荧光透视监视下气动肠套叠复位术中送气改良技术的评估

Assessment of a modified technique for air delivery during fluoroscopic-monitored pneumatic intussusception reduction.
PubMed 文献 · Pediatr Radiol · 2017-07-06

该研究对Shiels肠套叠复位装置进行改良,利用医院医用空气供应系统代替手动送气皮球送气,并回顾性比较改良装置与标准装置在18岁以下患者中的序贯荧光透视监视下气动肠套叠复位术。结果显示两种技术在手术时间、透视时间、辐射剂量及患者结局方面均无统计学显著差异,装置改良可提高操作者舒适度。

要点

  • 改良装置用医院医用空气供应系统替代手动送气皮球,省去术中持续按压皮球。
  • 改良与标准装置在手术时间、透视时间及辐射剂量方面无统计学显著差异。
  • 两组完全复位、不完全复位、3天内复发及后续手术例数均无统计学显著差异。

核心结论

  • 装置改良可提高操作者舒适度,所评估的手术参数和患者结局似乎与标准技术相似。

研究设计与样本

  • 回顾性比较序贯荧光透视监视下气动肠套叠复位术,患者年龄小于18岁,分别使用标准或改良装置,比较手术参数和患者结局。

局限性

  • 原文为回顾性比较序贯病例。
  • 原文结论使用「appear similar」表述,提示结果相似但非确定性结论。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Intussusception is one of the most common causes of acute abdominal emergency in children. Image-monitored pressure reduction is the first line of treatment.

OBJECTIVE: We report on a modified technique of air delivery during fluoroscopic-monitored pneumatic intussusception reductions, and compare it with an established technique.

MATERIALS AND METHODS: We modified the Shiels intussusception reduction device so that the air used for intussusception reduction is delivered not by the device's insufflator bulb, but rather by the hospital medical air supply system, eliminating the need for continuous pumping of the insufflator bulb during the procedure. Subsequently, we retrospectively compared sequential fluoroscopy-monitored pneumatic intussusception reduction procedures performed in patients younger than 18 years using either the standard or modified devices, evaluating technical procedure parameters (i.e. median procedure time length, fluoroscopy time length and radiation dose) and patient outcomes (i.e. number of complete intussusception reductions, number of incomplete intussusception reductions, number of intussusception recurrences within 3 days of the procedure, number of procedures followed by surgery).

RESULTS: We found no statistically significant differences between procedures performed with the standard and modified techniques.

CONCLUSION: The device modification allows for increased operator comfort. Evaluated procedure parameters and patient outcomes appear similar to those of the standard technique.

来源:Pediatr Radiol,查看原文。版权归原作者所有。