腹主动脉瘤腔内修复术后内漏:噪声优化虚拟单能双能CT提高检出率
本研究纳入75例腹主动脉瘤腔内修复术后患者,行动脉期双能CT血管成像,比较标准线性融合M_0.5、噪声优化虚拟单能图像(VMI+)及传统虚拟单能图像(VMI)40-100keV的图像质量与内漏诊断效能。34例存在内漏;40keV VMI+的对比噪声比最高(21.3±11.1),内漏检出AUC达0.992,显著高于70keV VMI(0.914)和M_0.5(0.916)。结论认为40keV噪声优化VMI+可提高内漏检出和排除的诊断准确率。
要点
- 40keV噪声优化VMI+的对比噪声比最高(21.3±11.1),优于M_0.5(10.9±5.5)及传统VMI各能级。
- 40keV VMI+内漏检出AUC为0.992,显著高于70keV VMI(0.914)和M_0.5(0.916)。
- 75例中34例存在内漏,提示EVAR术后随访中内漏并不少见。
核心结论
- 40keV噪声优化VMI+系列可提高EVAR术后内漏检出和排除的诊断准确率。
研究设计与样本
- 75例(男42例,年龄66.2±11.7岁)EVAR术后患者行双能CT血管成像,动脉期双能模式采集,重建标准线性融合M_0.5、VMI+及传统VMI(40-100keV,每10keV间隔);由三名盲法放射科医师采用各重建技术中客观最佳系列评估内漏检出诊断准确率。
关键影像表现/征象
- 内漏患者中,40keV VMI+的对比噪声比最高(21.3±11.1)。
- M_0.5的对比噪声比为10.9±5.5。
- 传统VMI系列在70keV时对比噪声比最高(13.5±6.6)。
诊断标准与关键数值
- 内漏检出ROC分析:40keV VMI+的AUC为0.992。
- 70keV VMI的AUC为0.914,M_0.5的AUC为0.916,均显著低于40keV VMI+(P≤0.039)。
- 40keV VMI+对比噪声比21.3±11.1 vs M_0.5 10.9±5.5及传统VMI最高值70keV 13.5±6.6,均P<0.001。
临床应用与阅片要点
- EVAR术后内漏检出可优先采用40keV噪声优化VMI+系列,以提高对比噪声比和诊断准确率。
- 三名盲法医师采用各重建技术中客观最佳系列评估,提示日常阅片可结合客观图像质量参数选择重建系列。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To assess image quality and diagnostic performance of a noise-optimized algorithm to reconstruct virtual monoenergetic images (VMI+) for the detection of endoleaks after endovascular abdominal aortic aneurysm repair (EVAR) using dual-energy CT angiography (DE-CTA).
MATERIALS AND METHODS: Seventy-five patients (42 men; 66.2±11.7years) underwent DE-CTA following EVAR. Arterial phase images were acquired in dual-energy mode for the reconstruction of standard linearly-blended M_0.5, VMI+ and traditional monoenergetic images (VMI) at 40-100keV in 10-keV intervals. Contrast-to-noise ratios (CNR) were calculated for the area of leakage in patients with endoleaks. Diagnostic accuracy for endoleak detection was evaluated by three blinded radiologists using the objectively best series for each reconstruction technique.
RESULTS: Thirty-four out of 75 patients showed endoleaks. Quantitative image parameters were highest at 40-keV VMI+ (CNR, 21.3±11.1), compared to M_0.5 (CNR, 10.9±5.5) and all VMI series that showed highest values at 70keV (CNR, 13.5±6.6; all P<0.001). ROC analysis for endoleak detection revealed an area under the curve (AUC) of 0.992 for 40-keV VMI+ series, which was significantly higher (P≤0.039) compared to 70-keV VMI (0.914) and M_0.5 series (0.916).
CONCLUSIONS: Noise-optimized VMI+ series at 40keV improve diagnostic accuracy for the detection and rule-out of endoleaks after EVAR.