流速、壁运动滤波、脉冲重复频率与混叠对能量多普勒图像定量影响的研究
该研究利用血流模拟流体流经体模,采集射束形成超声射频数据,评估流速(5–40 cm/s)、脉冲重复频率(0.6–10 kHz)和壁运动滤波截止频率(50–250 Hz)对能量多普勒定量的影响。结果显示壁运动滤波和脉冲重复频率显著影响能量多普勒定量,尤其在流速低于20 cm/s时,且脉冲重复频率低于1.3 kHz时可观察到混叠效应的影响。
要点
- 壁运动滤波截止频率和脉冲重复频率会显著影响能量多普勒信号强度定量。
- 流速低于20 cm/s时,壁运动滤波和脉冲重复频率对定量的影响尤为明显。
- 脉冲重复频率低于1.3 kHz时,能量多普勒定量可能受混叠效应影响。
核心结论
- 壁运动滤波和脉冲重复频率会显著影响能量多普勒定量,尤其在流速低于20 cm/s时;脉冲重复频率低于1.3 kHz时定量可受混叠效应影响。
研究设计与样本
- 体模实验研究,血流模拟流体流经体模,使用射束形成超声射频数据,脉冲重复频率0.6–10 kHz,6种流速5–40 cm/s,壁运动滤波截止频率50–250 Hz。
关键影像表现/征象
- 能量多普勒幅度受壁运动滤波截止频率的深度影响。
- 不同壁运动滤波值的效果随脉冲重复频率变化,能量多普勒信号强度依赖于脉冲重复频率。
- 能量多普勒定量可受混叠效应影响,尤其在使用低于1.3 kHz的脉冲重复频率时。
临床应用与阅片要点
- 作者指出实验在非临床环境中进行,但所评估参数与临床实践中使用的参数等效,有助于未来研究中相关数据的解读。
局限性
- 实验在非临床环境中进行。
- 许多既往研究使用专有软件输出,设置和参数未知,结论可能无法推广。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Although power Doppler imaging has been used to quantify tissue and organ vascularity, many studies showed that limitations in defining adequate ultrasound machine settings and attenuation make such measurements complex to be achieved. However, most of these studies were conducted by using the output of proprietary software, such as Virtual Organ computer-aided analysis (GE Healthcare, Kretz, Zipf, Austria); therefore, many conclusions may not be generalizable because of unknown settings and parameters used by the software. To overcome this limitation, our goal was to evaluate the impact of the flow velocity, pulse repetition frequency (PRF), and wall motion filter (WMF) on power Doppler image quantification using beam-formed ultrasonic radiofrequency data.
METHODS: The setup consisted of a blood-mimicking fluid flowing through a phantom. Radiofrequency signals were collected using PRFs ranging from 0.6 to 10 kHz for 6 different flow velocities (5-40 cm/s). Wall motion filter cutoff frequencies were varied between 50 and 250 Hz.
RESULTS: The power Doppler magnitude was deeply influenced by the WMF cutoff frequency. The effect of using different WMF values varied with the PRF; therefore, the power Doppler signal intensity was dependent on the PRF. Finally, we verified that power Doppler quantification can be affected by the aliasing effect, especially when using a PRF lower than 1.3 kHz.
CONCLUSIONS: The WMF and PRF greatly influenced power Doppler quantification, mainly when flow velocities lower than 20 cm/s were used. Although the experiments were conducted in a nonclinical environment, the evaluated parameters are equivalent to those used in clinical practice, which makes them valuable for aiding the interpretation of related data in future research.
引用
- Martins MR, Martins WP, Soares CAM, et al. Understanding the Influence of Flow Velocity, Wall Motion Filter, Pulse Repetition Frequency, and Aliasing on Power Doppler Image Quantification. J Ultrasound Med. 2017. PMID: 28736982. doi:10.1002/jum.14338
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。