文献速递神经介入US

频率与焦点间距对经颅组织摧毁术血栓液化影响的体外研究

Effect of Frequency and Focal Spacing on Transcranial Histotripsy Clot Liquefaction, Using Electronic Focal Steering.
PubMed 文献 · Ultrasound Med Biol · 2017-07-14

本体外研究使用250和500 kHz的256阵元半球形换能器,经3例离体人颅骨对4 cm直径血栓进行组织摧毁术治疗,比较不同焦点间距(λ/2、2λ/3、λ)的液化效果。最佳参数为500 kHz下λ间距,经3例颅骨的液化量分别为23.1±4.0、37.1±16.9和25.4±16.9 mL,速率分别为3.2±0.6、5.1±2.3和3.5±0.4 mL/min,颅骨温升均低于4°C。

要点

  • 500 kHz联合λ焦点间距为最佳参数组合,经3例颅骨均实现较大液化量。
  • 经3例颅骨的温升均低于4°C,治疗中监测颅骨加热。
  • 液化血栓经导管和注射器引流,范围6–59 mL,用时0.9–42.4 min。

核心结论

  • 500 kHz下λ焦点间距产生最大液化量和较快液化速率,且颅骨温升低于4°C。

研究设计与样本

  • 体外研究,使用250和500 kHz的256阵元半球形换能器(孔径30 cm),经3例离体人颅骨对4 cm直径体外血栓进行组织摧毁术治疗,焦点间距为λ/2、2λ/3和λ,每个位置50个脉冲,容积内脉冲重复频率200 Hz。

诊断标准与关键数值

  • 焦点间距按频率比例设置为λ/2、2λ/3和λ。
  • 每个焦点位置施加50个脉冲,容积内脉冲重复频率200 Hz。
  • 250 kHz占空比为0.08%,500 kHz占空比为0.04%(各焦点位置)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

This in vitro study investigated the effects of ultrasound frequency and focal spacing on blood clot liquefaction via transcranial histotripsy. Histotripsy pulses were delivered using two 256-element hemispherical transducers of different frequency (250 and 500 kHz) with 30-cm aperture diameters. A 4-cm diameter spherical volume of in vitro blood clot was treated through 3 excised human skullcaps by electronically steering the focus with frequency proportional focal spacing: λ/2, 2 λ/3 and λ with 50 pulses per location. The pulse repetition frequency across the volume was 200 Hz, corresponding to a duty cycle of 0.08% (250 kHz) and 0.04% (500 kHz) for each focal location. Skull heating during treatment was monitored. Liquefied clot was drained via catheter and syringe in the range of 6-59 mL in 0.9-42.4 min. The fastest rate was 16.6 mL/min. The best parameter combination was λ spacing at 500 kHz, which produced large liquefaction through 3 skullcaps (23.1 ± 4.0, 37.1 ± 16.9 and 25.4 ± 16.9 mL) with the fast rates (3.2 ± 0.6, 5.1 ± 2.3 and 3.5 ± 0.4 mL/min). The temperature rise through the 3 skullcaps remained below 4°C.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Gerhardson T, Sukovich JR, Pandey AS, et al. Effect of Frequency and Focal Spacing on Transcranial Histotripsy Clot Liquefaction, Using Electronic Focal Steering. Ultrasound Med Biol. 2017. PMID: 28716432. doi:10.1016/j.ultrasmedbio.2017.06.010

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。