文献速递心血管外周血管CT

计算机断层扫描血管成像中主动脉瓣环的定量:全自动方法的验证

Quantification of aortic annulus in computed tomography angiography: Validation of a fully automatic methodology.
PubMed 文献 · Eur J Radiol · 2017-04-26

本研究回顾性分析26例患者的CTA数据集,提出基于知识分割主动脉根部和检测瓣环方向与大小的全自动方法,并与半自动勾画和手动定义瓣环平面的参考标准比较。自动瓣环直径与观察者1的差值为0.2±1.0mm,观察者间变异性为1.2±0.6mm,直径的Pearson相关系数良好(观察者1为0.92);首次提出并验证全自动评估最佳投影曲线工具,基于自动瓣环平面计算的最佳投影曲线与参考标准在头/尾方向平均差6.4°,每数据集平均计算时间约60秒。

要点

  • 全自动方法测量主动脉瓣环直径与观察者1的差值为0.2±1.0mm。
  • 观察者间变异性为1.2±0.6mm,直径Pearson相关系数观察者1为0.92。
  • 自动瓣环平面与参考标准的最佳投影曲线头/尾方向平均差6.4°。

核心结论

  • 新型全自动快速方法不仅提供与有经验观察者相当的主动脉瓣环尺寸精确测量,还能从CTA图像预测最佳X射线投影曲线。

研究设计与样本

  • 回顾性分析26例患者的CTA数据集,以半自动勾画主动脉根部并手动定义瓣环平面为参考标准。

诊断标准与关键数值

  • 自动瓣环直径与观察者1参考标准的差值为0.2±1.0mm。
  • 观察者间变异性为1.2±0.6mm。
  • 直径的Pearson相关系数观察者1为0.92。
  • 基于自动瓣环平面计算的最佳投影曲线与参考标准在头/尾方向平均差6.4°。
  • 每数据集平均计算时间约60秒。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Automatic accurate measuring of the aortic annulus and determination of the optimal angulation of X-ray projection are important for the trans-catheter aortic valve replacement (TAVR) procedure. The objective of this study was to present a novel fully automatic methodology for the quantification of the aortic annulus in computed tomography angiography (CTA) images.

METHODS: CTA datasets of 26 patients were analyzed retrospectively with the proposed methodology, which consists of a knowledge-based segmentation of the aortic root and detection of the orientation and size of the aortic annulus. The accuracy of the methodology was determined by comparing the automatically derived results with the reference standard obtained by semi-automatic delineation of the aortic root and manual definition of the annulus plane.

RESULTS: The difference between the automatic annulus diameter and the reference standard by observer 1 was 0.2±1.0mm, with an inter-observer variability of 1.2±0.6mm. The Pearson correlation coefficient for the diameter was good (0.92 for observer 1). For the first time, a fully automatic tool to assess the optimal projection curves was presented and validated. The mean difference between the optimal projection curves calculated based on the automatically defined annulus plane and the reference standard was 6.4° in the cranial/caudal (CRA/CAU) direction. The mean computation time was short with around 60s per dataset.

CONCLUSION: The new fully automatic and fast methodology described in this manuscript not only provided precise measurements about the aortic annulus size with results comparable to experienced observers, but also predicted optimal X-ray projection curves from CTA images.

来源:Eur J Radiol,查看原文。版权归原作者所有。