文献速递腹部MR

健康志愿者中刚性及柔性驱动在呼气末与吸气末行二维和三维肝脏磁共振弹性成像的可重复性

Repeatability and reproducibility of 2D and 3D hepatic MR elastography with rigid and flexible drivers at end-expiration and end-inspiration in healthy volunteers.
PubMed 文献 · Abdom Radiol (NY) · 2017-01-01

研究纳入9名健康志愿者,同日两次检查,比较2D与3D肝脏磁共振弹性成像在刚性/柔性驱动及呼气末/吸气末条件下的可重复性。3D磁共振弹性成像的重复性系数较低、组内相关系数较高;吸气末硬度值高于呼气末,3D差异显著;驱动类型间无偏倚。作者指出呼吸时相可能是混杂因素,需更大样本验证。

要点

  • 3D磁共振弹性成像的重复性系数为0.16-0.26,组内相关系数为0.84-0.96,优于2D的0.29-0.49与0.71-0.91。
  • 吸气末肝脏硬度值高于呼气末,3D磁共振弹性成像差异显著(p < 0.01),呼吸时相可能为混杂因素。
  • 刚性驱动与柔性驱动之间未发现偏倚,两者硬度测量可重复。

核心结论

  • 初步研究提示2D和3D磁共振弹性成像在多数条件下具有良好重复性,刚性与柔性驱动测量的硬度可重复,但呼吸时相间可重复性中等,提示呼吸时相可能是磁共振弹性成像硬度测量的混杂因素,需更大规模研究验证这些初步结果。

研究设计与样本

  • 9名健康志愿者接受两次同日磁共振弹性成像检查,间隔5-10分钟休息;每次检查均行2D和3D磁共振弹性成像,各在4种条件下采集(2种驱动类型[刚性、柔性]×2个屏气时相[呼气末、吸气末])。

诊断标准与关键数值

  • 2D磁共振弹性成像重复性系数0.29-0.49、组内相关系数0.71-0.91;3D磁共振弹性成像重复性系数0.16-0.26、组内相关系数0.84-0.96;吸气末与呼气末比较的可重复性系数0.30-0.54、一致性相关系数0.61-0.72;刚性驱动与柔性驱动比较的可重复性系数0.10-0.44、一致性相关系数0.79-0.94。

临床应用与阅片要点

  • 作者指出吸气末硬度值高于呼气末且3D磁共振弹性成像差异显著(p < 0.01),呼吸时相可能是磁共振弹性成像硬度测量的混杂因素,但原文未据此给出具体阅片或报告推荐。

局限性

  • 原文为初步研究,样本量为9名健康志愿者;作者指出需更大规模研究验证这些初步结果。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To evaluate the repeatability and reproducibility of 2D and 3D hepatic MRE with rigid and flexible drivers at end-expiration and end-inspiration in healthy volunteers.

MATERIALS AND METHODS: Nine healthy volunteers underwent two same-day MRE exams separated by a 5- to 10-min break. In each exam, 2D and 3D MRE scans were performed, each under four conditions (2 driver types [rigid, flexible] × 2 breath-hold phases [end-expiration, end-inspiration]). Repeatability (measurements under identical conditions) and reproducibility (measurements under different conditions) were analyzed by calculating bias, limit of agreement, repeatability coefficient (RC), reproducibility coefficient (RDC), intraclass correlation coefficient (ICC), and concordance correlation coefficient (CCC), as appropriate.

RESULTS: For 2D MRE, RCs and ICCs range between 0.29-0.49 and 0.71-0.91, respectively. For 3D MRE, RCs and ICCs range between 0.16-0.26 and 0.84-0.96, respectively. Stiffness values were biased by breath-hold phase, being higher at end-inspiration than end-expiration, and the differences were significant for 3D MRE (p < 0.01). No bias was found between driver types. Inspiration vs. expiration RDCs and CCCs ranged between 0.30-0.54 and 0.61-0.72, respectively. Rigid vs. flexible driver RDCs and CCCs ranged between 0.10-0.44 and 0.79-0.94, respectively.

CONCLUSION: This preliminary study suggests that 2D MRE and 3D MRE under most conditions potentially have good repeatability. Our result also points to the possibility that stiffness measured with the rigid and flexible drivers is reproducible. Reproducibility between breath-hold phases was modest, suggesting breath-hold phase might be a confounding factor in MRE-based stiffness measurement. However, larger studies are required to validate these preliminary results.

来源:Abdom Radiol (NY),查看原文。版权归原作者所有。