采用迭代射束硬化校正面重建的锡滤过CT冠状动脉钙化评分
本研究回顾性分析62例患者的影像数据,比较标准120kVp与Sn100kVp(采用迭代射束硬化校正钙质材料重建)的CT冠状动脉钙化评分。结果显示两种采集方法的Agatston评分相关性极好(r=0.99),风险分类一致性高(ĸ=0.99),且Sn100kVp的有效辐射剂量显著降低(0.19±0.07 mSv vs. 0.83±0.33 mSv,降低75%)。
要点
- Sn100kVp联合迭代射束硬化校正重建与120kVp的Agatston评分相关性极好(r=0.99,p<0.0001)。
- 两种采集方法的Agatston评分分类和基于百分位的心脏风险分类一致性高(ĸ=1.00和ĸ=0.99)。
- Sn100kVp的有效辐射剂量较120kVp降低75%(0.19±0.07 mSv vs. 0.83±0.33 mSv,p<0.0001)。
核心结论
- 采用专用迭代射束硬化校正钙质材料重建算法的低电压锡滤过CT冠状动脉钙化评分,与标准120kVp采集在Agatston评分和心脏风险分类方面显示极好的相关性和一致性,同时辐射剂量显著降低75%至胸部X线水平。
研究设计与样本
- 回顾性分析62例临床指示行标准120kVp冠状动脉钙化评分采集并额外行Sn100kVp扫描的患者影像数据,使用专用光谱迭代射束硬化校正冠状动脉钙化评分重建恢复120kVp光谱响应,比较Agatston评分、风险分类及辐射剂量。
诊断标准与关键数值
- 迭代射束硬化校正Sn100kVp与120kVp采集的Agatston评分中位数分别为31.7和34.1(p=0.057)。
- 两种采集的Agatston评分相关性极好(r=0.99,p<0.0001)。
- Agatston评分分类和基于百分位的心脏风险分类一致性高(ĸ=1.00和ĸ=0.99),使用迭代射束硬化校正冠状动脉钙化评分重建导致低心脏风险重分类率为1.6%。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To investigate the diagnostic accuracy of CT coronary artery calcium scoring (CACS) with tin pre-filtration (Sn100kVp) using iterative beam-hardening correction (IBHC) calcium material reconstruction compared to the standard 120kVp acquisition.
BACKGROUND: Third generation dual-source CT (DSCT) CACS with Sn100kVp acquisition allows significant dose reduction. However, the Sn100kVp spectrum is harder with lower contrast compared to 120kVp, resulting in lower calcium score values. Sn100kVp spectral correction using IBHC-based calcium material reconstruction may restore comparable calcium values.
METHODS: Image data of 62 patients (56% male, age 63.9±9.2years) who underwent a clinically-indicated CACS acquisition using the standard 120kVp protocol and an additional Sn100kVp CACS scan as part of a research study were retrospectively analyzed. Datasets of the Sn100kVp scans were reconstructed using a dedicated spectral IBHC CACS reconstruction to restore the spectral response of 120kVp spectra. Agatston scores were derived from 120kVp and IBHC reconstructed Sn100kVp studies. Pearson's correlation coefficient was assessed and Agatston score categories and percentile-based risk categorization were compared.
RESULTS: Median Agatston scores derived from IBHC Sn100kVp scans and 120kVp acquisition were 31.7 and 34.1, respectively (p=0.057). Pearson's correlation coefficient showed excellent correlation between the acquisitions (r=0.99, p<0.0001). Agatston score categories and percentile-based cardiac risk categories showed excellent agreement (ĸ=1.00 and ĸ=0.99), resulting in a low cardiac risk reclassification of 1.6% with the use of IBHC CACS reconstruction. Image noise was 24.9±3.6HU in IBHC Sn100kVp and 17.1±3.9HU in 120kVp scans (p<0.0001). The dose-length-product was 13.2±3.4mGycm with IBHC Sn100kVp and 59.1±22.9mGycm with 120kVp scans (p<0.0001), resulting in a significantly lower effective radiation dose (0.19±0.07mSv vs. 0.83±0.33mSv, p<0.0001) for IBHC Sn100kVp scans.
CONCLUSION: Low voltage CACS with tin filtration using a dedicated IBHC CACS material reconstruction algorithm shows excellent correlation and agreement with the standard 120kVp acquisition regarding Agatston score and cardiac risk categorization, while radiation dose is significantly reduced by 75% to the level of a chest x-ray.