一项协调多中心脑MRI研究对一名多发性硬化症患者的容积分析

Volumetric Analysis from a Harmonized Multisite Brain MRI Study of a Single Subject with Multiple Sclerosis.
PubMed 文献 · AJNR Am J Neuroradiol · 2017-06-22

北美多发性硬化症影像合作指导委员会制定统一的高分辨率3T MR方案,在美国7个中心对一名复发缓解型MS志愿者进行扫描-再扫描。结果显示,中心间专家勾画的T1和T2病灶体积存在显著系统偏倚(P < .01),中心因素解释>90%的变异;多数自动容积测量中心因素解释>80%的变异。结论认为,即使场强和厂商一致且方案协调,系统差异仍可导致容积分析严重偏倚。

要点

  • 多中心脑MRI容积分析中,中心间系统偏倚可解释病灶体积变异>90%。
  • 专家勾画的T1和T2病灶体积中心间差异显著(P < .01)。
  • 自动容积测量结果按扫描仪型号聚类,Skyra与其他2款机型结果相似。

核心结论

  • 即使多中心研究在方案协调后扫描仪场强和厂商一致,系统差异仍可导致容积分析严重偏倚。

研究设计与样本

  • 多中心扫描-再扫描研究,7个美国中心,1名复发缓解型MS志愿者,全部使用Siemens扫描仪(4台Skyra、2台Tim Trio、1台Verio),3T高分辨率MR协议,参考标准为专家手动分割的T1低信号和T2(FLAIR)高信号病灶。

诊断标准与关键数值

  • 专家勾画T1病灶体积中心间范围13.0-16.4 mL,中心因素解释>90%变异,P < .01。
  • 专家勾画T2病灶体积中心间范围15.9-20.1 mL,中心因素解释>90%变异,P < .01。
  • 多数自动容积测量中心因素解释>80%的变异。
  • 自动测量结果按扫描仪型号聚类,Skyra与其他2款机型结果相似。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: MR imaging can be used to measure structural changes in the brains of individuals with multiple sclerosis and is essential for diagnosis, longitudinal monitoring, and therapy evaluation. The North American Imaging in Multiple Sclerosis Cooperative steering committee developed a uniform high-resolution 3T MR imaging protocol relevant to the quantification of cerebral lesions and atrophy and implemented it at 7 sites across the United States. To assess intersite variability in scan data, we imaged a volunteer with relapsing-remitting MS with a scan-rescan at each site.

MATERIALS AND METHODS: All imaging was acquired on Siemens scanners (4 Skyra, 2 Tim Trio, and 1 Verio). Expert segmentations were manually obtained for T1-hypointense and T2 (FLAIR) hyperintense lesions. Several automated lesion-detection and whole-brain, cortical, and deep gray matter volumetric pipelines were applied. Statistical analyses were conducted to assess variability across sites, as well as systematic biases in the volumetric measurements that were site-related.

RESULTS: Systematic biases due to site differences in expert-traced lesion measurements were significant (P < .01 for both T1 and T2 lesion volumes), with site explaining >90% of the variation (range, 13.0-16.4 mL in T1 and 15.9-20.1 mL in T2) in lesion volumes. Site also explained >80% of the variation in most automated volumetric measurements. Output measures clustered according to scanner models, with similar results from the Skyra versus the other 2 units.

CONCLUSIONS: Even in multicenter studies with consistent scanner field strength and manufacturer after protocol harmonization, systematic differences can lead to severe biases in volumetric analyses.