可吸收下腔静脉滤器预防猪肺栓塞的安全性与有效性
本研究在11头猪下腔静脉置入聚二氧环己酮缝线制成的可吸收滤器,随访至32周,未发现下腔静脉血栓、装置移位、腔静脉穿孔或肺栓塞;滤器逐渐吸收,植入后至少5周内可有效捕获血栓。
要点
- 可吸收下腔静脉滤器在猪模型中可安全置入,随访32周未见下腔静脉血栓、装置移位、腔静脉穿孔或肺栓塞。
- 滤器在置入后至少5周内可有效捕获自下腔静脉注入的自体血栓,捕获后1至4周内随滤器吸收。
- 组织学显示置入2周内出现下腔静脉滤器周围新生内膜增生,32周时腔静脉壁内仍可见聚二氧环己酮缝线残留微小碎片。
核心结论
- 可吸收下腔静脉滤器可安全置入猪体内,并在32周测试期内逐渐吸收;该装置在猪置入后至少5周内可有效预防肺栓塞。
研究设计与样本
- 研究经机构动物护理与使用委员会批准,为猪模型研究。
- 11头猪经导管在下腔静脉置入11枚由聚二氧环己酮缝线制成的可吸收下腔静脉滤器。
- 滤器留置时间分别为2周(n=2)、5周(n=2)、12周(n=2)、24周(n=2)和32周(n=3)。
- 7头猪在滤器置入后0至35天从滤器下方注入自体血栓。
- 滤器置入后第1至6周每周、第7至20周每两周、第21至32周每月进行透视和CT随访;尸检取肾下段下腔静脉、肺、心脏、肝、肾和脾。
关键影像表现/征象
- 随访期间未见下腔静脉血栓、装置移位、腔静脉穿孔或肺栓塞的证据。
- 大体病理分析显示装置在置入后至32周逐渐吸收。
- 组织学评估显示下腔静脉滤器置入后2周内出现滤器周围新生内膜增生,32周时腔静脉壁内仍可见聚二氧环己酮缝线残留微小碎片。
- 每例人工注入的血栓均被滤器成功捕获直至吸收(范围1至4周)。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Purpose To evaluate the immediate and long-term safety as well as thrombus-capturing efficacy for 5 weeks after implantation of an absorbable inferior vena cava (IVC) filter in a swine model. Materials and Methods This study was approved by the institutional animal care and use committee. Eleven absorbable IVC filters made from polydioxanone suture were deployed via a catheter in the IVC of 11 swine. Filters remained in situ for 2 weeks (n = 2), 5 weeks (n = 2), 12 weeks (n = 2), 24 weeks (n = 2), and 32 weeks (n = 3). Autologous thrombus was administered from below the filter in seven swine from 0 to 35 days after filter placement. Fluoroscopy and computed tomography follow-up was performed after filter deployment from weeks 1-6 (weekly), weeks 7-20 (biweekly), and weeks 21-32 (monthly). The infrarenal IVC, lungs, heart, liver, kidneys, and spleen were harvested at necropsy. Continuous variables were evaluated with a Student t test. Results There was no evidence of IVC thrombosis, device migration, caval penetration, or pulmonary embolism. Gross pathologic analysis showed gradual device resorption until 32 weeks after deployment. Histologic assessment demonstrated neointimal hyperplasia around the IVC filter within 2 weeks after IVC filter deployment with residual microscopic fragments of polydioxanone suture within the caval wall at 32 weeks. Each iatrogenic-administered thrombus was successfully captured by the filter until resorbed (range, 1-4 weeks). Conclusion An absorbable IVC filter can be safely deployed in swine and resorbs gradually over the 32-week testing period. The device is effective for the prevention of pulmonary embolism for at least 5 weeks after placement in swine. © RSNA, 2017.
引用
- Huang SY, Eggers M, McArthur MJ, et al. Safety and Efficacy of an Absorbable Filter in the Inferior Vena Cava to Prevent Pulmonary Embolism in Swine. Radiology. 2017. PMID: 28708470. doi:10.1148/radiol.2017161880
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。