筛查CT检出的肺结节体积测量变异随结节大小变化的研究
本研究基于低剂量CT筛查数据库,选取至少三次扫描、2年内大小稳定的实性肺结节,用Lung VCAR与syngo.via两种软件测量最近两次0.625 mm层厚CT的结节体积,计算体积变化百分比的标准差(SD)。共171个稳定结节(117名受试者,中位年龄61岁),直径2.2–18.7 mm。两种软件的体积变化百分比SD均随结节直径增大而减小,且两种软件间存在差异。
要点
- 结节越小,体积测量变异越大:Lung VCAR在<4 mm组体积变化百分比SD为20.4%,≥10 mm组仅3.7%。
- syngo.via在<4 mm组SD达59.5%,明显高于Lung VCAR同组20.4%。
- 两种软件体积变化百分比SD均随结节直径增大而下降,但软件间数值不同。
核心结论
- 两种软件测量体积变化百分比SD均随结节直径增大而减小,且两种软件间存在差异。
研究设计与样本
- 回顾性分析大型CT筛查数据库,纳入至少三次扫描、2年内大小稳定的实性肺结节,用Lung VCAR和syngo.via测量最近两次0.625 mm层厚CT的结节体积并计算体积变化百分比SD。
诊断标准与关键数值
- Lung VCAR体积变化百分比SD:<4 mm组20.4%,4-5 mm组17.7%,6-9 mm组14.6%,≥10 mm组3.7%。
- syngo.via体积变化百分比SD:<4 mm组59.5%,4-5 mm组24.3%,6-9 mm组9.1%,≥10 mm组6.2%。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The objective of this study is to evaluate measurement variability in volumetric assessment of pulmonary nodules on low-dose CT images with a view toward determining how this variability is influenced by nodule size.
MATERIALS AND METHODS: A large CT screening database was reviewed to identify solid pulmonary nodules that had remained stable in size on the basis of findings from at least three scans obtained over a 2-year period. Two software packages (Lung VCAR and syngo.via) were used to assess the nodule volume on the two most recent CT scans, which were obtained at a slice thickness of 0.625 mm. The percentage of volume change was calculated for each nodule. The SD of the percentage of volume change was determined for nodules in each of the following nodule diameter size categories: less than 4 mm, 4-5 mm, 6-9 mm, and 10 mm or larger. The diameter was the mean of the length and width in the CT image that represented the largest cross-sectional area of the nodule.
RESULTS: The 171 stable nodules that were identified in 117 CT screening participants (median age, 61 years) ranged in size from 2.2 to 18.7 mm. The time between acquisition of the first and last CT images ranged from 3.7 to 17.8 years (median, 11.5 years). For each of the four categories of diameter size (< 4, 4-5, 6-9, and ≥ 10 mm), the SD of the percentage of volume change was 20.4%, 17.7%, 14.6%, and 3.7%, with the use of Lung VCAR, and 59.5%, 24.3%, 9.1%, and 6.2%, with the use of syngo.via, respectively. The SD decreased with increasing nodule diameter, with the use of both software packages.
CONCLUSION: Measurement variability decreased with increasing nodule diameter for both software packages and was different between the two software packages.