四维引导三维彩色多普勒超声定量颈动脉狭窄具有高可重复性与准确性
前瞻性纳入93例患者122处颈内动脉狭窄,最终80例103处狭窄纳入分析,比较四维实时灰阶引导三维彩色多普勒超声(4D/3DC-US)与导管血管造影(CA)及双功能超声(DUS)对狭窄分级的定量的二分类准确性。4D/3DC-US与CA的一致性相关系数为0.89,检测50%和70%狭窄的敏感度分别为97%和87%,特异度分别为92%和84%,50%狭窄检测准确率4D/3DC-US为96% vs DUS 91%。
要点
- 4D/3DC-US与CA定量颈内动脉狭窄的一致性相关系数为0.89,差值标准差8.1%,偏倚+1.7%。
- 4D/3DC-US检测50%和70%狭窄的敏感度分别为97%和87%,特异度分别为92%和84%。
- 50%狭窄检测准确率4D/3DC-US为96%,DUS为91%,4D/3DC-US呈趋势性更高。
核心结论
- 4D/3DC-US方法可提供可靠且准确的狭窄定量和二分类,与CA和DUS相比具有良好的诊断准确性。
研究设计与样本
- 前瞻性研究,93例患者122处颈内动脉狭窄行4D/3DC-US筛查,排除图像质量不足、既往血运重建及CA禁忌者后,80例患者103处狭窄纳入分析;超声检查(10 MHz)包括共识一致的DUS检查和独立实时4D引导灰阶视图定位,随后离线多平面重建进行静态3DC-US NASCET狭窄百分比定量;同批狭窄以专用软件盲法行多平面选择性CA定量。
诊断标准与关键数值
- 以CA为参考,103处狭窄平均狭窄程度65±17%,4D/3DC-US与CA的一致性相关系数0.89,差值标准差8.1%,偏倚+1.7%;4D/3DC-US检测50%狭窄的敏感度97%、特异度92%,检测70%狭窄的敏感度87%、特异度84%;50%狭窄检测准确率4D/3DC-US为96% vs DUS 91%。
临床应用与阅片要点
- 4D/3DC-US可用于颈内动脉狭窄的定量评估和二分类(50%和70%阈值),与CA和DUS相比诊断准确性良好。
- 52处狭窄CA的观察者间差值标准差为7.2%,与4D/3DC-US和CA之间的差值标准差相比无显著差异(p=0.274)。
局限性
- 研究为前瞻性设计,但排除了图像质量不足、既往血运重建及CA禁忌的患者,最终仅80例患者103处狭窄纳入分析。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The purpose was to analyze the agreement and binary accuracy of the degree of internal carotid artery stenosis (ICAS) as determined by 4-dimensionally (4D) real-time gray-scale guided 3-dimensional (3D) color-Doppler ultrasonography (3DC-US) (4D/3DC-US) compared with catheter angiography (CA) and duplex ultrasonography (DUS). This study hypothesized that 4D/3DC-US is noninferior to CA and DUS in grading ICAS in selected patients.
BACKGROUND: Clinical stratification in patients with ICAS largely depends on a patient's symptomatic status and the degree of stenosis.
METHODS: Screening with 4D/3DC-US was prospectively performed in 93 study patients (with 122 ICASs), thus yielding 80 patients for analysis (with 103 ICASs) after excluding patients with insufficient image quality, previous revascularization, and contraindications to CA. The ultrasound examination (10 MHz) consisted of consensus conform DUS examination and independent real-time 4D-guided gray-scale views for orientation followed by static 3DC-US NASCET (North American Symptomatic Carotid Endarterectomy Trial) percent stenosis quantification using off-line multiplanar rendering. Multiplanar selective CA of the same ICASs was quantified with dedicated software in a blinded fashion.
RESULTS: Quantitative CA of 103 stenoses with a mean degree of 65 ± 17% was compared with 4D/3DC-US, with a resulting concordance correlation coefficient of 0.89 and a standard deviation of differences (SDD) of 8.1% at a bias of +1.7%. Binary 50% and 70% stenosis detection with 4D/3DC-US revealed a sensitivity of 97% and 87%, respectively, and a specificity of 92% and 84%, respectively. Interobserver SDD for CA of 52 stenoses (7.2%) did not differ from SDD for 4D/3DC-US and CA (p = 0.274). Accuracy of 50% stenosis detection by 4D/3DC-US was tendentially higher compared with DUS (96% vs. 91%).
CONCLUSIONS: The 4D/3DC-US method provides reliable and accurate stenosis quantification and binary classification with good diagnostic accuracy compared with CA and DUS.