儿童合成脑磁共振成像的图像质量评估
本研究纳入29名儿童(中位年龄6岁,范围0-16岁),对比合成MRI与传统序列的图像质量。结果显示合成T1和T2加权图像诊断可接受,灰白质区分和病灶显示与常规相当;但合成FLAIR图像质量明显降低,诊断不可接受。合成序列采集时间约为常规序列的三分之二。
要点
- 合成T1和T2加权图像在儿童中诊断可接受,但FLAIR图像质量显著降低。
- 合成序列采集时间约为常规序列的三分之二。
- 合成T1和T2图像的灰白质区分与常规相当或更好,病灶显示无差异。
核心结论
- 合成T1和T2加权图像诊断可接受,但合成FLAIR图像诊断不可接受;病灶显示和灰白质区分与常规MRI相当。
研究设计与样本
- 研究纳入29名儿童(中位年龄6岁,范围0-16岁),均接受合成和常规脑MRI;生成参数匹配的合成T2加权、T1加权和FLAIR图像;采用5分制评估整体图像质量、灰白质区分、病灶显示和图像退化,定量评估相对对比度,并比较两种技术的采集时间。
关键影像表现/征象
- 合成图像因更明显的图像退化而质量较差,但在小于2岁儿童中T1和T2加权图像无显著差异。
- 合成T1和T2加权图像的灰白质区分与常规相当或更好,而FLAIR图像灰白质区分较差。
- 病灶显示未受影响,合成图像相对对比度等于或高于常规图像。
诊断标准与关键数值
- 图像质量采用5分制评估,合成T1和T2加权图像质量在诊断可接受范围内,FLAIR图像质量大大降低。
- 相对对比度定量评估显示合成图像等于或高于常规图像。
临床应用与阅片要点
- 合成T1和T2加权图像可用于儿童脑MRI,但合成FLAIR图像质量不足,需谨慎使用。
- 合成序列采集时间约为常规序列的三分之二,可能提高检查效率。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: The clinical application of the multi-echo, multi-delay technique of synthetic magnetic resonance imaging (MRI) generates multiple sequences in a single acquisition but has mainly been used in adults.
OBJECTIVE: To evaluate the image quality of synthetic brain MR in children compared with that of conventional images.
MATERIALS AND METHODS: Twenty-nine children (median age: 6 years, range: 0-16 years) underwent synthetic and conventional imaging. Synthetic (T2-weighted, T1-weighted and fluid-attenuated inversion recovery [FLAIR]) images with settings matching those of the conventional images were generated. The overall image quality, gray/white matter differentiation, lesion conspicuity and image degradations were rated on a 5-point scale. The relative contrasts were assessed quantitatively and acquisition times for the two imaging techniques were compared.
RESULTS: Synthetic images were inferior due to more pronounced image degradations; however, there were no significant differences for T1- and T2-weighted images in children <2 years old. The quality of T1- and T2-weighted images were within the diagnostically acceptable range. FLAIR images showed greatly reduced quality. Gray/white matter differentiation was comparable or better in synthetic T1- and T2-weighted images, but poorer in FLAIR images. There was no effect on lesion conspicuity. Synthetic images had equal or greater relative contrast. Acquisition time was approximately two-thirds of that for conventional sequences.
CONCLUSION: Synthetic T1- and T2-weighted images were diagnostically acceptable, but synthetic FLAIR images were not. Lesion conspicuity and gray/white matter differentiation were comparable to conventional MRI.
引用
- Lee SM, Choi YH, Cheon JE, et al. Image quality at synthetic brain magnetic resonance imaging in children. Pediatr Radiol. 2017. PMID: 28638982. doi:10.1007/s00247-017-3913-y
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。