文献速递神经外周血管CTDSA

基于体模的CT血管造影图像标准化用于点征检测

Phantom-based standardization of CT angiography images for spot sign detection.
PubMed 文献 · Neuroradiology · 2017-07-20

本研究在大规模多中心脑出血临床试验中,设计含已知碘浓度的定制体模并在各机构按卒中CT方案扫描,开发软件计算亨氏单位与碘浓度的关系,比较校准与未校准密度阈值对点征识别的影响。结果显示校准后点征阳性扫描数更高(67.8% vs 38.9%,p < 0.001),对血肿扩大的敏感度更高(76% vs 52%)但特异度更低(35% vs 63%)。

要点

  • 校准后点征阳性扫描比例67.8%,高于未校准的38.9%(p < 0.001)。
  • 校准后对血肿扩大的敏感度76%,高于未校准的52%。
  • 校准后对血肿扩大的特异度35%,低于未校准的63%。

核心结论

  • 使用体模数据对CTA图像进行归一化是获得跨机构一致图像定量以用于点征分析的可行策略,并可能提高识别血肿扩大的敏感度。

研究设计与样本

  • 研究为多中心脑出血临床试验中的体模校准研究,共90例受试者符合条件,其中83例纳入分析。
  • 定制体模含已知碘浓度,在各机构按卒中CT方案扫描。
  • 开发定制软件读取CT容积数据集并计算每个体模扫描的亨氏单位随碘浓度的函数。
  • 由两名训练有素的读者分析症状发作8小时内获得的CTA图像,比较校准与未校准密度阈值对点征识别的影响。
  • 血肿扩大定义为血肿体积增长>33%。

关键影像表现/征象

  • 校准分析中点征阳性扫描数高于未校准分析(67.8% vs 38.9%,p < 0.001)。
  • 未校准分析中识别的所有点征在校准后仍为阳性。

诊断标准与关键数值

  • 点征识别采用校准与未校准的密度阈值进行比较。
  • 校准CTA图像对血肿扩大的敏感度为76%,未校准为52%。
  • 校准CTA图像对血肿扩大的特异度为35%,未校准为63%。
  • 血肿扩大定义为血肿体积增长>33%。

临床应用与阅片要点

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: The CT angiography (CTA) spot sign is a strong predictor of hematoma expansion in intracerebral hemorrhage (ICH). However, CTA parameters vary widely across centers and may negatively impact spot sign accuracy in predicting ICH expansion. We developed a CT iodine calibration phantom that was scanned at different institutions in a large multicenter ICH clinical trial to determine the effect of image standardization on spot sign detection and performance.

METHODS: A custom phantom containing known concentrations of iodine was designed and scanned using the stroke CT protocol at each institution. Custom software was developed to read the CT volume datasets and calculate the Hounsfield unit as a function of iodine concentration for each phantom scan. CTA images obtained within 8 h from symptom onset were analyzed by two trained readers comparing the calibrated vs. uncalibrated density cutoffs for spot sign identification. ICH expansion was defined as hematoma volume growth >33%.

RESULTS: A total of 90 subjects qualified for the study, of whom 17/83 (20.5%) experienced ICH expansion. The number of spot sign positive scans was higher in the calibrated analysis (67.8 vs 38.9% p < 0.001). All spot signs identified in the non-calibrated analysis remained positive after calibration. Calibrated CTA images had higher sensitivity for ICH expansion (76 vs 52%) but inferior specificity (35 vs 63%) compared with uncalibrated images.

CONCLUSION: Normalization of CTA images using phantom data is a feasible strategy to obtain consistent image quantification for spot sign analysis across different sites and may improve sensitivity for identification of ICH expansion.

来源:Neuroradiology,查看原文。版权归原作者所有。