冠状动脉CT血管造影评估生物可吸收血管支架的诊断准确性

Diagnostic Accuracy of Coronary CT Angiography for the Evaluation of Bioresorbable Vascular Scaffolds.
PubMed 文献 · JACC Cardiovasc Imaging · 2017-07-19

在ABSORB II研究中,238例患者(258处病变)接受生物可吸收血管支架治疗,3年随访时行冠状动脉造影、血管内超声和冠状动脉CT血管造影。冠状动脉CT血管造影检测支架内狭窄的曲线下面积为0.88(以冠状动脉造影直径狭窄≥50%为参考),以血管内超声最小管腔面积≤2.5 mm²为参考时曲线下面积为0.83。冠状动脉CT血管造影与冠状动脉造影识别显著病变的诊断准确性相似(p=0.75)。

要点

  • 冠状动脉CT血管造影检测生物可吸收血管支架内狭窄的曲线下面积为0.88,敏感度80%、特异度100%(以冠状动脉造影为参考)
  • 以血管内超声最小管腔面积≤2.5 mm²为参考时,曲线下面积为0.83,敏感度71%、特异度82%
  • 冠状动脉CT血管造影与冠状动脉造影识别显著病变的诊断准确性相似(p=0.75)

核心结论

  • 冠状动脉CT血管造影对检测生物可吸收血管支架置入后支架内管腔阻塞及评估管腔尺寸具有良好诊断准确性,可能成为评估此类患者的首选方法。

研究设计与样本

  • ABSORB II研究为随机对照试验,患者随机接受生物可吸收血管支架或依维莫司洗脱金属支架治疗;3年随访时238例患者(258处病变)接受生物可吸收血管支架治疗并完成冠状动脉造影、血管内超声和冠状动脉CT血管造影检查,以冠状动脉造影和血管内超声为参考标准评估冠状动脉CT血管造影的诊断准确性。

关键影像表现/征象

  • 冠状动脉CT血管造影与定量冠状动脉造影相比,最小管腔直径平均差值为-0.14 mm(一致性界限-0.88至0.60 mm)
  • 冠状动脉CT血管造影与血管内超声相比,最小管腔面积平均差值为0.73 mm²(一致性界限-1.85至3.30 mm²)

诊断标准与关键数值

  • 以冠状动脉造影直径狭窄≥50%为参考,冠状动脉CT血管造影检测支架内狭窄的曲线下面积为0.88(95%CI:0.82至0.92),敏感度80%(95%CI:28%至99%),特异度100%(95%CI:98%至100%)
  • 以血管内超声最小管腔面积≤2.5 mm²为参考,冠状动脉CT血管造影检测支架内狭窄的曲线下面积为0.83(95%CI:0.77至0.88),敏感度71%(95%CI:44%至90%),特异度82%(95%CI:75%至87%)
  • 冠状动脉CT血管造影与冠状动脉造影识别显著病变的诊断准确性相似(p=0.75)

临床应用与阅片要点

  • 冠状动脉CT血管造影可用于评估生物可吸收血管支架置入后的支架内管腔阻塞和管腔尺寸
  • 作者认为冠状动脉CT血管造影可能成为评估生物可吸收血管支架治疗患者的首选方法

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The purpose of this study was to assess the diagnostic accuracy of coronary computed tomography angiography (CTA) for bioresorbable vascular scaffold (BVS) evaluation.

BACKGROUND: Coronary CTA has emerged as a noninvasive method to evaluate patients with suspected or established coronary artery disease. The diagnostic accuracy of coronary CTA to evaluate angiographic outcomes after BVS implantation has not been well established.

METHODS: In the ABSORB II (A Bioresorbable Everolimus-Eluting Scaffold Versus a Metallic Everolimus-Eluting Stent II) study, patients were randomized either to receive treatment with the BVS or everolimus-eluting metallic stent. At the 3-year follow-up, 238 patients (258 lesions) treated with BVS underwent coronary angiography with intravascular ultrasound (IVUS) evaluation and coronary CTA. The diagnostic accuracy of coronary CTA was assessed by the area under the receiver-operating characteristic curve with coronary angiography and IVUS as references.

RESULTS: The mean difference in coronary CTA-derived minimal luminal diameter was -0.14 mm (limits of agreement -0.88 to 0.60) with quantitative coronary angiography as reference, whereas the mean difference in minimal lumen area was 0.73 mm2 (limits of agreement -1.85 to 3.30) with IVUS as reference. The per-scaffold diagnostic accuracy of coronary CTA for detecting stenosis based on coronary angiography diameter stenosis of ≥50% revealed an area under the receiver-operating characteristic curve of 0.88 (95% confidence interval [CI]: 0.82 to 0.92) with a sensitivity of 80% (95% CI: 28% to 99%) and a specificity of 100% (95% CI: 98% to 100%), whereas diagnostic accuracy based on IVUS minimal lumen area ≤2.5 mm2 showed an area under the receiver-operating characteristic curve of 0.83 (95% CI: 0.77 to 0.88) with a sensitivity of 71% (95% CI: 44% to 90%) and a specificity of 82% (95% CI: 75% to 87%). The diagnostic accuracy of coronary CTA was similar to coronary angiography in its ability to identify patients with a significant lesion based on the IVUS criteria (p = 0.75).

CONCLUSIONS: Coronary CTA has good diagnostic accuracy to detect in-scaffold luminal obstruction and to assess luminal dimensions after BVS implantation. Coronary angiography and coronary CTA yielded similar diagnostic accuracy to identify the presence and severity of obstructive disease. Coronary CTA might become the method of choice for the evaluation of patients treated with BVS.

来源:JACC Cardiovasc Imaging,查看原文。版权归原作者所有。