冠状动脉钙化积分中的剂量降低技术:迭代重建联合低管电压对胸部体模钙化积分的影响

Dose reduction techniques in coronary calcium scoring: The effect of iterative reconstruction combined with low tube voltage on calcium scores in a thoracic phantom.
PubMed 文献 · Eur J Radiol · 2017-06-04

在含100个钙化灶的胸部体模上,以120kVp、90参考mAs、FBP为参考方案,测试30种低剂量方案(70-110kVp,FBP及迭代重建1-5级)。90kVp-IR3和100kVp-IR1方案钙化检出率和Agatston积分与参考方案相似,剂量分别降低60.6%和43.6%。

要点

  • 90kVp联合迭代重建3级与参考120kVp方案钙化检出率和Agatston积分相似。
  • 100kVp联合迭代重建1级同样与参考方案检出率和积分相似。
  • 迭代重建级别和管电压升高均使钙化检出率与积分下降。

核心结论

  • 90kVp-IR3和100kVp-IR1方案可获得与参考120kVp方案相似的钙化检出率、Agatston积分和图像质量,剂量降低最高达60.6%。

研究设计与样本

  • 体模研究,使用含100个钙化灶的胸部体模,在第三代双源CT上扫描参考方案(120kVp、90参考mAs、FBP)和30种低剂量方案(70-110kVp、90参考mAs、FBP及迭代重建1-5级),并对与参考方案检出率和钙化积分相似的方案追加降低参考mAs的扫描。

诊断标准与关键数值

  • 90kVp-IR3方案与参考方案相比,钙化检出率和钙化积分差异无统计学意义(p>0.206),剂量降低60.6%。
  • 100kVp-IR1方案与参考方案相比,钙化检出率和钙化积分差异无统计学意义(p>0.206),剂量降低43.6%。
  • 在90kVp-IR3和100kVp-IR1方案中,降低管电流不影响检出率和Agatston积分(p>0.206),CNR和SNR与参考方案相似或更高(0.008<p<0.206)。
  • 随着迭代重建级别升高(τb<-0.825,p<0.001)和管电压升高(τb<-0.679,p<0.001),钙化检出率和钙化积分下降。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To define a dose-reduced coronary calcium CT protocol that provides similar calcium score values as a conventional 120kVp protocol.

METHODS: A thorax phantom containing 100 calcifications was scanned with the reference protocol (120kVp, 90 ref mAs, FBP) and 30 dose-reduced protocols (70-110kVp, 90 ref mAs, FBP and iterative reconstruction (IR) levels 1-5) with 3rd generation dual-source CT. For protocols that yielded similar detectability and calcium scores as the reference protocol, additional scans were acquired at reduced ref mAs. Kendall's τb and independent-samples-median test were used to determine trends and differences in contrast/signal-to-noise ratio (CNR and SNR), dose and calcium scores.

RESULTS: The detectability and calcium scores decreased at increasing IR levels (τb<-0.825, p<0.001) and increasing tube voltage (τb<-0.679, p<0.001). For 90kVp-IR3 and 100kVp-IR1, similar detectability and calcium score was found compared to the reference protocol (p>0.206). For these protocols, lower tube currents did not affect the detectability and Agatston score (p>0.206), while CNR and SNR were similar/higher compared to the reference protocol (0.008<p<0.206). Dose reduction was 60.6% (90kVp-IR3) and 43.6% (100kVp-IR1).

CONCLUSIONS: The protocol of 90kVp-IR3 and 100kVp-IR1 yielded similar calcium detectability, Agatston score and image quality as the reference protocol, with dose reduction up to 60.6%.

来源:Eur J Radiol,查看原文。版权归原作者所有。