文献速递腹部技术与安全CT

第三代供应商特定基于模型的迭代重建(Veo 3.0):使用新降噪预设和不同层面优化对腹部CT图像质量的评估

Third version of vendor-specific model-based iterativereconstruction (Veo 3.0): evaluation of CT image quality in the abdomen using new noise reduction presets and varied slice optimization.
PubMed 文献 · Br J Radiol · 2017-07-14

本回顾性研究利用30例连续腹部CT扫描原始数据,重建Veo 3.0四种临床预设的3.75mm轴位图像,由两名独立读者盲法随机评分并测量噪声、对比噪声比和空间分辨率。NR05预设总体图像质量和肝脏评估排名最佳,NR40在所有评估类别中显著较差。Veo 3.0预设允许选择图像噪声与空间分辨率的平衡。

要点

  • 在对比增强腹部CT中,5%降噪预设联合3.75mm层面优化(NR05)的总体图像质量和肝脏评估定性排名最佳。
  • 40%降噪预设(NR40)在所有评估类别中排名显著较差。
  • 与RP05n相比,NR05平均噪声降低21.01%,RP05c降低26.88%,NR40降低50.86%。

核心结论

  • Veo 3.0临床预设允许选择图像噪声与空间分辨率的平衡;对于对比增强腹部CT评估,5%降噪预设联合3.75mm层面优化(NR05)定性排名总体更优,且相对于其他序列在图像噪声和空间分辨率方面处于中间位置。

研究设计与样本

  • 回顾性研究,经机构审查委员会批准并符合健康保险携带和责任法案;使用30例连续腹部CT扫描患者的原始数据重建四种Veo 3.0临床预设的3.75mm轴位图像:RP05n(5%分辨率偏好)、NR05(5%降噪)和NR40(40%降噪)联合新的3.75mm层面优化,以及RP05联合常规0.625mm层面优化(RP05c);由两名独立读者盲法随机采用5点Likert量表和5点比较量表评估;在Catphan 600体模7 lp cm-1条形图案上绘制线轮廓评估空间分辨率。

关键影像表现/征象

  • 图像噪声和空间分辨率沿谱系增加,RP05n最高,NR40最低。
  • NR05图像集总体图像质量排名最佳(平均排名差异0.48,95%CI [0.081-0.88],p=0.01),肝脏评估排名最佳(平均差异0.46,95%CI [0.030-0.89],p=0.03),均与各自类别中第二佳序列比较。
  • RP05n在骨骼评估中排名最佳。
  • NR40在所有评估类别中排名显著较差。
  • NR05和RP05c图像集对比噪声比和空间分辨率几乎相同,但NR05通常更受偏好。

诊断标准与关键数值

  • 与RP05n相比,NR05平均噪声降低21.01%,RP05c降低26.88%,NR40降低50.86%。
  • NR05总体图像质量平均排名差异0.48(95%CI [0.081-0.88],p=0.01),肝脏评估平均差异0.46(95%CI [0.030-0.89],p=0.03)。

临床应用与阅片要点

  • 对于对比增强腹部CT评估,5%降噪预设联合3.75mm层面优化(NR05)定性排名总体更优。
  • Veo 3.0临床预设允许选择图像噪声与空间分辨率的平衡。
  • 本研究结果提供重要参数,指导临床和研究目的的预设选择。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To qualitatively and quantitatively compare abdominal CT images reconstructed with a newversion of model-based iterative reconstruction (Veo 3.0; GE Healthcare Waukesha, WI) utilizing varied presetsof resolution preference, noise reduction and slice optimization.

METHODS: This retrospective study was approved by our Institutional Review Board and was Health Insurance Portability and Accountability Act compliant. The raw datafrom 30 consecutive patients who had undergone CT abdomen scanning were used to reconstructfour clinical presets of 3.75mm axial images using Veo 3.0: 5% resolution preference (RP05n), 5%noise reduction (NR05) and 40% noise reduction (NR40) with new 3.75mm "sliceoptimization," as well as one set using RP05 with conventional 0.625mm "slice optimization" (RP05c). The images were reviewed by two independent readers in a blinded, randomized manner using a 5-point Likert scale as well as a 5-point comparative scale. Multiple two-dimensional circular regions of interest were defined for noise and contrast-to-noise ratio measurements. Line profiles were drawn across the 7 lp cm-1 bar pattern of the Catphan 600 phantom for evaluation of spatial resolution.

RESULTS: The NR05 image set was ranked as the best series in overall image quality (mean difference inrank 0.48, 95% CI [0.081-0.88], p = 0.01) and with specific reference to liver evaluation (meandifference 0.46, 95% CI [0.030-0.89], p = 0.03), when compared with the secondbest series ineach category. RP05n was ranked as the best for bone evaluation. NR40 was ranked assignificantly inferior across all assessed categories. Although the NR05 and RP05c image setshad nearly the same contrast-to-noise ratio and spatial resolution, NR05 was generally preferred. Image noise and spatial resolution increased along a spectrum with RP05n the highest and NR40the lowest. Compared to RP05n, the average noise was 21.01% lower for NR05, 26.88%lower for RP05c and 50.86% lower for NR40.

CONCLUSION: Veo 3.0 clinical presets allow for selection of image noise and spatial resolution balance; for contrast-enhanced CT evaluation of the abdomen, the 5% noise reduction preset with 3.75 mm slice optimization (NR05) was generally ranked superior qualitatively and, relative to other series, was in the middle of the spectrum with reference to image noise and spatial resolution. Advances in knowledge: To our knowledge, this is the first study of Veo 3.0 noise reduction presets and varied slice optimization. This study provides insight into the behaviour of slice optimization and documents the degree of noise reduction and spatial resolution changes that users can expect across various Veo 3.0 clinical presets. These results provide important parameters to guide preset selection for both clinical and research purposes.

来源:Br J Radiol,查看原文。版权归原作者所有。

引用

  1. Telesmanich ME, Jensen CT, Enriquez JL, et al. Third version of vendor-specific model-based iterativereconstruction (Veo 3.0): evaluation of CT image quality in the abdomen using new noise reduction presets and varied slice optimization. Br J Radiol. 2017. PMID: 28707531. doi:10.1259/bjr.20170188

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。