专用乳腺PET与全身PET/CT对乳腺癌病灶显示清晰度的比较
本研究回顾性分析150例女性179个组织学证实的乳腺癌病灶,比较专用乳腺PET(dbPET)与全身PET/CT基于视觉分析的敏感度及SUVmax肿瘤/本底比值(T/B比值)。dbPET与WB-PET/CT在患者/乳腺/病灶基于的敏感度上无显著差异,但dbPET的SUVmax和T/B比值显著更高,提示dbPET肿瘤清晰度更高。
要点
- dbPET病灶基于敏感度92%,WB-PET/CT为88%,两者无显著差异。
- dbPET的SUVmax均值13.0±9.7,WB-PET/CT为6.4±4.8。
- dbPET的T/B比值8.1±7.1显著高于WB-PET/CT的5.1±4.5。
核心结论
- 使用更新重建模式的dbPET与WB-PET/CT在敏感度上无显著差异,但dbPET的T/B比值显著更高,提示dbPET肿瘤清晰度更高。
研究设计与样本
- 回顾性分析150例女性179个组织学证实的乳腺癌病灶,均接受18F-FDG WB-PET/CT和dbPET检查。
诊断标准与关键数值
- 视觉分析敏感度:dbPET患者基于95%、乳腺基于95%、病灶基于92%;WB-PET/CT患者基于95%、乳腺基于94%、病灶基于88%。FDG高摄取肿瘤SUVmax均值±标准差:dbPET 13.0±9.7,WB-PET 6.4±4.8。T/B比值:dbPET 8.1±7.1 vs WB-PET/CT 5.1±4.5。亚组分析中dbPET的T/B比值在cT1c、cT2、cT3、浸润性癌、非特殊型浸润性癌、黏液癌及Grade 1-3均显著高于WB-PET/CT。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The purpose of this study was to evaluate lesion detectability of a dedicated breast positron-emission tomography (dbPET) scanner for breast cancers with an updated reconstruction mode, comparing it to whole-body positron-emission tomography/computed tomography (WB-PET/CT).
MATERIALS AND METHODS: A total of 179 histologically-proven breast cancer lesions in 150 females who underwent both WB-PET/CT and dbPET with 18F-fluorodeoxyglucose were retrospectively analyzed. The patient/breast/lesion-based sensitivities based on visual analysis were compared between dbPET and WB-PET/CT. For lesions visible on both PET images, SUVmax values of the tumors were measured, and tumor-to-background ratios (T/B ratios) of SUVmax were compared between the two scans. Subgroup analyses according to clinical tumor stage, histopathology and histological grade were also performed.
RESULTS: Patient/breast/lesion-based sensitivities were 95%, 95%, and 92%, respectively, for dbPET, and 95%, 94%, and 88%, respectively, for WB-PET/CT. Mean±standard deviation SUVmax values of FDG-avid tumors were 13.0±9.7 on dbPET and 6.4±4.8 on WB-PET. T/B ratios were also significantly higher in dbPET than in WB-PET/CT (8.1±7.1 vs. 5.1±4.5). In the subgroup analysis, no significant differences in sensitivities between dbPET and WB-PET/CT were found. However, T/B ratios of dbPET were significantly higher than those of WB-PET/CT in cT1c, cT2, cT3, invasive cancer, invasive carcinoma of no special type, mucinous carcinoma and Grades 1-3.
CONCLUSION: No significant differences in sensitivities were identified between dbPET using an updated reconstruction mode and WB-PET/CT; however, T/B ratios of dbPET were significantly higher than those of WB-PET/CT, indicating higher tumor conspicuity on dbPET.
引用
- Nishimatsu K, Nakamoto Y, Miyake KK, et al. Higher breast cancer conspicuity on dbPET compared to WB-PET/CT. Eur J Radiol. 2017. PMID: 28583624. doi:10.1016/j.ejrad.2017.02.046
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。