改良运动敏感驱动平衡准备对比增强三维涡轮自旋回波成像的最大强度投影在脑转移瘤检测中的效能

Efficacy of Maximum Intensity Projection of Contrast-Enhanced 3D Turbo-Spin Echo Imaging with Improved Motion-Sensitized Driven-Equilibrium Preparation in the Detection of Brain Metastases.
PubMed 文献 · Korean J Radiol · 2017-05-19

研究评估5 mm最大强度投影的改良运动敏感驱动平衡准备对比增强3D T1加权涡轮自旋回波成像(MIP iMSDE-TSE)检测脑转移瘤的诊断价值,并与1 mm iMSDE-TSE、1 mm对比增强3D T1加权梯度回波成像(非MIP 3D-GRE)及5 mm MIP 3D-GRE比较。30例患者共460个强化脑转移瘤(>3 mm者150个,≤3 mm者310个)纳入分析,两名神经放射科医师独立阅片。MIP iMSDE-TSE的检出效能与非MIP iMSDE-TSE相当,但优于非MIP及MIP 3D-GRE,且阅读时间更短。

要点

  • MIP iMSDE-TSE对全部病灶及≤3 mm病灶的敏感度与FOM均与非MIP iMSDE-TSE相当,且优于非MIP/MIP 3D-GRE。
  • MIP iMSDE-TSE阅读时间显著短于非MIP iMSDE-TSE(读者1:47.7±35.9秒 vs 78.8±43.7秒,p=0.01;读者2:44.7±23.6秒 vs 82.9±39.9秒,p<0.001)。
  • MIP iMSDE-TSE的假阳性结果较多,作者建议与1 mm序列联合使用以获得高诊断效能和低假阳性率。

核心结论

  • MIP iMSDE-TSE对脑转移瘤的检出效能高,与非MIP iMSDE-TSE相当,但优于非MIP/MIP 3D-GRE;其阅读时间更短,但假阳性结果较多,作者建议与1 mm序列联合使用可提供高诊断效能和低假阳性率。

研究设计与样本

  • 研究纳入2014年10月至2015年7月间30例患者共460个强化脑转移瘤(>3 mm者150个,≤3 mm者310个),行非MIP iMSDE-TSE和非MIP 3D-GRE扫描,并进行5 mm MIP重建,由两名神经放射科医师独立阅片比较四种序列。

诊断标准与关键数值

  • MIP iMSDE-TSE平均FOM:全部病灶0.984,≤3 mm病灶0.980;敏感度:读者1全部病灶97.3%、≤3 mm病灶96.2%,读者2全部病灶97.0%、≤3 mm病灶95.8%。
  • 非MIP iMSDE-TSE平均FOM:全部病灶0.985,≤3 mm病灶0.977;敏感度:读者1全部病灶96.7%、≤3 mm病灶99.0%,读者2全部病灶97%、≤3 mm病灶95.3%。
  • MIP iMSDE-TSE的FOM和敏感度优于非MIP和MIP 3D-GRE(均p<0.001)。
  • 两种序列对所有病灶的观察者间一致性均极好(κ>0.75)。

临床应用与阅片要点

  • 作者建议MIP iMSDE-TSE与1 mm序列联合使用,可提供高诊断效能和低假阳性率。

局限性

  • MIP iMSDE-TSE的假阳性结果较多。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To evaluate the diagnostic benefits of 5-mm maximum intensity projection of improved motion-sensitized driven-equilibrium prepared contrast-enhanced 3D T1-weighted turbo-spin echo imaging (MIP iMSDE-TSE) in the detection of brain metastases. The imaging technique was compared with 1-mm images of iMSDE-TSE (non-MIP iMSDE-TSE), 1-mm contrast-enhanced 3D T1-weighted gradient-echo imaging (non-MIP 3D-GRE), and 5-mm MIP 3D-GRE.

MATERIALS AND METHODS: From October 2014 to July 2015, 30 patients with 460 enhancing brain metastases (size > 3 mm, n = 150; size ≤ 3 mm, n = 310) were scanned with non-MIP iMSDE-TSE and non-MIP 3D-GRE. We then performed 5-mm MIP reconstruction of these images. Two independent neuroradiologists reviewed these four sequences. Their diagnostic performance was compared using the following parameters: sensitivity, reading time, and figure of merit (FOM) derived by jackknife alternative free-response receiver operating characteristic analysis. Interobserver agreement was also tested.

RESULTS: The mean FOM (all lesions, 0.984; lesions ≤ 3 mm, 0.980) and sensitivity ([reader 1: all lesions, 97.3%; lesions ≤ 3 mm, 96.2%], [reader 2: all lesions, 97.0%; lesions ≤ 3 mm, 95.8%]) of MIP iMSDE-TSE was comparable to the mean FOM (0.985, 0.977) and sensitivity ([reader 1: 96.7, 99.0%], [reader 2: 97, 95.3%]) of non-MIP iMSDE-TSE, but they were superior to those of non-MIP and MIP 3D-GREs (all, p < 0.001). The reading time of MIP iMSDE-TSE (reader 1: 47.7 ± 35.9 seconds; reader 2: 44.7 ± 23.6 seconds) was significantly shorter than that of non-MIP iMSDE-TSE (reader 1: 78.8 ± 43.7 seconds, p = 0.01; reader 2: 82.9 ± 39.9 seconds, p < 0.001). Interobserver agreement was excellent (κ > 0.75) for all lesions in both sequences.

CONCLUSION: MIP iMSDE-TSE showed high detectability of brain metastases. Its detectability was comparable to that of non-MIP iMSDE-TSE, but it was superior to the detectability of non-MIP/MIP 3D-GREs. With a shorter reading time, the false-positive results of MIP iMSDE-TSE were greater. We suggest that MIP iMSDE-TSE can provide high diagnostic performance and low false-positive rates when combined with 1-mm sequences.

来源:Korean J Radiol,查看原文。版权归原作者所有。