自动容积乳腺X线摄影乳腺密度测量可能低估高密度乳腺的乳腺密度百分比
本研究比较两种商用软件(Volpara和Quantra)与MRI分割所测乳腺密度百分比(PBD)的一致性。共纳入110例女性220侧乳腺,结果显示Volpara和Quantra均倾向于低估PBD,且在较高密度时差异最明显;不同厂商的PBD值不可互换使用。
要点
- Volpara与MRI的PBD总体无显著差异,但Quantra与MRI差异小而有统计学意义。
- 两种软件在高密度乳腺中均倾向低估PBD,Volpara略低估、Quantra略高估。
- 不同厂商自动容积软件的PBD值不可互换使用。
核心结论
- Volpara和Quantra均倾向于低估PBD,在较高密度时最为明显;自动容积软件可准确测量PBD,但不同厂商的数值不应互换使用。
研究设计与样本
- 经机构审查委员会批准、符合健康保险流通与责任法案的回顾性研究;连续纳入6个月内MRI筛查阴性或良性 findings 的110例女性共220侧乳腺,比较MRI分割与两种商用软件(Volpara和Quantra)从乳腺X线摄影「for processing」图像得出的PBD。
诊断标准与关键数值
- PBD总体MRI均值10%(范围1%-41%)与Volpara均值10%(范围3%-29%)无差异;MRI与Quantra差异4.0%(95%CI:2.9至5.0,P<0.001)。总乳腺体积差异Volpara和Quantra与MRI均无显著差异(P=0.89和0.35)。
临床应用与阅片要点
- 阅片和报告中使用自动容积软件测量PBD时,应注意高密度乳腺可能被低估,且不同厂商的PBD值不可互换。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: The purpose of this study was to evaluate discrepancy in breast composition measurements obtained from mammograms using two commercially available software methods for systematic trends in overestimation or underestimation compared to magnetic resonance-derived measurements.
MATERIALS AND METHODS: An institutional review board-approved, Health Insurance Portability and Accountability Act-compliant retrospective study was performed to calculate percent breast density (PBD) by quantifying fibroglandular volume and total breast volume derived from magnetic resonance imaging (MRI) segmentation and mammograms using two commercially available software programs (Volpara and Quantra). Consecutive screening MRI exams from a 6-month period with negative or benign findings were used. The most recent mammogram within 9 months was used to derive mean density values from "for processing" images at the per breast level. Bland-Altman statistical analyses were performed to determine the mean discrepancy and the limits of agreement.
RESULTS: A total of 110 women with 220 breasts met the study criteria. Overall, PBD was not different between MRI (mean 10%, range 1%-41%) and Volpara (mean 10%, range 3%-29%); a small but significant difference was present in the discrepancy between MRI and Quantra (4.0%, 95% CI: 2.9 to 5.0, P < 0.001). Discrepancy was highest at higher breast densities, with Volpara slightly underestimating and Quantra slightly overestimating PBD compared to MRI. The mean discrepancy for both Volpara and Quantra for total breast volume was not significantly different from MRI (p = 0.89, 0.35, respectively). Volpara tended to underestimate, whereas Quantra tended to overestimate fibroglandular volume, with the highest discrepancy at higher breast volumes.
CONCLUSIONS: Both Volpara and Quantra tend to underestimate PBD, which is most pronounced at higher densities. PBD can be accurately measured using automated volumetric software programs, but values should not be used interchangeably between vendors.
引用
- Rahbar K, Gubern-Merida A, Patrie JT, et al. Automated Volumetric Mammographic Breast Density Measurements May Underestimate Percent Breast Density for High-density Breasts. Acad Radiol. 2017. PMID: 28754209. doi:10.1016/j.acra.2017.06.002
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。