文献速递外周血管CTDSA

基于体重指数调整的超低剂量CT血管成像方案用于外周动脉:图像质量、诊断准确性与辐射暴露

A BMI-adjusted ultra-low-dose CT angiography protocol for the peripheral arteries-Image quality, diagnostic accuracy and radiation exposure.
PubMed 文献 · Eur J Radiol · 2017-06-03

这项前瞻性研究纳入40例外周动脉疾病患者,采用80kV管电压、按体重指数(BMI)个体化设置管电流调制参考幅值(BMI≤25为120mAs,25<BMI≤30为150mAs)并应用迭代重建的超低剂量CT血管成像方案,以数字减影血管造影为参考标准。该方案诊断准确率达94%(阅片者1)和93%(阅片者2),相比低剂量机构标准方案显著降低CTDIvol(1.97±0.55mGy vs 4.18±0.62mGy)和DLP(256±81mGy×cm vs 544±83mGy×cm),而图像质量相似(CNR的p=0.37)。

要点

  • 超低剂量方案诊断准确率达94%(阅片者1)和93%(阅片者2)。
  • 相比机构标准方案,CTDIvol和DLP显著降低(p<0.001),CNR相似(p=0.37)。
  • 不同BMI组间图像质量相似(CNR的p=0.86)。

核心结论

  • 采用低kV设置、按BMI个体化管电流调制参考幅值并联合迭代重建的CT方案,可实现极低辐射暴露的CTA,同时在外周动脉疾病评估中保持良好图像质量和高诊断准确性。

研究设计与样本

  • 前瞻性、经机构审查委员会批准的研究,纳入40例外周动脉疾病患者(30例男性,平均年龄72岁),在双源CT扫描仪上以80kV管电压行CTA,以数字减影血管造影为参考标准。

诊断标准与关键数值

  • 以数字减影血管造影为参考标准,超低剂量方案诊断准确率为94%(阅片者1)和93%(阅片者2)。
  • 管电流调制参考幅值按BMI个体化设置:BMI≤25为120mAs,25<BMI≤30为150mAs。
  • 与机构标准方案相比,CTDIvol为1.97±0.55mGy vs 4.18±0.62mGy,DLP为256±81mGy×cm vs 544±83mGy×cm,均显著降低(p<0.001)。
  • 与机构标准方案相比,CNR相似(p=0.37);不同BMI组间CNR相似(p=0.86)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To investigate radiation exposure, objective image quality, and the diagnostic accuracy of a BMI-adjusted ultra-low-dose CT angiography (CTA) protocol for the assessment of peripheral arterial disease (PAD), with digital subtraction angiography (DSA) as the standard of reference.

METHODS: In this prospective, IRB-approved study, 40 PAD patients (30 male, mean age 72 years) underwent CTA on a dual-source CT scanner at 80kV tube voltage. The reference amplitude for tube current modulation was personalized based on the body mass index (BMI) with 120 mAs for [BMI≤25] or 150 mAs for [25<BMI≤30]. Iterative image reconstruction was applied. The presence of significant stenoses (>70%) was assessed by two readers independently and compared to subsequent DSA. Radiation exposure was assessed with the computed tomography dose index (CTDIvol) and the dosis-length product (DLP). Objective image quality was assessed via contrast- and signal-to-noise ratio (CNR and SNR) measurements. Radiation exposure and image quality were compared between the BMI groups and between the BMI-adjusted ultra-low-dose protocol and the low-dose institutional standard protocol (ISP).

RESULTS: The BMI-adjusted ultra-low-dose protocol reached high diagnostic accuracy values of 94% for Reader 1 and 93% for Reader 2. Moreover, in comparison to the ISP, it showed significantly (p<0.001) lower CTDIvol (1.97±0.55mGy vs. 4.18±0.62 mGy) and DLP (256±81mGy x cm vs. 544±83mGy x cm) but similar image quality (p=0.37 for CNR). Furthermore, image quality was similar between BMI groups (p=0.86 for CNR).

CONCLUSIONS: A CT protocol that incorporates low kV settings with a personalized (BMI-adjusted) reference amplitude for tube current modulation and iterative reconstruction enables very low radiation exposure CTA, while maintaining good image quality and high diagnostic accuracy in the assessment of PAD.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Schreiner MM, Platzgummer H, Unterhumer S, et al. A BMI-adjusted ultra-low-dose CT angiography protocol for the peripheral arteries-Image quality, diagnostic accuracy and radiation exposure. Eur J Radiol. 2017. PMID: 28668409. doi:10.1016/j.ejrad.2017.06.002

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。