文献速递骨肌儿科影像介入CTDSA

介入放射学套件中使用导航软件进行儿童经皮骨活检的实时透视针引导

Real-time fluoroscopic needle guidance in the interventional radiology suite using navigational software for percutaneous bone biopsies in children.
PubMed 文献 · Pediatr Radiol · 2017-05-04

本研究前瞻性纳入2011至2016年使用导航软件(Syngo iGuide)行儿童经皮骨活检的26例患者,并与2008至2016年解剖匹配的33例CT引导骨活检进行回顾性比较。结果显示,导航软件辅助活检的技术成功率为100%(26/26),诊断率为96%(25/26),CT引导活检技术成功率为100%(33/33),诊断率为88%(29/33);导航软件组总体有效辐射剂量显著低于CT引导组(3.0±3.4 vs. 6.6±7.7 mSv,P=0.02),手术时间也显著缩短(91±54 vs. 141±68 min,P=0.005)。

要点

  • 导航软件辅助儿童骨活检的技术成功率为100%(26/26),诊断率为96%(25/26)。
  • 导航软件组总体有效辐射剂量显著低于CT引导组(3.0±3.4 vs. 6.6±7.7 mSv,P=0.02)。
  • 导航软件组手术时间显著短于CT引导组(91±54 vs. 141±68 min,P=0.005)。

核心结论

  • 作者经验认为,介入放射学套件中的导航软件技术是儿童骨活检中CT引导的有前景的替代方案,提供相当的技术成功率和诊断准确率,同时辐射剂量和手术时间更低,并提供实时透视针引导。

研究设计与样本

  • 前瞻性纳入2011至2016年使用导航软件(Syngo iGuide,Siemens Healthcare)行儿童经皮骨活检的26例患者,回顾性复习2008至2016年解剖匹配的CT引导骨活检33例,经机构审查委员会批准;导航软件辅助病例使用机构开发的低剂量(0.1/0.17 μGy/投影)、常规剂量(0.36 μGy/投影)或低剂量/常规剂量联合的C臂CT协议。

诊断标准与关键数值

  • 导航软件辅助活检26例中25例(96%)诊断,CT引导活检33例中29例(88%)诊断;两者技术成功率均为100%。
  • 导航软件组总体有效辐射剂量3.0±3.4 mSv vs. CT引导组6.6±7.7 mSv,P=0.02。
  • 使用低剂量C臂CT的导航软件病例有效剂量1.2±1.8 mSv vs. CT引导6.6±7.7 mSv,P=0.001。
  • 使用低剂量/常规剂量联合C臂CT的导航软件病例有效剂量1.9±2.4 mSv vs. CT引导6.6±7.7 mSv,P=0.04。
  • 使用常规剂量C臂CT的导航软件病例有效剂量6.0±3.5 mSv vs. CT引导6.6±7.7 mSv,P=0.7。

临床应用与阅片要点

  • 导航软件辅助活检组手术时间91±54 min,CT引导组141±68 min,P=0.005。
  • 导航软件组活检部位:骨盆15例、腰椎7例、下肢4例;CT引导组:骨盆22例、腰椎7例、下肢4例。
  • 导航软件组26例患者中男性15例,平均年龄10岁;CT引导组33例患者中男性13例,平均年龄9岁。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Navigational software provides real-time fluoroscopic needle guidance for percutaneous procedures in the Interventional Radiology (IR) suite.

OBJECTIVE: We describe our experience with navigational software for pediatric percutaneous bone biopsies in the IR suite and compare technical success, diagnostic accuracy, radiation dose and procedure time with that of CT-guided biopsies.

MATERIALS AND METHODS: Pediatric bone biopsies performed using navigational software (Syngo iGuide, Siemens Healthcare) from 2011 to 2016 were prospectively included and anatomically matched CT-guided bone biopsies from 2008 to 2016 were retrospectively reviewed with institutional review board approval. C-arm CT protocols used for navigational software-assisted cases included institution-developed low-dose (0.1/0.17 μGy/projection), regular-dose (0.36 μGy/projection), or a combination of low-dose/regular-dose protocols. Estimated effective radiation dose and procedure times were compared between software-assisted and CT-guided biopsies.

RESULTS: Twenty-six patients (15 male; mean age: 10 years) underwent software-assisted biopsies (15 pelvic, 7 lumbar and 4 lower extremity) and 33 patients (13 male; mean age: 9 years) underwent CT-guided biopsies (22 pelvic, 7 lumbar and 4 lower extremity). Both modality biopsies resulted in a 100% technical success rate. Twenty-five of 26 (96%) software-assisted and 29/33 (88%) CT-guided biopsies were diagnostic. Overall, the effective radiation dose was significantly lower in software-assisted than CT-guided cases (3.0±3.4 vs. 6.6±7.7 mSv, P=0.02). The effective dose difference was most dramatic in software-assisted cases using low-dose C-arm CT (1.2±1.8 vs. 6.6±7.7 mSv, P=0.001) or combined low-dose/regular-dose C-arm CT (1.9±2.4 vs. 6.6±7.7 mSv, P=0.04), whereas effective dose was comparable in software-assisted cases using regular-dose C-arm CT (6.0±3.5 vs. 6.6±7.7 mSv, P=0.7). Mean procedure time was significantly lower for software-assisted cases (91±54 vs. 141±68 min, P=0.005).

CONCLUSION: In our experience, navigational software technology in the IR suite is a promising alternative to CT guidance for pediatric bone biopsies providing comparable technical success and diagnostic accuracy with lower radiation dose and procedure time, in addition to providing real-time fluoroscopic needle guidance.

来源:Pediatr Radiol,查看原文。版权归原作者所有。

引用

  1. Shellikeri S, Setser RM, Hwang TJ, et al. Real-time fluoroscopic needle guidance in the interventional radiology suite using navigational software for percutaneous bone biopsies in children. Pediatr Radiol. 2017. PMID: 28474255. doi:10.1007/s00247-017-3830-0

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。