胎儿生物测量超声质量控制:INTERGROWTH-21st项目结果

Quality control of ultrasound for fetal biometry: results from the INTERGROWTH-21st Project.
PubMed 文献 · Ultrasound Obstet Gynecol · 2018-07-26

在INTERGROWTH-21st项目胎儿生长纵向研究中,对8个地理位置的4321例胎儿14-41周孕期的20313张超声图像进行前瞻性质量控制,包括定性视觉评估与定量测量变异分析。自评与外评一致性高(κ 0.96-0.99),观察者内与观察者间95%一致性界限分别达±3.3%-6.2%与±4.4%-6.0%,31名超声医师中3名三次测量SD分布显示变异过大并被及时识别和再培训。作者推荐将该QC系统用于未来研究和临床实践。

要点

  • 定性QC采用特定标准视觉评估图像,10%图像由牛津超声质量单位复评,调整κ达0.96-0.99。
  • 定量QC通过三次重复测量评估观察者内变异,10%重新放置游标评估观察者间变异。
  • 31名超声医师中3名三次测量SD分布显示变异过大,系统可及时识别并再培训。

核心结论

  • 定性与定量QC监测在大规模多中心研究中可行且高度可重复,有助于产出高质量超声图像;作者推荐将该QC系统应用于未来研究和临床实践。

研究设计与样本

  • INTERGROWTH-21st项目胎儿生长纵向研究中的前瞻性质量控制评估,纳入8个地理位置4321例胎儿14-41周孕期的20313张超声图像,对头围、腹围和股骨长度在分别生成的图像上盲法三次测量。

诊断标准与关键数值

  • 定性QC自评与外评一致性:头围κ=0.99(95%CI 0.98-0.99)、腹围κ=0.98(95%CI 0.97-0.99)、股骨长度κ=0.96(95%CI 0.95-0.98)。

临床应用与阅片要点

  • 作者推荐将所开发的QC系统应用于未来研究和临床实践。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To assess a comprehensive package of ultrasound quality control in the Fetal Growth Longitudinal Study of the INTERGROWTH-21st Project, a large multicenter study of fetal growth.

METHODS: Quality control (QC) measures were performed for 20 313 ultrasound scan images obtained prospectively from 4321 fetuses at 14-41 weeks' gestation in eight geographical locations. At the time of each ultrasound examination, three fetal biometric variables (head circumference (HC), abdominal circumference (AC) and femur length (FL)) were measured in triplicate on separately generated images. All measurements were taken in a blinded fashion. QC had two elements: (1) qualitative QC: visual assessment by sonographers at each study site of their images based on specific criteria, with 10% of images being re-assessed at the Oxford-based Ultrasound Quality Unit (compared using an adjusted kappa statistic); and (2) quantitative QC: assessment of measurement data by comparing the first, second and third measurements (intraobserver variability), remeasurement of caliper replacement in 10% (interobserver variability), both by Bland-Altman plots and plotting frequency histograms of the SD of triplicate measurements and assessing how many were above or below 2 SD of the expected distribution. The system allowed the sonographers' performances to be monitored regularly.

RESULTS: A high level of agreement between self- and external scoring was demonstrated for all measurements (κ = 0.99 (95% CI, 0.98-0.99) for HC, 0.98 (95% CI, 0.97-0.99) for AC and 0.96 (95% CI, 0.95-0.98) for FL). Intraobserver 95% limits of agreement (LoA) of ultrasound measures for HC, AC and FL were ± 3.3%, ± 5.6% and ± 6.2%, respectively; the corresponding values for interobserver LoA were ± 4.4%, ± 6.0% and ± 5.6%. The SD distribution of triplicate measurements for all biometric variables showed excessive variability for three of 31 sonographers, allowing prompt identification and retraining.

CONCLUSIONS: Qualitative and quantitative QC monitoring was feasible and highly reproducible in a large multicenter research study, which facilitated the production of high-quality ultrasound images. We recommend that the QC system we developed is implemented in future research studies and clinical practice. Copyright © 2017 ISUOG. Published by John Wiley & Sons Ltd.