慢性肝炎肝纤维化评估:剪切波弹性成像与瞬时弹性成像的比较
该前瞻性横断面研究以肝活检为金标准,纳入240例慢性乙型或丙型肝炎患者,同日行剪切波弹性成像(SWE)与瞬时弹性成像(TE),比较两者诊断肝纤维化的效能。SWE对进展期纤维化/肝硬化的诊断表现优于TE,联合两种方法可提高特异度。
要点
- SWE在鉴别F2与F3-4纤维化时,以7.0 kPa为阈值,敏感度81.3%、特异度77.6%。
- TE以8.3 kPa为阈值鉴别F2与F3-4时,敏感度81.8%、特异度83.1%。
- SWE与TE联合应用时,各纤维化分期的敏感度下降而特异度提高。
核心结论
- SWE是评估肝纤维化的准确技术,其表现与TE相当,尤其在预测进展期纤维化/肝硬化方面更优;联合SWE与TE可进一步提高特异度。
研究设计与样本
- 前瞻性横断面研究,纳入连续慢性乙型/丙型肝炎患者240例(CHB 172例,CHC 68例),男176例、女64例,平均年龄32.6±11.6岁;以肝活检为金标准,SWE、TE与活检于同日进行。
诊断标准与关键数值
- SWE鉴别F2与F3-4纤维化:以7.0 kPa为阈值,敏感度81.3%、特异度77.6%。
- TE鉴别F2与F3-4纤维化:以8.3 kPa为阈值,敏感度81.8%、特异度83.1%。
- SWE鉴别F3与F4:敏感度83.3%、特异度90.7%。
- TE鉴别F3与F4:以14.8 kPa为阈值,敏感度83.3%、特异度96.5%。
- 无组织学纤维化(F0)患者平均LSM:SWE 5.0±0.7 kPa,TE 5.1±1.4 kPa。
临床应用与阅片要点
- SWE与TE联合应用时,所有肝纤维化分期的敏感度下降而特异度提高。
- SWE与TE测值间存在显著相关性(r=0.33,p<0.001)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To evaluate the diagnostic accuracy of shear wave elastography (SWE) and transient elastography (TE) in the evaluation of liver fibrosis in chronic hepatitis B (CHB) and C (CHC) patients taking liver biopsy as gold standard.
METHODS: Ethics committee approved this prospective cross-sectional study. Between October 2012 and December 2014, consecutive CHB/CHC patients fulfilling the inclusion criteria were included-age more than 18 years, informed written consent, willing and suitable for liver biopsy. SWE, TE, and biopsy were performed the same day. Liver stiffness measurement (LSM) cut-offs for various stages of fibrosis were generated for SWE and TE. AUC, sensitivity, specificity, and positive/negative predictive values were estimated individually or in combination.
RESULTS: CH patients (n = 240, CHB 172, CHC 68), 176 males, 64 females, mean age 32.6 ± 11.6 years were enrolled. Mean LSM of patients with no histological fibrosis (F0) was 5.0 ± 0.7 and 5.1+1.4 kPa on SWE and TE, respectively. For differentiating F2 and F3-4 fibrosis on SWE, at 7.0 kPa cut-off, the sensitivity was 81.3% and specificity 77.6%. For TE, at 8.3 kPa cut-off, sensitivity was 81.8% and specificity 83.1%. For F3 vs. F4, SWE sensitivity was 83.3% and specificity 90.7%. At 14.8 kPa cut-off, TE showed similar sensitivity (83.3%) but specificity increased to 96.5%. Significant correlation between SWE and TE was observed (r = 0.33, p < 0.001). On combining SWE and TE, a drop in sensitivity with increased specificity for all stages of liver fibrosis occured.
CONCLUSION: SWE is an accurate technique for evaluating liver fibrosis. SWE compares favorably with TE especially for predicting advanced fibrosis/cirrhosis. Combining SWE and TE further improves specificity.
引用
- Paul SB, Das P, Mahanta M, et al. Assessment of liver fibrosis in chronic hepatitis: comparison of shear wave elastography and transient elastography. Abdom Radiol (NY). 2017. PMID: 28643136. doi:10.1007/s00261-017-1213-5
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。