文献速递腹部泌尿生殖CT

CT结肠成像中结直肠息肉的计算机辅助检测:前瞻性临床性能与第三方报销

Computer-Aided Detection of Colorectal Polyps at CT Colonography: Prospective Clinical Performance and Third-Party Reimbursement.
PubMed 文献 · AJR Am J Roentgenol · 2017-01-01

研究回顾性评估了FDA批准的CAD系统在第二读者模式下对347例连续成人CTC检查的前瞻性临床性能,并以经验丰富的CTC放射科医师前瞻性判读联合后续光学结肠镜为参考标准;对≥6 mm息肉,患者层面CAD敏感度为91.3%,息肉层面为88.4%,平均每序列假阳性标记4.4±3.1个;另对1225例CTC病例评估报销,CAD判读总费用中31.0%已从多家第三方支付方收回。

要点

  • CAD在第二读者模式下对≥6 mm结直肠息肉的 patient 层面敏感度为91.3%(63/69)。
  • 息肉层面CAD单独敏感度为88.4%(114/129),6–9 mm为88.3%(83/94),≥10 mm为88.6%(31/35)。
  • 平均每序列CAD假阳性标记为4.4±3.1个,回顾性复习发现3个≥6 mm息肉在CTC被当作CAD假阳性而忽略。

核心结论

  • 在作者所在机构的常规临床实践中,CAD对检测≥6 mm结直肠息肉显示出良好敏感度,假阳性标记数量可接受;CAD已在部分第三方支付方获得报销。

研究设计与样本

  • 回顾性评估FDA批准的CAD系统在第二读者模式下的前瞻性临床性能,纳入5个月期间连续347例成人(平均年龄57.6岁;205名女性、142名男性)CTC检查;参考标准为经验丰富的CTC放射科医师前瞻性判读联合后续光学结肠镜(如实施);另评估18个月期间CAD的第三方报销。

诊断标准与关键数值

  • 以光学结肠镜联合CTC放射科医师前瞻性判读为参考标准,CAD患者层面敏感度91.3%(63/69),息肉层面敏感度88.4%(114/129);按息肉大小,6–9 mm为88.3%(83/94),≥10 mm为88.6%(31/35);平均每序列假阳性标记4.4±3.1个。

临床应用与阅片要点

  • 在CTC第二读者模式下使用CAD辅助检测≥6 mm结直肠息肉;回顾性复习发现3个≥6 mm息肉在光学结肠镜可见且被CAD标记,但CTC判读时被当作CAD假阳性而忽略。

局限性

  • 原文为回顾性评估前瞻性临床性能。
  • 原文报告了CAD假阳性标记,平均每序列4.4±3.1个。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: We assessed the initial clinical performance and third-party reimbursement rates of supplementary computer-aided detection (CAD) at CT colonography (CTC) for detecting colorectal polyps 6 mm or larger in routine clinical practice.

MATERIALS AND METHODS: We retrospectively assessed the prospective clinical performance of a U.S. Food and Drug Administration-approved CAD system in second-reader mode in 347 consecutive adults (mean age, 57.6 years; 205 women, 142 men) undergoing CTC evaluation over a 5-month period. The reference standard consisted of the prospective interpretation by experienced CTC radiologists combined with subsequent optical colonoscopy (OC), if performed. We also assessed third-party reimbursement for CAD for studies performed over an 18-month period.

RESULTS: In all, 69 patients (mean [± SD] age, 59.0 ± 7.7 years; 32 men, 37 women) had 129 polyps ≥ 6 mm. Per-patient CAD sensitivity was 91.3% (63 of 69). Per-polyp CAD-alone sensitivity was 88.4% (114 of 129), including 88.3% (83 of 94) for 6- to 9-mm polyps and 88.6% (31 of 35) for polyps 10 mm or larger. On retrospective review, three additional polyps 6 mm or larger were seen at OC and marked by CAD but dismissed as CAD false-positives at CTC. The mean number of false-positive CAD marks was 4.4 ± 3.1 per series. Of 1225 CTC cases reviewed for reimbursement, 31.0% of the total charges for CAD interpretation had been recovered from a variety of third-party payers.

CONCLUSION: In our routine clinical practice, CAD showed good sensitivity for detecting colorectal polyps 6 mm or larger, with an acceptable number of false-positive marks. Importantly, CAD is already being reimbursed by some third-party payers in our clinical CTC practice.

来源:AJR Am J Roentgenol,查看原文。版权归原作者所有。

引用

  1. Ziemlewicz TJ, Kim DH, Hinshaw JL, et al. Computer-Aided Detection of Colorectal Polyps at CT Colonography: Prospective Clinical Performance and Third-Party Reimbursement. AJR Am J Roentgenol. 2017. PMID: 28753031. doi:10.2214/AJR.16.17499

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。