频率与焦点间距对经颅组织摧毁术血栓液化影响的体外研究
本体外研究使用250和500 kHz的256阵元半球形换能器,经3例离体人颅骨对4 cm直径血栓进行组织摧毁术治疗,比较不同焦点间距(λ/2、2λ/3、λ)的液化效果。最佳参数为500 kHz下λ间距,经3例颅骨的液化量分别为23.1±4.0、37.1±16.9和25.4±16.9 mL,速率分别为3.2±0.6、5.1±2.3和3.5±0.4 mL/min,颅骨温升均低于4°C。
要点
- 500 kHz联合λ焦点间距为最佳参数组合,经3例颅骨均实现较大液化量。
- 经3例颅骨的温升均低于4°C,治疗中监测颅骨加热。
- 液化血栓经导管和注射器引流,范围6–59 mL,用时0.9–42.4 min。
核心结论
- 500 kHz下λ焦点间距产生最大液化量和较快液化速率,且颅骨温升低于4°C。
研究设计与样本
- 体外研究,使用250和500 kHz的256阵元半球形换能器(孔径30 cm),经3例离体人颅骨对4 cm直径体外血栓进行组织摧毁术治疗,焦点间距为λ/2、2λ/3和λ,每个位置50个脉冲,容积内脉冲重复频率200 Hz。
诊断标准与关键数值
- 焦点间距按频率比例设置为λ/2、2λ/3和λ。
- 每个焦点位置施加50个脉冲,容积内脉冲重复频率200 Hz。
- 250 kHz占空比为0.08%,500 kHz占空比为0.04%(各焦点位置)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
This in vitro study investigated the effects of ultrasound frequency and focal spacing on blood clot liquefaction via transcranial histotripsy. Histotripsy pulses were delivered using two 256-element hemispherical transducers of different frequency (250 and 500 kHz) with 30-cm aperture diameters. A 4-cm diameter spherical volume of in vitro blood clot was treated through 3 excised human skullcaps by electronically steering the focus with frequency proportional focal spacing: λ/2, 2 λ/3 and λ with 50 pulses per location. The pulse repetition frequency across the volume was 200 Hz, corresponding to a duty cycle of 0.08% (250 kHz) and 0.04% (500 kHz) for each focal location. Skull heating during treatment was monitored. Liquefied clot was drained via catheter and syringe in the range of 6-59 mL in 0.9-42.4 min. The fastest rate was 16.6 mL/min. The best parameter combination was λ spacing at 500 kHz, which produced large liquefaction through 3 skullcaps (23.1 ± 4.0, 37.1 ± 16.9 and 25.4 ± 16.9 mL) with the fast rates (3.2 ± 0.6, 5.1 ± 2.3 and 3.5 ± 0.4 mL/min). The temperature rise through the 3 skullcaps remained below 4°C.
引用
- Gerhardson T, Sukovich JR, Pandey AS, et al. Effect of Frequency and Focal Spacing on Transcranial Histotripsy Clot Liquefaction, Using Electronic Focal Steering. Ultrasound Med Biol. 2017. PMID: 28716432. doi:10.1016/j.ultrasmedbio.2017.06.010
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。