频率可调电容式微机械超声换能器探头原型的临床前测试
该工作在心内超声心动图(ICE)临床前场景下测试两种频率可调探头原型:直径11 mm刚性探头及新型12-Fr柔性可操控ICE导管,二者均采用前向32阵元电容式微机械超声换能器阵列(孔径2×2 mm²),塌陷模式下可在6–18 MHz范围调频。刚性探头在离体被动心脏平台高穿透(6 MHz)、通用(12 MHz)和高分辨(18 MHz)模式下采集主动脉瓣图像,兼具满意图像质量与2.5–10 cm穿透深度;ICE导管在猪模型体内多深度采集主动脉瓣图像,前向设计与调频能力结合可显示距导管尖端不同距离、不同大小的心内结构,提供宽视野与细节特写。
要点
- 两种探头原型均采用前向32阵元CMUT阵列(孔径2×2 mm²),塌陷模式下可在6–18 MHz范围调频。
- 刚性探头离体被动心脏平台高穿透6 MHz、通用12 MHz、高分辨18 MHz模式图像兼具满意质量与2.5–10 cm穿透深度。
- 12-Fr柔性可操控ICE导管在猪模型体内多深度采集主动脉瓣图像,前向设计与调频结合可显示不同距离和大小的心内结构。
核心结论
- 前向设计与频率可调能力的结合使探头可在不同距离显示不同大小的心内结构,提供宽视野与细节特写。
研究设计与样本
- 临床前测试研究:刚性探头(直径11 mm)在离体被动心脏平台测试,新型12-Fr柔性可操控ICE导管在猪动物模型体内测试。
关键影像表现/征象
- 刚性探头高穿透6 MHz、通用12 MHz、高分辨18 MHz模式采集的主动脉瓣图像兼具满意图像质量与2.5–10 cm穿透深度。
- ICE导管置于升主动脉多深度,在3种成像模式下采集主动脉瓣图像,前向设计与调频能力结合可显示距导管尖端不同距离、不同大小的心内结构。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
In intracardiac echocardiography (ICE) it may be beneficial to generate ultrasound images acquired at multiple frequencies, having the possibility of high penetration or high-resolution imaging in a single device. The objective of the presented work is to test two frequency-tunable probe prototypes in a preclinical setting: a rigid probe having a diameter of 11 mm and a new flexible and steerable 12-Fr ICE catheter. Both probes feature a forward-looking 32-element capacitive micromachined ultrasonic transducer array (aperture of 2 × 2 mm2) operated in collapse mode, which allows for frequency tuning in the 6-MHz-18-MHz range. The rigid probe prototype is tested ex vivo in a passive heart platform. Images of an aortic valve acquired in high-penetration (6 MHz), generic (12 MHz) and high-resolution (18 MHz) mode combine satisfying image quality and penetration depth between 2.5 cm and 10 cm. The ICE catheter prototype is tested in vivo using a porcine animal model. Images of an aortic valve are acquired in the 3 imaging modes with the ICE catheter placed in an ascending aorta at multiple depths. It was found that the combination of the forward-looking design and frequency-tuning capability allows visualizing intracardiac structures of various sizes at different distances relative to the catheter tip, providing both wide overviews and detailed close-ups.
引用
- Pekař M, Kolen AF, Belt H, et al. Preclinical Testing of Frequency-Tunable Capacitive Micromachined Ultrasonic Transducer Probe Prototypes. Ultrasound Med Biol. 2017. PMID: 28645796. doi:10.1016/j.ultrasmedbio.2017.05.005
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。