文献速递神经头颈外周血管CT

光子计数计算机断层扫描用于头颈部血管成像:首次在体人体研究结果

Photon-Counting Computed Tomography for Vascular Imaging of the Head and Neck: First In Vivo Human Results.
PubMed 文献 · Invest Radiol · 2018-01-01

本研究纳入16例无症状受试者(7例男性,平均年龄58岁,范围46-75岁),在同一受试者中对比光子计数探测器(PCD)CT与能量积分探测器(EID)单能CT的头颈部增强扫描图像质量。结果显示PCD图像质量评分显著更高,图像噪声降低9.1%(8.0±1.3 HU vs 8.8±1.5 HU,P<0.001),伪影更少(P<0.001),碘图对比噪声比(CNR)较非光谱PCD提高20.7%(65.2±9.0 vs 54.0±4.5,P=0.01)。作者结论认为光子计数CT有潜力改善颈动脉及颅内CT血管成像的图像质量。

要点

  • PCD图像噪声较EID降低9.1%(8.0±1.3 HU vs 8.8±1.5 HU,P<0.001),图像质量评分更高。
  • PCD图像伪影评分显著优于EID(P<0.001),EID上因线束硬化出现的伪影在PCD图像上未见。
  • 碘图CNR较非光谱PCD提高20.7%(65.2±9.0 vs 54.0±4.5,P=0.01),70 keV虚拟单能图CNR与非光谱图像相似(52.6±4.2 vs 54.0±4.5,P=0.39)。

核心结论

  • 光子计数CT相比单能EID CT有潜力改善颈动脉及颅内CT血管成像的图像质量。

研究设计与样本

  • 经机构审查委员会批准的研究,16例无症状受试者(7例男性,平均年龄58岁,范围46-75岁)均接受PCD与EID头颈部增强CT扫描;管设置EID为120 kVp/160 mA,PCD为140 kVp/108 mA;两名对探测器类型设盲的放射科医师独立评分;采用线性混合模型、线性回归及Bland-Altman分析(95%一致性界限)。

关键影像表现/征象

  • EID图像上动脉节段因线束硬化出现伪影,而PCD图像上未见这些伪影。
  • PCD图像噪声较EID降低9.1%(8.0±1.3 HU vs 8.8±1.5 HU,P<0.001)。
  • PCD图像质量评分显著高于EID,图像噪声更低(P<0.01)、伪影更少(P<0.001)。

诊断标准与关键数值

  • 头模PCD图像中实际碘浓度与计算碘浓度相关性极好(R=0.998),无显著偏倚(偏倚-0.4 mg/mL,95%一致性界限-1.1至0.4 mg/mL)。
  • 碘图CNR较非光谱PCD提高20.7%(65.2±9.0 vs 54.0±4.5,P=0.01)。
  • 70 keV虚拟单能图CNR与非光谱图像相似(52.6±4.2 vs 54.0±4.5,P=0.39)。
  • 两名放射科医师对PCD与EID图像各动脉节段图像质量、诊断可接受性、图像噪声及伪影严重程度进行评分,采用组内相关系数评估可重复性。

临床应用与阅片要点

  • 作者结论认为光子计数CT有潜力改善颈动脉及颅内CT血管成像的图像质量,尚未给出具体阅片流程建议。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: The purpose of this study was to evaluate image quality of a spectral photon-counting detector (PCD) computed tomography (CT) system for evaluation of major arteries of the head and neck compared with conventional single-energy CT scans using energy-integrating detectors (EIDs).

METHODS: In this institutional review board-approved study, 16 asymptomatic subjects (7 men) provided informed consent and received both PCD and EID contrast-enhanced CT scans of the head and neck (mean age, 58 years; range, 46-75 years). Tube settings were (EID: 120 kVp/160 mA vs PCD: 140 kVp/108 mA) for all volunteers. Quantitative analysis included measurements of mean attenuation, image noise, and contrast-to-noise ratio (CNR). Spectral PCD data were used to reconstruct virtual monoenergetic images and iodine maps. A head phantom was used to validate iodine concentration measurements in PCD images only. Two radiologists blinded to detector type independently scored the image quality of different segments of the arteries, as well as diagnostic acceptability, image noise, and severity of artifacts of the PCD and EID images. Reproducibility was assessed with intraclass correlation coefficient. Linear mixed models that account for within-subject correlation of analyzed arterial segments were used. Linear regression and Bland-Altman analysis with 95% limits of agreement were used to calculate the accuracy of material decomposition.

RESULTS: Photon-counting detector image quality scores were significantly higher compared with EID image quality scores with lower image noise (P < 0.01) and less image artifacts (P < 0.001). Photon-counting detector image noise was 9.1% lower than EID image noise (8.0 ± 1.3 HU vs 8.8 ± 1.5 HU, respectively, P < 0.001). Arterial segments showed artifacts on EID images due to beam hardening that were not present on PCD images. On PCD images of the head phantom, there was excellent correlation (R = 0.998) between actual and calculated iodine concentrations without significant bias (bias: -0.4 mg/mL [95% limits of agreements: -1.1 to 0.4 mg/mL]). Iodine maps had 20.7% higher CNR compared with nonspectral PCD (65.2 ± 9.0 vs 54.0 ± 4.5, P = 0.01), and virtual monoenergetic image at 70 keV showed similar CNR to nonspectral images (52.6 ± 4.2 vs 54.0 ± 4.5, P = 0.39).

CONCLUSIONS: Photon-counting CT has the potential to improve the image quality of carotid and intracranial CT angiography compared with single-energy EID CT.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Symons R, Reich DS, Bagheri M, et al. Photon-Counting Computed Tomography for Vascular Imaging of the Head and Neck: First In Vivo Human Results. Invest Radiol. 2018. PMID: 28926370. doi:10.1097/RLI.0000000000000418

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。