扫描时间适配的低管电压CT血管造影低容量对比剂注射方案:一项体外研究
本研究利用自制动态血流体模,探讨不同管电压下小容量对比剂(≤30mL)对动脉峰值强化和诊断性强化持续时间的影响,并优化长扫描时间CT血管造影的多团注方案。短扫描方案中,20mL对比剂在70和80kVp下、30mL在70、80和100kVp下可达诊断性峰值强化;长扫描方案中,优化多团注方案在100kVp下将T200较线性注射延长6秒(40%)。结论认为,优化CTA方案可同时节省辐射剂量和对比剂容量。
要点
- 20mL对比剂在70和80kVp下即可达到诊断性峰值强化。
- 长扫描时间下多团注方案较线性注射使T200延长6秒(40%)。
- 优化CTA方案可同时节省辐射剂量和对比剂容量。
核心结论
- 优化CTA方案结合管电压和对比剂注射方案的交替变化,可同时节省辐射剂量和对比剂容量。
研究设计与样本
- 体外研究,采用自制动态血流体模;短扫描时间方案比较70-120kVp下20和30mL碘比醇350mg I/mL、流率5mL/s的动脉强化曲线;长扫描时间方案采用13个「微团注」对比剂(总量60mL)与生理盐水(总量24mL)快速切换、流率4mL/s的优化多团注技术。
诊断标准与关键数值
- 短扫描时间方案:20mL对比剂在70和80kVp下峰值动脉强化分别为327±10和255±15 HU;30mL对比剂在70、80和100kVp下峰值动脉强化分别为451±10、367±9和253±15 HU;长扫描时间方案:优化多团注方案在100kVp下T200为21±1s,线性注射方案为15±1s(p<0.001),优化方案延长6s(40%)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: The aims of this study were twofold. First, we investigated the extent of changes in arterial peak enhancement and changes in the duration of a diagnostic arterial enhancement when small amounts of CA volumes (≤30mL) were administered at varying tube voltages. Second, we investigated how to optimize CA injection protocols for CT-angiography with long scan times at various tube voltages to achieve optimal vascular enhancement at the lowest reasonable CA dose.
MATERIALS AND METHODS: Measurements were performed with a custom-made dynamic flow phantom. For CTA protocols with a short scan time, we investigated the effect of various tube voltages (70-120kVp) on the arterial enhancement profile with very small CA volumes (20 and 30mL of Iobitridol 350mg I/mL) at a flow rate of 5mL/s. For CTA protocols with a long scan time, we utilized an optimized multi-bolus technique switching rapidly between 13 "micro-boli" of CA (total, 60mL) and saline (total, 24mL) at a flow rate of 4mL/s. The peak arterial enhancement (PAE) and the time period of diagnostic aortic enhancement ≥200 HU (T200) were analyzed.
RESULTS: For the short scan time protocols, a diagnostic peak enhancement was achieved using 20mL of CA at 70 and 80kVp (PAE: 327±10 and 255±15 HU, respectively) or 30mL of CA at 70, 80 and 100kVp (PAE 451±10, 367±9, and 253±15 HU). For the long scan time, the optimized multi-bolus injection protocol extended T200 at 100kVp by 6s (40%) compared to a linear injection protocol (21±1s and 15±1s, respectively; p<0.001).
CONCLUSION: Optimized CTA protocols comprising alternations of tube voltage and the CA injection protocol can save radiation doses and CA volumes at the same time.
引用
- Benz MR, Szucs-Farkas Z, Froehlich JM, et al. Scan time adapted contrast agent injection protocols with low volume for low-tube voltage CT angiography: An in vitro study. Eur J Radiol. 2017. PMID: 28668433. doi:10.1016/j.ejrad.2017.05.017