双层探测器CT虚拟单能图像减少金属伪影
研究纳入35例在科隆大学医院接受双层探测器CT检查且扫描区域含骨科植入物的患者,重建64keV、70keV、105keV、140keV、200keV及个体化优化光子能量虚拟单能图像,由三名盲法观察者评估。高光子能量下诊断图像质量Likert评分从4.3(±0.83)降至2.3(±1.02),伪影评分从4.3(±0.66)降至2.6(±2.57),平均优化光子能量为149.2±39.4keV。结论认为虚拟单能图像可减少金属植入物相关伪影,140keV似为最佳预设,20%患者个体化优化keV可较140keV进一步改善图像质量。
要点
- 高光子能量虚拟单能图像可显著改善金属植入物区域的诊断图像质量并减少伪影。
- 140keV作为虚拟单能图像预设似可提供最佳伪影减少效果。
- 约20%患者个体化优化keV可较140keV进一步改善图像质量。
核心结论
- 使用虚拟单能图像可通过减少金属植入物相关伪影获得诊断图像质量的改善,140keV预设似提供最佳伪影减少,20%患者个体化优化keV可较140keV进一步改善图像质量。
研究设计与样本
- 研究纳入35例在科隆大学医院接受双层探测器CT检查且扫描区域含骨科植入物的患者,对每个双层探测器CT重建64keV、70keV、105keV、140keV、200keV及个体化优化光子能量的虚拟单能图像,由三名盲法观察者分析主观指标(伪影范围、诊断图像质量)和客观指标(伪影宽度和密度)。
关键影像表现/征象
- 21例植入物位于脊柱,8例位于骨盆,6例位于四肢。
- 高光子能量下诊断图像质量Likert评分从4.3(±0.83)改善至2.3(±1.02)。
- 伪影评分从4.3(±0.66)降至2.6(±2.57)。
- 最明显伪影的密度从-374.6±251.89HU降至-12.5±205.84HU。
- 最明显伪影的宽度从14.5±8.74mm降至6.4±10.76mm。
诊断标准与关键数值
- 平均优化光子能量为149.2±39.4keV。
- 140keV作为虚拟单能图像预设似可提供最佳伪影减少。
- 20%患者个体化优化keV可较140keV进一步改善图像质量。
临床应用与阅片要点
- 双层探测器CT虚拟单能图像可用于减少骨科金属植入物相关伪影并改善诊断图像质量。
- 140keV可作为虚拟单能图像的预设 photon energy,似提供最佳伪影减少。
- 对约20%患者,个体化优化keV可较140keV进一步改善图像质量。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The aim of the study was to investigate the performance and diagnostic value of metal artifact reduction in virtual monoenergetic images generated from dual-layer computed tomography (DLCT).
METHODS: 35 patients that received a DLCT at the University Hospital Cologne and had an orthopedic implant in the examined region were included in this study. For each DLCT virtual monoenergetic images of different energy levels (64keV, 70keV, 105keV, 140keV, 200keV and an optimized photon energy) were reconstructed and analyzed by three blinded observers. Images were analyzed with regard to subjective criteria (extent of artifacts, diagnostic image quality) and objective criteria (width and density of artifacts).
RESULTS: 21 patients had implants in the spine, 8 in the pelvis and 6 patients in the extremities. Diagnostic image quality improved significantly at high photon energies from a Likert-score of 4.3 (±0.83) to 2.3 (±1.02) and artifacts decreased significantly from a score of 4.3 (±0.66) to 2.6 (±2.57). The average optimized photon energy was 149.2±39.4keV. The density as well as the width of the most pronounced artifacts decreased from-374.6±251.89HU to -12.5±205.84HU and from 14.5±8.74mm to 6.4±10.76mm, respectively.
CONCLUSION: Using virtual monoenergetic images valuable improvements of diagnostic image quality can be achieved by reduction of artifacts associated with metal implants. As preset for virtual monoenergetic images, 140keV appear to provide optimal artifact reduction. In 20% of the patients, individually optimized keV can lead to a further improvement of image quality compared to 140keV.
引用
- Neuhaus V, Große Hokamp N, Abdullayev N, et al. Metal artifact reduction by dual-layer computed tomography using virtual monoenergetic images. Eur J Radiol. 2017. PMID: 28668408. doi:10.1016/j.ejrad.2017.05.013