荧光透视引导脊柱手术的机构(本地)诊断参考水平
该研究记录了156例颈椎与胸腰椎介入手术的透视时间、空气比释动能面积乘积与累积剂量,并计算机构本地诊断参考水平与行动水平。颈椎与胸腰椎的本地诊断参考水平分别为透视时间0.15min与0.29min、空气比释动能面积乘积0.10Gycm2与0.71Gycm2、累积剂量0.47mGy与3.24mGy;年龄与治疗节段仅对颈椎介入的参考剂量值有显著影响。作者认为所报告的本地诊断参考水平与行动水平可有助于建立国家诊断参考水平并提高神经外科医生对患者剂量与辐射防护的意识。
要点
- 颈椎介入本地诊断参考水平:透视时间0.15min、空气比释动能面积乘积0.10Gycm2、累积剂量0.47mGy。
- 胸腰椎介入本地诊断参考水平:透视时间0.29min、空气比释动能面积乘积0.71Gycm2、累积剂量3.24mGy。
- 体重分层法所得参考值与全组患者相当,尺寸校正法所得参考值较低。
核心结论
- 所报告的本地诊断参考水平与行动水平可有助于建立国家诊断参考水平并提高神经外科医生对患者剂量与辐射防护的意识。
研究设计与样本
- 研究记录了156例接受颈椎与胸腰椎介入手术患者的透视时间、空气比释动能面积乘积、累积剂量及影响手术复杂度的解剖、临床与技术因素。
- 基于空气比释动能面积乘积测量估算了患者入射表面剂量、有效剂量、甲状腺吸收剂量与性腺剂量。
- 本地诊断参考水平与行动水平按全组患者透视时间、空气比释动能面积乘积与累积剂量的第75百分位与第10百分位计算,并采用体重分层法与尺寸校正法计算。
诊断标准与关键数值
- 全组患者颈椎介入本地诊断参考水平:透视时间0.15min、空气比释动能面积乘积0.10Gycm2、累积剂量0.47mGy。
- 全组患者胸腰椎介入本地诊断参考水平:透视时间0.29min、空气比释动能面积乘积0.71Gycm2、累积剂量3.24mGy。
- 全组患者颈椎介入行动水平:透视时间0.03min、空气比释动能面积乘积0.01Gycm2、累积剂量0.05mGy。
- 全组患者胸腰椎介入行动水平:透视时间0.03min、空气比释动能面积乘积0.07Gycm2、累积剂量0.33mGy。
- 年龄与治疗节段仅对颈椎介入的参考剂量值有显著影响,其余纳入因素对颈椎与胸腰椎介入均无统计学显著关联。
临床应用与阅片要点
- 颈椎介入平均入射表面剂量1.58mGy(范围0.02-13.58mGy),平均有效剂量0.012mSv(范围0.001-0.097mSv)。
- 胸腰椎介入平均入射表面剂量23mGy(范围0.004-390.3mGy),平均有效剂量0.124mSv(范围0.00002-2.11mSv)。
- 平均甲状腺剂量0.14mGy(范围0.002-1.12mGy),平均性腺剂量0.044mGy(范围0.000003-1.56mGy)。
- 体重分层法所得参考值与全组患者所得参考值相当,尺寸校正法所得参考值较低。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
In this study, institutional (local) diagnostic reference levels (LDRLs) and action levels (ALs) for spine interventional procedures are reported. Fluoroscopy time (FT), kerma area product (KAP), cumulative dose (CD), as well as anatomical, clinical and technical factors affecting procedure complexity were recorded for 156 patients who underwent cervical and thoraco-lumbar interventions. Patient entrance surface dose (ESD), effective dose (ED), thyroid absorbed dose and gonadal dose were also estimated, based on KAP measurements. The LDRLs and ALs were calculated as the 75th and 10th percentile of FT, KAP and CD values for the total group of patients, as well as utilizing the weight banding method and the size correction method. For the total distribution of patients, the LDRLs for cervical and thoraco-lumbar interventions are 0.15min and 0.29min for FT values, 0.10Gycm2 and 0.71Gycm2 for KAP values, as well as 0.47mGy and 3.24mGy for CD values, respectively. The corresponding ALs are 0.03min and 0.03min, 0.01Gycm2 and 0.07Gycm2, as well as 0.05mGy and 0.33mGy for FT, KAP and CD values, respectively. The age and treated levels had a significant influence on the reference dose values only for cervical interventions, whereas none of the other included factors showed statistically significant association for both cervical and thoraco-lumbar interventions. The weight banding method resulted to reference values comparable to those obtained for the whole group of patients, while the size correction method resulted to lower values. The mean ESD values were 1.58mGy (range 0.02-13.58mGy) for cervical and 23mGy (range 0.004-390.3mGy) for thoraco-lumbar interventions. The corresponding mean ED values were 0.012mSv (range 0.001-0.097mSv) and 0.124mSv (range 0.00002-2.11mSv), respectively. The mean thyroid and gonadal doses were 0.14mGy (range 0.002-1.12mGy) and 0.044mGy (range 0.000003-1.56mGy), respectively. The LDRLs and ALs reported could contribute in the effort for establishing national DRLs and for increasing neurosurgeons awareness regarding patient dose and radiation protection issues during spine interventional procedures.