文献速递胸部CT

光子计数探测器低剂量胸部CT的可行性:人体初步结果

Feasibility of Dose-reduced Chest CT with Photon-counting Detectors: Initial Results in Humans.
PubMed 文献 · Radiology · 2017-07-28

本研究在30名无症状志愿者中比较了低剂量光子计数探测器(PCD)与常规能量积分探测器(EID)胸部CT的图像质量。PCD的诊断为量更高,图像噪声更低,肺结节对比噪声比更高(12.1±1.7 vs 10.0±1.8),且线束硬化伪影更少。

要点

  • 低剂量PCD胸部CT的图像噪声低于EID,肺结节对比噪声比更高(12.1±1.7 vs 10.0±1.8)。
  • PCD在肺、软组织、骨及线束硬化伪影方面图像质量评分均优于EID(P < .001)。
  • 三位阅片者均认为PCD的诊断质量高于EID(P = .016、.016、.013)。

核心结论

  • 低剂量PCD胸部CT的初步人体经验显示,与常规EID CT相比,图像噪声更低,诊断质量和肺结节对比噪声比更好。

研究设计与样本

  • 经机构审查委员会批准、符合HIPAA的研究,30名无症状志愿者行低剂量螺旋平扫肺部EID和PCD CT,采用制造商推荐的肺癌筛查指南(管电压120 kVp,参考管电流-时间乘积20 mAs)。

关键影像表现/征象

  • PCD图像在肺、软组织和骨方面的图像质量评分显著优于EID,且线束硬化伪影更少(均P < .001)。
  • 所有BMI组的PCD图像噪声均显著低于EID(第1组和第3组P < .001,第2组P < .01)。
  • 肺结节对比噪声比PCD为12.1±1.7,EID为10.0±1.8(P < .001)。

诊断标准与关键数值

  • PCD诊断质量高于EID,阅片者1、2、3的P值分别为.016、.016和.013。
  • 阅片者间和阅片者内重复性良好(所有ICC > 0.800)。

临床应用与阅片要点

  • 低剂量PCD胸部CT在无症状志愿者中显示图像噪声低于EID,可能有助于改善低剂量肺癌筛查的图像质量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Purpose To investigate whether photon-counting detector (PCD) technology can improve dose-reduced chest computed tomography (CT) image quality compared with that attained with conventional energy-integrating detector (EID) technology in vivo. Materials and Methods This was a HIPAA-compliant institutional review board-approved study, with informed consent from patients. Dose-reduced spiral unenhanced lung EID and PCD CT examinations were performed in 30 asymptomatic volunteers in accordance with manufacturer-recommended guidelines for CT lung cancer screening (120-kVp tube voltage, 20-mAs reference tube current-time product for both detectors). Quantitative analysis of images included measurement of mean attenuation, noise power spectrum (NPS), and lung nodule contrast-to-noise ratio (CNR). Images were qualitatively analyzed by three radiologists blinded to detector type. Reproducibility was assessed with the intraclass correlation coefficient (ICC). McNemar, paired t, and Wilcoxon signed-rank tests were used to compare image quality. Results Thirty study subjects were evaluated (mean age, 55.0 years ± 8.7 [standard deviation]; 14 men). Of these patients, 10 had a normal body mass index (BMI) (BMI range, 18.5-24.9 kg/m2; group 1), 10 were overweight (BMI range, 25.0-29.9 kg/m2; group 2), and 10 were obese (BMI ≥30.0 kg/m2, group 3). PCD diagnostic quality was higher than EID diagnostic quality (P = .016, P = .016, and P = .013 for readers 1, 2, and 3, respectively), with significantly better NPS and image quality scores for lung, soft tissue, and bone and with fewer beam-hardening artifacts (all P < .001). Image noise was significantly lower for PCD images in all BMI groups (P < .001 for groups 1 and 3, P < .01 for group 2), with higher CNR for lung nodule detection (12.1 ± 1.7 vs 10.0 ± 1.8, P < .001). Inter- and intrareader reproducibility were good (all ICC > 0.800). Conclusion Initial human experience with dose-reduced PCD chest CT demonstrated lower image noise compared with conventional EID CT, with better diagnostic quality and lung nodule CNR. © RSNA, 2017 Online supplemental material is available for this article.