优化超低剂量脉冲透视用于上消化道检查的可行性:体模研究及临床相关性

Feasibility of optimized ultralow-dose pulsed fluoroscopy for upper gastrointestinal tract examinations: a phantom study with clinical correlation.
PubMed 文献 · Radiol Med · 2017-07-22

本研究在体模上建立上消化道超低剂量数字脉冲透视方案(86.7 kV、77 mA、0.9 mm3、自动噪声与对比度适应),并测量剂量面积乘积与器官剂量;随后在41例患者中由两名放射科医师共识评估图像质量。与常规数字X线摄影方案相比,透视方案每幅图像剂量降低约96%,剂量面积乘积降低97%,器官剂量亦显著降低,且所有患者图像质量均满足诊断要求。

要点

  • 超低剂量脉冲透视方案每幅图像剂量较常规数字X线摄影降低约96%,剂量面积乘积降低97%。
  • 女性性腺剂量FP 2.4×10-3 vs DR 0.05,男性FP 8×10-4 vs DR 0.03(p≤0.0004)。
  • 所有患者图像质量均可满足诊断,但各评估参数评分均显著低于常规数字X线摄影(p<0.05)。

核心结论

  • 超低剂量脉冲透视可用于临床常规,在显著降低直接和散射剂量暴露的同时提供足够诊断图像质量。

研究设计与样本

  • 体模研究联合临床评估:采用Alderson-Rando体模和60个热释光剂量计测量剂量面积乘积与器官剂量。
  • 临床部分纳入41例接受对比增强脉冲透视的患者,由两名放射科医师共识评估图像质量。
  • 以常规数字X线摄影采集方案作为辐射剂量和图像质量分析的参考标准。

关键影像表现/征象

  • 体模测量显示脉冲透视方案较数字X线摄影标准每幅图像剂量降低约96%。
  • 脉冲透视方案剂量面积乘积可降低97%。
  • 直接与散射束器官剂量均显著降低,眶部剂量FP 5.6×10-3 vs DR 0.11(p=0.02)。
  • 所有患者脉冲透视均提供诊断级图像质量,但各评估参数评分显著低于数字X线摄影标准(p<0.05)。

诊断标准与关键数值

  • 主观图像质量、对比剂显示和图像噪声采用5分Likert量表评估(5=极好)。
  • 优化脉冲透视采集参数:86.7 kV、77 mA、0.9 mm3、自动图像噪声与对比度适应。

临床应用与阅片要点

  • 超低剂量脉冲透视方案可用于上消化道检查的临床常规,兼顾剂量降低与诊断图像质量。
  • 阅片时需注意脉冲透视图像各主观质量参数评分显著低于常规数字X线摄影,但所有患者仍满足诊断要求。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To establish an optimized ultralow-dose digital pulsed fluoroscopy (FP) protocol for upper gastrointestinal tract examinations and to investigate the radiation dose and image quality.

MATERIALS AND METHODS: An Alderson-Rando-Phantom with 60 thermoluminescent dosimeters was used for dose measurements to systematically evaluate the dose-area product (DAP) and organ doses of the optimized FP protocol with the following acquisition parameters: 86.7 kV; 77 mA; 0.9 mm3, automatic image noise and contrast adaption. Subjective image quality, depiction of contrast agent and image noise (5-point Likert scale; 5 = excellent) were assessed in 41 patients, who underwent contrast-enhanced FP with the aforementioned optimized protocol by two radiologists in consensus. A conventional digital radiograph (DR) acquisition protocol served as the reference standard for radiation dose and image quality analyses.

RESULTS: Phantom measurements revealed a general dose reduction of approximately 96% per image for the FP protocol as compared to the DR standard. DAP could be reduced by 97%. Significant dose reductions were also found for organ doses, both in the direct and scattered radiation beam with negligible orbital (FP 5.6 × 10-3 vs. DR 0.11; p = 0.02) and gonadal dose exposure (female FP 2.4 × 10-3 vs. DR 0.05; male FP 8 × 10-4 vs. DR 0.03; p ≤ 0.0004). FP provided diagnostic image quality in all patients, although reading scores were significantly lower for all evaluated parameters as compared to the DR standard (p < 0.05).

CONCLUSION: Ultralow-dose FP is feasible for clinical routine allowing a significant reduction of direct and scattered dose exposure while providing sufficient diagnostic image quality for reliable diagnosis.

来源:Radiol Med,查看原文。版权归原作者所有。