儿童双能量平扫脑CT的先进虚拟单能重建:与标准单能图像及传统多能CT的图像质量比较
回顾性重建23例儿童双源双能量平扫脑CT的40–100 keV(每5 keV)虚拟单能图像,比较传统单能重建(Mono)与先进单能重建(Mono+)及传统多能CT的图像质量。Mono+在60 keV处灰白质噪声比、信噪比和对比噪声比最优,70 keV处后颅窝伪影最少;Mono+各图像质量指标均优于Mono及多能CT,主观评分亦最优。
要点
- 儿童双能量平扫脑CT中,Mono+重建在60 keV处灰白质噪声比、信噪比和对比噪声比最优。
- Mono+在70 keV处后颅窝伪影最少,优于Mono(75 keV)及多能CT。
- Mono+图像质量各项指标及主观评分均优于Mono和多能CT。
核心结论
- 采用Mono+算法的优化虚拟单能成像在灰白质区分和后颅窝伪影减少方面表现出更好的图像质量。
研究设计与样本
- 回顾性研究,纳入23例儿童的双源双能量平扫脑CT扫描,重建40–100 keV(每5 keV)共13个虚拟单能能级,比较Mono+、Mono与传统多能CT的图像质量。
关键影像表现/征象
- Mono+在60 keV处灰白质噪声比、信噪比和对比噪声比最优,70 keV处后颅窝伪影最少。
- Mono在65–70 keV处图像质量指标最优,75 keV处后颅窝伪影最少。
- Mono+在图像质量指标上优于Mono和多能CT,主观分析亦评价Mono+优于其他图像集。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Advanced virtual monochromatic reconstruction from dual-energy brain CT has not been evaluated in children.
OBJECTIVE: To determine the most effective advanced virtual monochromatic imaging energy level for maximizing pediatric brain parenchymal image quality in dual-energy unenhanced brain CT and to compare this technique with conventional monochromatic reconstruction and polychromatic scanning.
MATERIALS AND METHODS: Using both conventional (Mono) and advanced monochromatic reconstruction (Mono+) techniques, we retrospectively reconstructed 13 virtual monochromatic imaging energy levels from 40 keV to 100 keV in 5-keV increments from dual-source, dual-energy unenhanced brain CT scans obtained in 23 children. We analyzed gray and white matter noise ratios, signal-to-noise ratios and contrast-to-noise ratio, and posterior fossa artifact. We chose the optimal mono-energetic levels and compared them with conventional CT.
RESULTS: For Mono+maximum optima were observed at 60 keV, and minimum posterior fossa artifact at 70 keV. For Mono, optima were at 65-70 keV, with minimum posterior fossa artifact at 75 keV. Mono+ was superior to Mono and to polychromatic CT for image-quality measures. Subjective analysis rated Mono+superior to other image sets.
CONCLUSION: Optimal virtual monochromatic imaging using Mono+ algorithm demonstrated better image quality for gray-white matter differentiation and reduction of the artifact in the posterior fossa.
引用
- Park J, Choi YH, Cheon JE, et al. Advanced virtual monochromatic reconstruction of dual-energy unenhanced brain computed tomography in children: comparison of image quality against standard mono-energetic images and conventional polychromatic computed tomography. Pediatr Radiol. 2017. PMID: 28656326. doi:10.1007/s00247-017-3908-8