高螺距螺旋双源CT血管成像在复杂先天性心脏病儿童中与回顾性门控螺旋采集的可行性比较
研究纳入50例小于4岁的复杂先天性心脏病儿童,随机分为高螺距(螺距3.4)螺旋双源CT组(A组,25例)和回顾性螺旋双源CT组(B组,25例)。两组图像质量、诊断准确率及冠状动脉起源与走行显示无显著差异,但A组有效剂量显著低于B组(0.40±0.20 mSv vs 2.7±1.0 mSv,p<0.05)。作者认为高螺距心电触发双源CT能以低辐射剂量和良好图像质量清晰显示心内外畸形及冠状动脉异常。
要点
- A组有效剂量0.40±0.20 mSv,显著低于B组2.7±1.0 mSv(p<0.05)。
- 两组图像质量评分分别为A组1.3±0.4、B组1.1±0.3(p=0.2),无显著差异。
- 两组诊断准确率均为100%,冠状动脉可追踪率A组80%、B组84%(p=0.7)。
核心结论
- 高螺距心电触发双源CT可在复杂先天性心脏病儿童中以低辐射剂量和良好图像质量清晰显示心内外畸形、冠状动脉起源及走行异常。
研究设计与样本
- 研究为随机分组对照研究,2014年7月至2016年6月连续纳入50例小于4岁的复杂先天性心脏病儿童,随机分为高螺距(螺距3.4)螺旋双源CT组(A组,25例)和回顾性螺旋双源CT组(B组,25例),均行心电触发双源CT检查,比较图像质量、诊断准确率、冠状动脉起源与走行显示及辐射剂量。
关键影像表现/征象
- A组发现1例单冠状动脉,B组发现1例左前降支起源于右冠状动脉。
- 两组图像质量评分分别为A组1.3±0.4、B组1.1±0.3(p=0.2),无显著差异。
- 两组冠状动脉可追踪率分别为A组80%、B组84%(p=0.7),无显著差异。
诊断标准与关键数值
- A组诊断准确率100%,B组诊断准确率100%。
- A组有效剂量0.40±0.20 mSv,B组有效剂量2.7±1.0 mSv,两组差异有统计学意义(p<0.05)。
临床应用与阅片要点
- 作者认为高螺距心电触发双源CT可用于复杂先天性心脏病儿童,在低辐射剂量下获得良好图像质量并清晰显示冠状动脉起源与走行异常。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM: To investigate the use of second-generation dual-source high-pitch computed tomography in obtaining confident diagnostic image quality using a low radiation dose in young patients with congenital heart disease (CHD).
MATERIALS AND METHODS: From July 2014 to June 2016, 50 consecutive children <4 years with complex CHD underwent electrocardiography (ECG)-triggered dual-source computed tomography (CT). The patients were assigned randomly to two groups: high-pitch (pitch 3.4) spiral dual-source CT acquisition (group A) and retrospectively spiral dual-source CT acquisition (group B). The image quality, diagnostic accuracy, coronary artery origin, course demonstration, and radiation exposure were compared between the two groups.
RESULTS: Fifty examinations were performed (group A, 25; group B, 25). There were no significant differences in image quality, diagnostic accuracy, coronary artery origin, and course demonstration between the two groups. The image quality scores were 1.3±0.4 in group A and 1.1±0.3 in group B (p=0.2). The diagnostic accuracy was 100% in both groups. The coronary arteries were traceable in 80% in group A and 84% in group B (p=0.7). A single coronary artery was identified in one case in group A and the left anterior descending (LAD) branch originated from the right coronary artery (RCA) in one case in group B. There were significant differences in the effective doses between the two groups (0.40±0.20 mSv in group A and 2.7±1.0 mSv in group B, p<0.05).
CONCLUSION: Intra-cardiac and extra-cardiac malformation, coronary artery origin, and course malformation can be visualised clearly using a high-pitch ECG-triggered dual-source CT with a low radiation dose and good image quality in patients with CHD.
引用
- Li T, Zhao S, Liu J, et al. Feasibility of high-pitch spiral dual-source CT angiography in children with complex congenital heart disease compared to retrospective-gated spiral acquisition. Clin Radiol. 2017. PMID: 28576437. doi:10.1016/j.crad.2017.05.005
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。