文献速递外周血管技术与安全USMR

三维超声向量血流成像测量颈总动脉血流并与磁共振成像验证

Common Carotid Artery Flow Measured by 3-D Ultrasonic Vector Flow Imaging and Validated with Magnetic Resonance Imaging.
PubMed 文献 · Ultrasound Med Biol · 2017-07-12

研究比较了颈总动脉三维超声向量血流成像与穿透平面磁共振成像序列,用首位志愿者数据作训练集,对其余八位志愿者计算血流量和峰值流速。三维超声所测收缩期峰值流速平均较MRI高34%,血流量平均较MRI低24%;PSV相关性高(r=0.79),血流量相关性较低(r=0.43)。改变回波消除滤波器参数重新处理后,血流量差异缩小至-9%,总体标准差改善至±4.7%。

要点

  • 三维超声所测颈总动脉PSV平均较MRI高34%,血流量平均较MRI低24%。
  • PSV相关性高(r=0.79),血流量相关性较低(r=0.43)。
  • 改变回波消除滤波器参数后,三维超声与MRI血流量差异缩小至-9%。

核心结论

  • 三维超声用于体内精确且角度无关的血流量和PSV估测是可行的。

研究设计与样本

  • 将颈总动脉超声检查与穿透平面磁共振成像序列比较,以验证近期提出的三维超声向量血流成像技术;首位志愿者数据用作训练集,其余八位志愿者计算血流量和峰值流速。

诊断标准与关键数值

  • 三维超声所测收缩期峰值流速平均较MRI高34%。
  • 三维超声所测血流量平均较MRI低24%。
  • PSV相关性r=0.79,血流量相关性r=0.43。
  • 三维超声血流量和PSV总体标准差均为±5.7%,MRI分别为±2.7%和±3.2%。
  • 改变回波消除滤波器参数重新处理后,三维向量血流成像与MRI血流量差异缩小至-9%,总体标准差改善至±4.7%。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Ultrasound (US) examination of the common carotid artery was compared with a through-plane magnetic resonance imaging (MRI) sequence to validate a recently proposed technique for 3-D US vector flow imaging. Data from the first volunteer examined were used as the training set, before volume flow and peak velocities were calculated for the remaining eight volunteers. Peak systolic velocities (PSVs) and volume flow obtained with 3-D US were, on average, 34% higher and 24% lower than those obtained with MRI, respectively. A high correlation was observed for PSV (r = 0.79), whereas a lower correlation was observed for volume flow (r = 0.43). The overall standard deviations were ±5.7% and ±5.7% for volume flow and PSV with 3-D US, compared with ±2.7% and ±3.2% for MRI. Finally, the data were re-processed with a change in the parameter settings for the echo-canceling filter to investigate its influence on overall performance. PSV was less affected by the re-processing, whereas the difference in volume flow between 3-D vector flow imaging and MRI was reduced to -9%, and with an improved overall standard deviation of ±4.7%. The results illustrate the feasibility of using 3-D US for precise and angle-independent volume flow and PSV estimation in vivo.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Holbek S, Hansen KL, Bouzari H, et al. Common Carotid Artery Flow Measured by 3-D Ultrasonic Vector Flow Imaging and Validated with Magnetic Resonance Imaging. Ultrasound Med Biol. 2017. PMID: 28711283. doi:10.1016/j.ultrasmedbio.2017.06.007

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。