机器人驱动的CT引导针插入:体模和动物实验的初步结果
在体模实验中比较机器人远控与手动CT引导针插入的准确性,18次机器人插入与18次手动插入平均误差分别为1.6 mm和1.4 mm,差异95%CI为-0.3至0.6 mm;医生有效剂量机器人插入始终为0 μSv,手动插入平均5.7 μSv(P < .001)。动物实验中机器人插入可行,总体平均误差3.2 mm,出现3例轻微操作相关并发症。
要点
- 体模中机器人插入与手动插入准确性相当,平均误差1.6 mm对1.4 mm
- 机器人插入时医生有效剂量始终为0 μSv,手动插入平均5.7 μSv
- 动物实验中机器人插入总体平均误差3.2 mm,出现3例轻微并发症
核心结论
- 机器人插入在体模中表现出与手动插入相当的准确性,且医生无辐射暴露;在动物体内操作中也被发现是准确的。
研究设计与样本
- 体模实验:18次机器人插入与18次手动插入,使用19G针,CT透视引导,以1.0 mm界值评估准确性等效性。动物实验:经机构动物护理与使用委员会批准,3头猪的肝脏、肾脏、肺和髋部肌肉目标,使用19G或17G针,常规CT引导,尝试5次机器人插入,评估可行性、安全性和准确性。
诊断标准与关键数值
- 体模实验以1.0 mm界值评估机器人插入与手动插入的准确性等效性
- 体模中机器人插入平均误差1.6 mm,手动插入平均误差1.4 mm,平均差值95%CI为-0.3至0.6 mm
- 体模中两种方法在针插入时间、CT透视时间或体模辐射暴露方面无显著差异
- 医生有效剂量:机器人插入始终为0 μSv,手动插入平均5.7 μSv(P < .001)
- 动物实验中机器人插入总体平均误差3.2 mm,出现3例轻微操作相关并发症
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Purpose To evaluate the accuracy of the remote-controlled robotic computed tomography (CT)-guided needle insertion in phantom and animal experiments. Materials and Methods In a phantom experiment, 18 robotic and manual insertions each were performed with 19-gauge needles by using CT fluoroscopic guidance for the evaluation of the equivalence of accuracy of insertion between the two groups with a 1.0-mm margin. Needle insertion time, CT fluoroscopy time, and radiation exposure were compared by using the Student t test. The animal experiments were approved by the institutional animal care and use committee. In the animal experiment, five robotic insertions each were attempted toward targets in the liver, kidneys, lungs, and hip muscle of three swine by using 19-gauge or 17-gauge needles and by using conventional CT guidance. The feasibility, safety, and accuracy of robotic insertion were evaluated. Results The mean accuracies of robotic and manual insertion in phantoms were 1.6 and 1.4 mm, respectively. The 95% confidence interval of the mean difference was -0.3 to 0.6 mm. There were no significant differences in needle insertion time, CT fluoroscopy time, or radiation exposure to the phantom between the two methods. Effective dose to the physician during robotic insertion was always 0 μSv, while that during manual insertion was 5.7 μSv on average (P < .001). Robotic insertion was feasible in the animals, with an overall mean accuracy of 3.2 mm and three minor procedure-related complications. Conclusion Robotic insertion exhibited equivalent accuracy as manual insertion in phantoms, without radiation exposure to the physician. It was also found to be accurate in an in vivo procedure in animals. © RSNA, 2017 Online supplemental material is available for this article.
引用
- Hiraki T, Kamegawa T, Matsuno T, et al. Robotically Driven CT-guided Needle Insertion: Preliminary Results in Phantom and Animal Experiments. Radiology. 2017. PMID: 28604237. doi:10.1148/radiol.2017162856
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。