自动乳腺容积扫描仪与手持超声在乳腺诊断检查中的一致性

Agreement Between an Automated Volume Breast Scanner and Handheld Ultrasound for Diagnostic Breast Examinations.
PubMed 文献 · J Ultrasound Med · 2017-06-01

90例拟行诊断性超声的患者随机先接受手持超声或自动乳腺容积扫描仪(ABVS)检查,ABVS由超声技师或乳腺摄影技师操作,2名放射科医生盲法判读。ABVS与手持超声的κ值为0.831(95%CI 0.744-0.925),7点BI-RADS评分全局一致性κ值为0.488(95%CI 0.372-0.560)。两种操作者间ABVS一致性无显著差异(P=0.47),ABVS与手持超声的诊断特征ROC曲线下面积无统计学差异。

要点

  • ABVS单视图与手持超声诊断工作流一致性良好,κ=0.831(95%CI 0.744-0.925)。
  • ABVS由乳腺摄影技师或超声技师操作,一致性无显著差异(P=0.47)。
  • 7点BI-RADS评分全局一致性κ=0.488(95%CI 0.372-0.560)。

核心结论

  • 单视图诊断性ABVS检查与手持超声诊断工作流一致性良好;ABVS由超声技师或乳腺摄影技师操作无差异。

研究设计与样本

  • 随机、盲法研究,90例拟行诊断性超声检查的患者随机先接受手持超声或ABVS检查,ABVS操作者随机为超声技师或乳腺摄影技师,由2名放射科医生盲法判读,最终诊断经核心活检或随访2年确定,病灶包括9例恶性和81例良性。

诊断标准与关键数值

  • ABVS与手持超声的一致性κ=0.831(95%CI 0.744-0.925)。
  • 7点BI-RADS评分全局一致性κ=0.488(95%CI 0.372-0.560)。
  • 乳腺摄影技师操作ABVS的κ=0.858(95%CI 0.723-0.963),超声技师操作ABVS的κ=0.803(95%CI 0.596-1.000),P=0.47。
  • ABVS特征描述ROC曲线下面积:读片者1为0.91(95%CI 0.84-0.96),读片者2为0.91(95%CI 0.83-0.96)。
  • 手持超声特征描述ROC曲线下面积:读片者1为0.91(95%CI 0.84-0.96),读片者2为0.83(95%CI 0.74-0.90),无统计学差异。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To compare the agreement and interobserver variability of diagnostic handheld ultrasound (US) and a single volume on an automated breast volume scanner (ABVS) and to determine whether there was a significant difference if the ABVS was used by a sonographer or mammographic technologist.

METHODS: Ninety patients scheduled for diagnostic US examinations were randomized to either handheld US or the ABVS first. The AVBS was randomized between a sonographer and a mammographic technologist performing the study. The studies were blinded, randomized, and read by 2 radiologists. The lesion with the highest Breast Imaging Reporting and Data System (BI-RADS) score was used in the analysis. Final diagnoses were made by core biopsy or follow-up for 2 years. Lesions included 9 malignant and 81 benign.

RESULTS: The 90 patients had a mean age ± SD of 53.1 ± 16.3 years. The κ value for agreement between the ABVS and handheld US was 0.831 (95% confidence interval, 0.744-0.925), whereas the global agreement for a 7-point BI-RADS score was 0.488 (0.372-0.560). The agreement between the ABVS and handheld US was nearly the same when the ABVS was used by a mammographic technologist (κ = 0.858 [0.723-0.963]) or sonographer (κ = 0.803 [0.596-1.000]; P = .47). The areas under the receiver operating characteristic curves for characterization by the ABVS were 0.91 (0.84-0.96) for reader 1 and 0.91 (0.83-0.96) for reader 2; those for handheld US were 0.91 (0.84-0.96) for reader 1 and 0.83 (0.74-0.90) for reader 2, with no statistical difference. The agreement based on pathologic images was κ = 0.831 (0.718-0.944); for handheld US, κ = 0.795 (0.623-0.967); and for the AVBS, κ = 0.869 (0.725-1.000).

CONCLUSIONS: Performing a single-view diagnostic ABVS examination has good agreement with a handheld diagnostic US workup. There is no difference if the ABVS is used by a sonographer or mammographic technologist.

来源:J Ultrasound Med,查看原文。版权归原作者所有。