cartesian acquisition sequences, StarVIBE and SpiralVIBE, provide a free-breathing imaging protocol of the lung with sufficient image quality to permit anatomical, structural and functional assessment during radiotherapy. Advances in knowledge: Novel application of non-Cartesian MRI sequences for lung cancer imaging for radiotherapy. Illustration of SpiralVIBE UTE sequence as a promising sequence for lung structural imaging during lung radiotherapy.

来源:Br J Radiol,查看原文。版权归原作者所有。

引用

  1. Kumar S, Rai R, Stemmer A, et al. Feasibility of free breathing Lung MRI for Radiotherapy using non-Cartesian k-space acquisition schemes. Br J Radiol. 2017. PMID: 28937270. doi:10.1259/bjr.20170037

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。