实时三维超声心动图无创评估右心室功能

Non-invasive Evaluation of Right Ventricular Function with Real-Time 3-D Echocardiography.
PubMed 文献 · Ultrasound Med Biol · 2017-07-14

研究将乳胶球囊置入20个新鲜离体猪心右心室,由搏动泵被动驱动,在不同每搏量(30–70 mL)下用实时三维超声心动图(3-DE)测量右心室整体纵向应变、整体圆周应变、整体面积应变和右心室射血分数,并与超声微测距法对比;3-DE各参数与微测距法均强相关(r=0.89、0.79、0.74、0.93,均P<0.001),3-DE评估右心室功能比2-DE更可行且更准确,整体纵向应变是量化右心室收缩功能的潜在替代参数。

要点

  • 3-DE derived RVEF与超声微测距法相关性最强(r=0.93,P<0.001),优于2-DE RVEF(r=0.80)。
  • 3-DE整体纵向应变与微测距法一致性优于整体圆周应变和2-DE RVEF。
  • 实时3-DE评估右心室功能比2-DE更可行且更准确,整体纵向应变是潜在替代参数。

核心结论

  • 实时三维超声心动图评估右心室功能比二维超声心动图更可行且更准确,整体纵向应变是量化右心室收缩功能的潜在替代参数。

研究设计与样本

  • 离体动物实验:20个新鲜离体猪心右心室置入乳胶球囊并由搏动泵被动驱动,在不同每搏量(30–70 mL)下用实时三维超声心动图测量整体纵向应变、整体圆周应变、整体面积应变和右心室射血分数,用二维超声心动图测量右心室射血分数,以超声微测距法数据为参考标准进行对比。

诊断标准与关键数值

  • 3-DE整体纵向应变、整体圆周应变、整体面积应变、2-DE右心室射血分数和3-DE右心室射血分数与超声微测距法数据的相关系数分别为r=0.89、0.79、0.74、0.80和0.93(均P<0.001);Bland-Altman分析显示超声衍生整体纵向应变、整体圆周应变、2-DE右心室射血分数和3-DE右心室射血分数较超声微测距法值轻度高估,偏倚分别为1.55、2.72、3.59和2.21。

临床应用与阅片要点

  • 作者认为实时三维超声心动图评估右心室功能比二维超声心动图更可行且更准确,整体纵向应变是量化右心室收缩功能的潜在替代参数。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

The aim of this study was to evaluate the accuracy and feasibility of real-time 3-D echocardiography (3-DE) in assessing right ventricular (RV) systolic function. A latex balloon was inserted into the right ventricle of 20 freshly harvested pig hearts which were then passively driven by a pulsatile pump apparatus. The RV global longitudinal strain (GLS), global circumferential strain (GCS), global area strain (GAS) and RV ejection fraction (RVEF), derived from 3-DE, as well as the RVEF obtained from 2-D echocardiography (2-DE) were quantified at different stroke volumes (30-70 mL) and compared with sonomicrometry data. In all comparisons, 3-D GLS, GCS, GAS, 2-D RVEF and 3-D RVEF exhibited strong correlations with sonomicrometry data (r = 0.89, 0.79, 0.74, 0.80, and 0.93, respectively; all p values < 0.001). Bland-Altman analyses revealed slight overestimations of echo-derived GLS, GCS, 2-DE RVEF and 3-DE RVEF compared with sonomicrometry values (bias = 1.55, 2.72, 3.59 and 2.21, respectively). Furthermore, there is better agreement among GLS, 3-D RVEF and the sonomicrometry values than between GCS and 2-D RVEF. Real-time 3-DE is more feasible and accurate for assessing RV function than 2-DE. GLS is a potential alternative parameter for quantifying RV systolic function.